Profibus DP/V1 Interface for Bayard-Alpert Pirani Gauge FRG-730

Technical Reference Summary

Covered Models & Document Scope

Profibus DP/V1 interface functionality and programming for the Agilent FRG-730 Bayard-Alpert Pirani combination gauge. Covers part numbers FRG730KF25SP (DN 25 ISO-KF) and FRG730CF35SP (DN 40 CF-R).

Section-by-Section Technical Breakdown

pp. 6-7
General Data

Defines the Profibus communication basics of the FRG-730: supported data rates up to 12 Mbaud, node address setting via two hexadecimal rotary switches, the PNO ident number, configuration data for standard telegrams, user parameter data for pressure units, and the distinction between cyclic and acyclic communication.

  • Supports all EN 50170 data rates up to 12 Mbaud with automatic or fixed data rate setting.
  • Node address 0…125dec (00…7Dhex) set via MSD/LSD ADDRESS switches; default 5Chex; values >7Dhex are ignored and the stored address remains valid.
  • Ident number 0x09AA.
  • Configuration data strings are provided for standard telegram combinations (e.g. 0x44, 0x84, 0x05, 0x05, 0x05, 0x03 for telegram 4).
  • User parameter data string selects pressure unit: COUNTS 00 00 00 03 E9, Torr 00 00 00 05 15, Micron 00 00 00 05 16, mbar 00 00 00 05 1C, Pascal 00 00 00 05 1D.
  • Acyclic traffic for device/process settings; cyclic traffic for continuous pressure and status exchange.
pp. 8-11
Data Exchange Mode

Describes the slot/index addressing scheme, the block model (Device Block in Slot 0, transducer and function blocks in Slot 1), the structure of cyclic input/output telegrams with an 8-byte parameter channel plus up to 15 bytes process data, and the PKE/IND/PWE parameter channel protocol including instruction signatures and error codes.

  • 254 indices per slot, index width up to 255 bytes; parameters numbered ascending from index 16, operations descending from index 15.
  • Input data: 8-byte parameter channel plus up to 15 bytes process data; output data: 8-byte parameter channel plus up to 10 bytes process data.
  • Parameter channel supports read/write of max 4 bytes; strings cannot be read.
  • Exactly one instruction per output telegram and one response per input telegram; 4 bytes of actual data per cycle.
  • PKE bits 15…12 = AK instruction/response signature, bits 7…0 = slot.
  • Error codes: 0 undefined slot, 1 parameter not changeable, 2 range limit overflow, 3 index error, 5 data type error, 17 instruction not allowed in this state, 18 other errors, 201 already in requested state, 202 object state conflict.
pp. 12
Cyclic Message Telegrams

Lists the implemented standard telegrams 1 through 7 with their byte-level meaning for master-to-slave and slave-to-master directions, including transition commands for hot ion gauges, parameter channel, exception status, One Of N status extension, PV selector, and UINT16 or float process values.

  • Telegram 1 (M→S): byte 0 transition command, byte 1 transition command value for hot ion gauges, bytes 2…9 parameter channel.
  • Telegram 4 (S→M): byte 0 exception status, byte 1 One Of N status extension, byte 2 One Of N PV selector, bytes 3…4 process value UINT16.
  • Telegram 5 (S→M): byte 0 exception status, byte 1 One Of N status extension, byte 2 One Of N PV selector, bytes 3…6 process value float.
  • Telegram 6 (S→M): bytes 0…7 parameter channel, byte 8 exception status, byte 9 One Of N status extension, byte 10 One Of N PV selector, bytes 11…12 process value UINT16.
  • Telegram 7 (S→M): bytes 0…7 parameter channel, byte 8 exception status, byte 9 One Of N status extension, byte 10 One Of N PV selector, bytes 11…14 process value float.
pp. 14-21
Device Block

Documents the Device Block parameters (IDs 15–28 and 204–211), including block type, device type, standard revision, manufacturer, model number, firmware/hardware revisions, serial number, device state, exception status, and exception detail alarm/warning records with their bit-level definitions.

