Cary 100/300/4000/5000/6000i/7000 分光光度計 ユーザーガイド

Technical Reference Summary

Covered Models & Document Scope

Covers the Agilent Cary 100, 300, 4000, 5000, 6000i and 7000 UV-Vis-NIR spectrophotometers. Variant-specific content includes Cary 100/300 (voltage selection switch, transport screw, cell holder base, extended sample compartment nitrogen purge) and Cary 4000/5000/6000i/7000 (universal power supply, lamp turret, solid sample holders, removable sample compartment base). Solid sample holders are standard on Cary 5000/6000i/7000 and optional on Cary 4000. Mercury lamp is standard on Cary 4000/5000/6000i/7000.

Section-by-Section Technical Breakdown

pp. 12
Electrical Connections

Describes the rear-panel IEEE 488 (GPIB) connector and the sample-compartment accessory connectors. The 15-pin D-range connector carries two high-voltage pins whose ratings differ between Cary 4000/5000/7000, Cary 6000i and Cary 100/300. Cary 4000/5000/6000i/7000 have the accessory controller function built into the standard instrument.

  • Rear panel: IEEE 488 (GPIB) for Cary 100/300/4000/5000/6000i/7000 system connection.
  • Sample compartment: 15-pin D-range connector with two high-voltage pins: -1000 VDC and +125 VDC (Cary 4000/5000/7000), +85 VDC (Cary 6000i), -1150 VDC and 125 VDC (Cary 100/300).
  • Cary 100/300 with accessory controller board, and all Cary 4000/5000/6000i/7000: three accessory connectors in the sample compartment (low-voltage DC and digital lines) plus one accessory connector on the front of the instrument.
  • No separate accessory controller board is required for Cary 4000/5000/6000i/7000; the function is included in the standard instrument.
pp. 14
Software Startup and Registration

Explains how to launch Cary WinUV, select an application and complete first-time software registration using the product key printed on the UFD case cover. Registration must be completed by the user of the Agilent Cary UV-Vis-NIR spectrophotometer.

  • Double-click the Cary WinUV folder on the desktop, then select the desired application.
  • On first launch the Software Registration dialog appears; click Register, then Next.
  • Enter the product key printed on the cover of the Agilent Cary WinUV software UFD case.
  • A message confirms successful Agilent software registration; the application then opens ready for data collection.
  • For computers without internet access, refer to the Software Registration Help.
pp. 15
GPIB Communication

The computer must be fitted with a USB-GPIB-HS+ converter to connect to the Cary spectrophotometer. Installation requires powering the computer off, connecting the converter between instrument and computer USB port, then powering on.

  • A National Instruments USB-GPIB-HS+ converter is required on the computer.
  • Ensure Cary WinUV software is installed before fitting the converter.
  • Power off the computer, connect one end of the converter to the instrument and the other to a USB port on the rear of the computer, then power the computer on.
pp. 16
Instrument Setup and Wiring

Covers the system connection diagram and minimum clearances for air circulation. The printer and spectrophotometer connect to the computer; the spectrophotometer uses a shielded IEEE-488 cable to the computer interface board.

  • Allow at least 50 mm clearance at the sides and 150 mm at the rear of the system for air circulation.
  • Connect the printer and spectrophotometer to the computer with cables plugged into the rear of the computer.
  • Connect the spectrophotometer to the computer with a shielded IEEE-488 cable: one end to the computer interface board, the other to the socket on the rear of the spectrophotometer.
pp. 17
Power Connection

Describes mains connection, voltage selection and fuse checks before powering the system. Cary 100/300 have a rear-panel voltage selection switch; Cary 4000/5000/6000i/7000 use a universal power supply. Two operator-accessible fuses are located at the rear.

  • Before connecting, ensure both instrument and mains are off and the rear-panel voltage selection switch (Cary 100/300 only) is set to the correct mains voltage per the rear-panel table.
  • Check the two operator-accessible fuses at the rear for correct type and rating before connecting to mains.
  • Connect the mains cord to the rear of the instrument and the free end to the mains, then switch on.
  • Cary 4000/5000/6000i/7000 have a universal power supply but the correct voltage must still be confirmed.
pp. 17
System Power-On

Pre-power checks and the power-on procedure, including the Cary 300 transport screw caution and the recommended 2-hour warm-up. The rocker switch marked '|' is at the front left of the instrument.

