Agilent NovoCyte Opteon Spectral Flow Cytometer Specifications
Technical Reference Summary
Covered Models & Document Scope
Agilent NovoCyte Opteon Spectral Flow Cytometer; configurations with four lasers (349 nm 20 mW, 405 nm 130 mW, 488 nm 100 mW, 561 nm 100 mW, 637 nm 120 mW) and 8, 11, 14, 18 or 19 detectors; custom configurations available on request
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Laser type | Solid state laser with onboard thermal-electric cooling p.1 ↗ | |
| Laser beam configuration | Spacially separated; fixed laser delay and no need for adjustment p.1 ↗ | |
| Photodetector | Avalanche photodiode (APD) p.1 ↗ | |
| Optical alignment | Fixed. No operator alignment required. p.1 ↗ | |
| Forward scatter detection laser | 488 nm p.1 ↗ | FSC |
| Side scatter detection laser | BSSC 488 nm; VSSC 405 nm nm p.1 ↗ | SSC |
| FSC sensitivity | 0.4 µm p.1 ↗ | |
| VSSC sensitivity | 0.08 µm p.1 ↗ | |
| BSSC sensitivity | 0.2 µm p.1 ↗ | |
| FITC fluorescence threshold sensitivity | ≤ 50 MESF p.1 ↗ | B543 channel |
| PE fluorescence threshold sensitivity | ≤ 10 MESF p.1 ↗ | Y582 channel |
| APC fluorescence threshold sensitivity | ≤ 10 MESF p.1 ↗ | R661 channel |
| Pacific Blue fluorescence threshold sensitivity | ≤ 30 MESF p.1 ↗ | V455 channel |
| Fluorescence resolution | < 3 % CV p.1 ↗ | CEN |
| Flow cell dimensions | 170 x 490 µm p.2 ↗ | rectangular quartz flow cell |
| Sample acquisition rate | 100,000 events/second p.2 ↗ | |
| Volumetric absolute count precision | ≤ 5 % p.2 ↗ | |
| Sample flow rate | 5 to 120 µL/min p.2 ↗ | 1 µL increments |
| Sample acquisition volume | 1 µL to 5 mL p.2 ↗ | |
| Sheath fluid container capacity | 6 L p.2 ↗ | |
| Waste fluid container capacity | 6 L p.2 ↗ | |
| Cleaning fluid container capacity | 500 mL p.2 ↗ | |
| Decontamination fluid container capacity | 500 mL p.2 ↗ | automatic liquid volume sensing |
| Number of detectors | 8, 11, 14, 18, 19 p.3 ↗ | four-laser configuration |
Operating Limits
- Sample flow rate range: 5 to 120 µL/min p.2 ↗
- Sample acquisition volume range: 1 µL to 5 mL p.2 ↗
- Volumetric absolute count precision: ≤ 5% p.2 ↗