  • Block Type ID 16 = 1 for Device Block; Device Type ID 17 = "CG" (Combination Gauge).
  • Standard Revision Level ID 18 = "E54-0997"; Device Manufacturer ID 19 = "Agilent".
  • Device State ID 25 values: 0 Undefined, 1 Self testing, 2 Idle, 3 Self test exception, 4 Executing, 5 Abort, 6 Critical fault.
  • Exception Status ID 26 bit 7 is constantly 1 (expanded format); bits 0–2 alarm groups, bits 4–6 warning groups.
  • Exception Detail Alarm ID 27 is 10 bytes; Exception Detail Warning ID 28 is 11 bytes.
  • Common Exception Detail bits: bit 2 EPROM exception, bit 3 EEPROM exception, bit 4 RAM exception.
  • Device Exception Detail Alarm byte 2 bit 0 = Sensor filament 1 Alarm, bit 1 = Sensor filament 2 Alarm (BA); byte 3 bit 1 = Electronics/sensor error.
  • Device Exception Detail Warning byte 7 bit 3 = Pressure too high for degas (BA); byte 8 bit 1 = Electronics/sensor warning.
  • Manufacturer Exception Detail 0 = 1 indicates internal communication error.
pp. 22-23
Device Block, Device Behavior

Describes the device state machine (NORMAL, RUNNING, SELFTESTING, IDLE, SELFTEST EXCEPTION, EXECUTING, ABORT, CRITICAL FAULT) and the Device Block State Command (ID 15) used to force transitions such as Reset, Abort, Recover, Execute, Stop, and Perform Diagnostic.

  • After power-up the gauge runs INIT and SELFTESTING, then goes to IDLE or SELFTEST EXCEPTION.
  • Cyclic data traffic automatically brings the gauge to EXECUTING; acyclic traffic requires an explicit START service.
  • In IDLE, SELFTEST EXCEPTION, and CRITICAL FAULT the Safe State value is output as measured value.
  • State Command values: 0 Inactive, 1 Reset, 2 Abort, 3 Recover, 4 Execute, 5 Stop, 6 Perform Diagnostic.
pp. 24
Analog Input Block

Covers the One Of N Analog Input Function Block in Slot 1, including AI Block Adjust Command, Block Type, Channel Instance Selector, and PV Selector, and defines the pressure ranges in which the Pirani or BA measuring system is active.

  • AI Block Adjust Command ID 15: value 1 starts Full Scale Adjust.
  • Block Type ID 16 = 3.
  • Channel Instance Selector ID 46 selects which instance's attributes are mapped into the address range.
  • PV Selector ID 47 determines which instance's measured value is copied into cyclic output data.
  • Pirani active for 5.5×10-3 < p ≤ 1000 mbar; BA active for 5×10-10 < p < 2×10-2 mbar.
  • In the overlap 5.5×10-3 … 2×10-2 mbar the two measurement principles are averaged and the active instance is Pirani.
  • Measured values of Instances 3 and 4 (switching thresholds) are not output in cyclic data.
pp. 25-27
Analog Sensor Input Function Block SLOT 1 / Instance 1

Defines the Pirani measuring-part instance attributes: Process Value, Status, Data Type, Data Units, Reading Valid, Full Scale, Safe State, Safe Value, Overrange, Underrange, and the Pirani-specific AI Block Adjust Command for zero and gain adjustment.

  • Process Value ID 19 is output in the selected data unit and data type; if not EXECUTING, the Safe State value is output.
  • COUNTS conversion: PVmbar = 10^(COUNTS/2000) − 12.5; PVTorr = 0.75006168 × PVmbar; PVMicron = 10^-3 × PVTorr; PVPa = 100 × PVmbar.
  • Data Type ID 21: 3 = Integer16, 8 = Float (default); changeable only in IDLE status.
  • Data Unit ID 22: 1001 COUNTS, 1301 Torr, 1302 mTorr (Micron), 1308 mbar, 1309 Pascal; changeable only in IDLE status.
  • Reading Valid ID 23 = 1 only when EXECUTING, no error, and measured value between underrange and overrange.
  • Safe State ID 39: 0 Zero, 1 Full Scale, 2 Hold Last Value, 3 Use Safe Value.
  • AI Block Adjust Command ID 15: 0 Zero Adjust (Pirani zero point), 1 Gain Adjust (Pirani fullscale).
pp. 28-29
Analog Sensor Input Function Block SLOT 1 / Instance 2

Defines the BA measuring-part instance with the same attribute set as Instance 1 (Process Value, Status, Data Type, Data Units, Reading Valid, Full Scale, Safe State, Safe Value, Overrange, Underrange) and the same COUNTS conversion formulas.