  • Verify components are interconnected, voltage tap settings checked, and the instrument connected to power as described on page 16.
  • Ensure the sample compartment is empty and its cover is closed.
  • For Cary 300, loosen the wavelength drive transport screw approximately 3 turns with a flat-blade screwdriver before powering on; failure to do so damages the instrument. The screw is on the right side of the rear panel (viewed from the rear) and is labelled.
  • Press the rocker switch marked '|' at the front left of the instrument to power on.
  • Allow the Cary instrument to warm up for 2 hours before use for optimum performance.
pp. 18
Instrument Performance Tests

The Cary WinUV validation application can run a number of performance tests to confirm that key instrument parameters meet specification. If any test fails, Agilent must be contacted to diagnose and correct the problem.

  • Use the validation application in Cary WinUV to verify a major subset of instrument parameters against specifications.
  • Many applications include validation files to confirm the software is installed correctly.
  • If any test does not meet specification, arrange for Agilent personnel to diagnose and correct the issue.
pp. 18-20
Cell Holder Installation

Installation of the cell holder base and single cell holders for Cary 100/300 and Cary 4000/5000/6000i/7000. Cary 100/300 ship with a cell holder base and two single cell holders installed; Cary 4000/5000/6000i/7000 ship with two single cell holders and a base plate.

  • Cary 100/300: remove the front panel, loosen the two black screws at the rear of the sample compartment and remove the front-right black screw, then fit the cell holder base with keyholes to the rear screws and the front onto the locating pin.
  • Cary 4000/5000/6000i/7000: place the cell holder base at the sample position on the base plate with the notched pillar on the right; tighten the two locating screws with a 1/8-inch hex ball driver.
  • In standard or double mode the sample position is at the front of the sample compartment; in reverse mode it is at the rear.
  • The cell holder fits in only one orientation; secure with the thumbscrew.
pp. 20-21
Single Cell Holder Alignment

Alignment procedure for all instruments: set zero order in Cary WinUV Align, place white paper in the light path and adjust the three adjustment screws until the beam passes through the centre of the cell holder aperture, then lock the adjustment screws with the lock nuts.

  • In Cary WinUV Align, select Zero Order under Instrument Parameters and click Apply.
  • Place white paper in the light path; the beam must pass through the centre of the cell holder aperture.
  • If off-centre, adjust the three adjustment screws on the cell holder base with a 2.5 mm hex ball driver.
  • Carefully tighten the lock nuts under the cell holder base to fix the adjustment screw height.
pp. 21-23
Solid Sample Holder (Cary 4000/5000/6000i/7000 only)

Cary 5000/6000i/7000 include two solid sample holders; for Cary 4000 the solid sample holder is optional. Each holder includes four solid sample slides with different aperture sizes (standard, 1, 5 and 10 mm), two V holders (large and small) and spacers for delicate samples.

  • Four solid sample slides with aperture sizes: standard, 1 mm, 5 mm and 10 mm.
  • Two V holders (large and small) secure the slide with two screws; spacers are placed between V holder and slide for delicate samples, which are held only at the edges.
  • Alternatively, a locking plate can be fitted to the slide using two 40 mm pins, a configuration suited to large samples.
  • The solid sample holder must be set on the base plate before mounting on the instrument.
  • Set the slide in the slide holder so the beam is blocked by the slide until it reaches the sample, minimising scattered light.
pp. 24
Solid Sample Holder Alignment

Alignment procedure for the solid sample holder: set zero order, place white paper in the light path, move the slide holder along the optical rail to centre it, then adjust the three adjustment screws on the optical rail base with a 2.5 mm hex ball driver.

  • Set the wavelength to 0 nm (zero order) in Cary WinUV and open Align.
  • Place white paper in the light path; the beam must pass through the centre of the aperture.
  • First move the slide holder along the optical rail toward the centre of the sample compartment and tighten the clamp screw.
  • Then adjust the three adjustment screws on the optical rail base with a 2.5 mm hex ball driver until the beam is centred.
  • Polarizer/depolarizer accessories and additional slide holders are available from Agilent.
pp. 26
Base Plate Installation

The base plate must be secured in position in the sample compartment using the lockdown mechanism: two posts at the rear and a clamp post at the front, with a lever at the front of the sample compartment.

  • Hold the base plate (with accessories) by hand with the posts toward the rear of the sample compartment.
  • Push the base plate rearward so the two rear posts engage under the clamps in the sample compartment floor.
  • Lower the front edge of the base plate so the front post engages the lockdown.
  • Move the lever at the front of the sample compartment to the left to lock the base plate in place.
  • After fitting, align the accessory as described on page 24.
pp. 27-29
Nitrogen Purge

Nitrogen purge is possible on Cary 100/300 only with the extended sample compartment, and on Cary 4000/5000/6000i/7000 using a user-supplied purge system. Recommended supply pressure is 83-172 kPa (12-25 psig); flow rates are 0-20 L/min for the instrument and 0-10 L/min for the sample compartment.

  • Cary 100/300: connect nitrogen tubing to the inlet tube under the extended sample compartment.
  • Cary 4000/5000/6000i/7000: Agilent does not supply a nitrogen purge system; obtain the items from a suitable supplier.
  • Recommended nitrogen: ultra-high-purity liquid nitrogen such as Nitrogen 5.0 (with heat exchanger); compressed nitrogen must be dry, oil-free and uncontaminated.
  • Do not use compressed nitrogen from suppliers whose compression process uses oil or water; use nitrogen filled by a liquid-nitrogen-lubricated immersion pump.
  • Operating pressure: 83-172 kPa (12-25 psig).
  • Supply tubing: clean flexible plastic (PVC or equivalent), 6 mm (1/4 inch) internal diameter; do not use rubber tubing because its internal talc can enter the optical system.
  • Flow rates: instrument 0-20 L/min; sample compartment 0-10 L/min.
  • Purge is recommended for measurements below 200 nm, across water absorption bands near 1370 and 1850 nm (Cary 5000/6000i/7000) and 2580 and 2750 nm (Cary 5000/7000), when the optics could be damaged by ambient atmosphere or particles, and for long-term continuous UV operation.
  • Purge success can be judged by scanning oxygen in the UV-Vis region and water vapour in the NIR region (Cary 5000/6000i/7000 only).
pp. 29-30
Sample Compartment Base Removal

Cary 100/300 have a rectangular base plate that must be removed to fit accessories connecting to sockets below the sample compartment and refitted when the accessory is removed to prevent leak damage. Cary 4000/5000/6000i/7000 have a fully removable sample compartment base for large accessories.

  • Cary 100/300: refit the rectangular base plate after removing an accessory to prevent connector damage from liquid leaks.
  • Cary 4000/5000/6000i/7000: the sample compartment base is fully removable to accommodate large accessories.
  • Removing the sample compartment floor disturbs optical alignment to the floor; realign the optics before microcell analysis.
  • Removal sequence: remove front door, remove cover plate between floor and horizontal projection, loosen two base screws and remove horizontal projection, remove lower cover plate, release fasteners holding the floor, remove the floor (the lockdown mechanism is also removed).
pp. 32
Cleaning

Liquid spills in the sample compartment must be wiped up immediately and deposits removed from the sample compartment windows. The exterior is cleaned with a soft cloth, moistened with water or mild detergent if necessary; organic solvents and abrasive cleaners must not be used.

  • Wipe up spills in the sample compartment immediately and remove deposits from the sample compartment windows.
  • Keep the exterior clean using a soft cloth; if needed, moisten with water or mild detergent and wring out well.
  • Do not use organic solvents or abrasive cleaning agents.
pp. 32
Light Source Lamps

Replacement and alignment of the Vis and UV lamps is described in the software help. Lamps and brackets remain hot after switch-off, and UV and mercury lamps operate at high voltage; the instrument must be switched off and the power cable removed before replacement.

  • Switch off the instrument and remove the power cable, then allow the lamp to cool before replacing the deuterium or Vis lamp.
  • Lamp surfaces and mounting brackets remain hot for some time after switch-off.
  • UV and mercury lamps operate at high voltage.
  • Do not touch the glass envelope of a new lamp; handle the lamp by its base using a soft cloth, since touching the glass reduces efficiency.
pp. 33
Fuses

The spectrophotometer has two fuses at the rear. Replacement requires unplugging the instrument and fitting fuses of the type and rating given in the pre-installation guide or on the rear panel. Repeated fuse blowing indicates another problem and may require service.

  • Two fuses are located at the rear of the instrument.
  • Unplug the spectrophotometer before replacing a fuse.
  • Replace with the type and rating stated in the pre-installation guide or on the rear panel; the fuse information printed on the rear panel is current.
  • Verify that the code stamped on the fuse matches the information next to the fuse holder.
  • To check a fuse: disconnect from mains, press and turn the cap counter-clockwise, pull the cap out (the fuse is inside), inspect/replace, refit and turn clockwise, then reconnect to mains.
  • If fuses blow repeatedly, the Cary instrument may have another fault and may need repair.
  • Do not access other internal fuses or circuit breakers; these are replaced by Agilent-certified personnel.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Minimum clearance for air circulation is 50 mm at the sides and 150 mm at the rear of the system. p.16 ↗ Bench space must be planned with these clearances to avoid overheating. whole series
The sample compartment 15-pin D-range connector carries high-voltage pins rated -1000 VDC and +125 VDC on Cary 4000/5000/7000, +85 VDC on Cary 6000i, and -1150 VDC and 125 VDC on Cary 100/300. p.12 ↗ Only Agilent internal and external DRA accessories may be connected to the high-voltage connector; other accessories must use the low-voltage DC and digital connectors. Cary 100/300/4000/5000/6000i/7000
Cary 4000/5000/6000i/7000 have a universal power supply, while Cary 100/300 have a rear-panel voltage selection switch that must be set to the local mains voltage. p.17 ↗ Cary 100/300 must be configured for the site mains before connection; Cary 4000/5000/6000i/7000 still require verification of the correct supply voltage. Cary 100/300 vs Cary 4000/5000/6000i/7000
On Cary 300 the wavelength drive transport screw must be loosened approximately 3 turns with a flat-blade screwdriver before power-on. p.17 ↗ Powering on a Cary 300 without loosening the transport screw will damage the instrument. Cary 300
A 2-hour warm-up is required before use for optimum performance. p.17 ↗ Schedule instrument power-on at least 2 hours before measurements. whole series
Nitrogen purge operating pressure is 83-172 kPa (12-25 psig). p.27 ↗ A suitable pressure regulator and gauge must be installed to hold this range. Cary 100/300 with extended sample compartment; Cary 4000/5000/6000i/7000
Nitrogen purge flow rates are 0-20 L/min for the instrument and 0-10 L/min for the sample compartment. p.29 ↗ Flow may need to be increased in the sample compartment when samples are frequently exchanged. Cary 100/300 with extended sample compartment; Cary 4000/5000/6000i/7000
Nitrogen supply tubing must be clean flexible plastic (PVC or equivalent) with 6 mm (1/4 inch) internal diameter; rubber tubing must not be used. p.27 ↗ Rubber tubing carries internal talc that can enter and contaminate the optical system. Cary 100/300 with extended sample compartment; Cary 4000/5000/6000i/7000
Compressed nitrogen must be dry, oil-free and uncontaminated; nitrogen from suppliers using oil- or water-lubricated compression must not be used. p.27 ↗ Oil mist in the purge gas forms oily deposits on optical surfaces and voids the instrument warranty. Cary 100/300 with extended sample compartment; Cary 4000/5000/6000i/7000
Cell holder base adjustment screws are adjusted with a 2.5 mm hex ball driver; locating screws are tightened with a 1/8-inch hex ball driver. p.20 ↗ These two tools are required for cell holder and solid sample holder installation and alignment. whole series
In standard or double mode the sample position is at the front of the sample compartment and the reference position at the rear; in reverse mode the positions are reversed. p.19 ↗ Accessory mounting positions must be chosen according to the operating mode. Cary 4000/5000/6000i/7000
Solid sample slides are supplied with aperture sizes standard, 1 mm, 5 mm and 10 mm; two V holders (large and small) and spacers are included. p.21 ↗ Aperture size and V holder size must be selected to suit the sample dimensions; delicate samples are held only at the edges with spacers. Cary 4000/5000/6000i/7000
Removing the sample compartment floor on Cary 4000/5000/6000i/7000 disturbs optical alignment to the floor. p.29 ↗ Optical alignment must be redone before microcell analysis after the floor has been removed. Cary 4000/5000/6000i/7000
The Cary 100/300 rectangular sample compartment base plate must be refitted after accessory removal. p.29 ↗ Leaving the plate out exposes the connectors below the sample compartment to liquid-leak damage. Cary 100/300
The spectrophotometer has two operator-accessible fuses at the rear; the fuse code stamped on the fuse must match the information next to the fuse holder. p.33 ↗ Incorrect fuse type or rating can reduce safety or cause nuisance blowing. whole series
A National Instruments USB-GPIB-HS+ converter is required on the computer for GPIB communication with the Cary spectrophotometer. p.15 ↗ The converter must be installed with the computer powered off and Cary WinUV already installed. whole series
The product key for Cary WinUV is printed on the cover of the software UFD case. p.14 ↗ The UFD case must be retained for first-time software registration. whole series
The Cary WinUV validation application can verify a major subset of instrument parameters against specification. p.18 ↗ A failed validation test requires Agilent personnel to diagnose and correct the problem. whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Fuse blows repeatedly p.33 ↗ The Cary instrument may have another problem. Arrange for repair service. Fuses, page 33
Performance test does not meet specification p.18 ↗ Instrument parameter out of specification (cause not further specified). Arrange for Agilent personnel to diagnose and correct the problem. Instrument Performance Tests, page 18