  • Process Value ID 19 contains the BA measured value in the selected data unit and data type.
  • If the device is not in EXECUTING, the Safe State value is output.
  • Data Type and Data Unit settings apply to all instances.
  • COUNTS conversion formulas identical to Instance 1.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
The FRG-730 supports all EN 50170 data rates up to 12 Mbaud with automatic data rate setting or a fixed rate. p.6 ↗ The Profibus master can be configured for any standard baud rate up to 12 Mbaud; no manual baud rate matching is required if automatic setting is enabled. FRG-730
Node address is set via two hexadecimal rotary switches (MSD high nibble, LSD low nibble) in the range 00…7Dhex (0…125dec); default is 5Chex. p.6 ↗ Address must be set on the gauge before installation; values above 7Dhex are ignored and the previously stored address remains active. FRG-730
The PNO ident number for the FRG-730 is 0x09AA. p.6 ↗ The Profibus master GSD configuration must use ident number 0x09AA to recognize the gauge. FRG-730
Configuration data for standard telegram 4 (slave→master only) is 0x44, 0x84, 0x05, 0x05, 0x05, 0x03. p.7 ↗ This exact byte string must be sent during the configuration phase when telegram 4 is selected. FRG-730
User parameter data string for mbar is 00 00 00 05 1C; for Torr 00 00 00 05 15; for Micron 00 00 00 05 16; for Pascal 00 00 00 05 1D; for COUNTS 00 00 00 03 E9. p.7 ↗ The pressure unit must be selected via the user parameter data string during configuration; it cannot be changed while cyclically exchanging data with a DP/V0 master. FRG-730
The parameter channel is 8 bytes and allows reading/writing slave parameters with a maximum data length of 4 bytes; strings cannot be read. p.10 ↗ Only byte, word, double-word, and float values can be transferred via the parameter channel; string parameters require acyclic DPV1 services. FRG-730
Exactly one instruction per output telegram and exactly one response per input telegram can be transmitted; 4 bytes of actual data per cycle. p.10 ↗ Parameter channel throughput is limited to one transaction per bus cycle; plan acyclic access for bulk parameter reads. FRG-730
Pirani is the active measuring system for 5.5×10-3 < p ≤ 1000 mbar; BA is active for 5×10-10 < p < 2×10-2 mbar; in the overlap 5.5×10-3 … 2×10-2 mbar the two principles are averaged and Pirani is the active instance. p.24 ↗ The PV Selector automatically switches the cyclic output between Pirani and BA values; the overlap region uses an averaged value. FRG-730
Data Type and Data Unit settings can only be modified when the gauge is in the IDLE status; selecting a standard telegram overwrites all previously made acyclic data type settings. p.25 ↗ Configure data type and pressure unit before starting cyclic data exchange, or stop cyclic traffic and bring the gauge to IDLE to change them. FRG-730
Reading Valid ID 23 is set to 1 only when the gauge is in EXECUTING status, there is no error, and the measured value is lower than the overrange value and higher than the underrange value. p.26 ↗ A Reading Valid of 0 requires checking Exception Status ID 26 for errors or One Of N Status Extension ID 120 for overrange/underrange. FRG-730
Safe State ID 39 options: 0 Zero (PV set to 0), 1 Full Scale (PV set to full scale ID 24), 2 Hold Last Value, 3 Use Safe Value (PV set to ID 40). p.27 ↗ The Safe State determines the pressure value output when the gauge is not EXECUTING or has a device error; choose according to fail-safe requirements. FRG-730
AI Block Adjust Command ID 15 for Pirani: value 0 = Zero Adjust (zero point adjustment), value 1 = Gain Adjust (fullscale adjustment). p.27 ↗ Pirani zero and fullscale calibration can be triggered over Profibus by writing the corresponding state command. FRG-730
Device Block State Command ID 15 values: 0 Inactive, 1 Reset, 2 Abort, 3 Recover, 4 Execute, 5 Stop, 6 Perform Diagnostic. p.23 ↗ The gauge can be forced into ABORT, recovered to IDLE, or made to run a selftest via acyclic writes to ID 15. FRG-730
Exception Status ID 26 bit 7 is constantly 1 and marks the expanded error output format; bits 0–2 indicate alarm groups and bits 4–6 indicate warning groups. p.16 ↗ A nonzero Exception Status with bit 7 set is normal; only bits 0–6 carry actual alarm/warning information. FRG-730
Common Exception Detail Alarm bits: bit 2 EPROM exception, bit 3 EEPROM exception, bit 4 RAM exception. p.17 ↗ These bits identify memory-related hardware faults in the gauge. FRG-730
Device Exception Detail Alarm byte 2 (BA): bit 0 Sensor filament 1 Alarm, bit 1 Sensor filament 2 Alarm; byte 3 (BA): bit 1 Electronics/sensor error. p.18 ↗ Filament rupture or electronics/sensor faults on the BA measuring part are reported in these bits. FRG-730
Device Exception Detail Warning byte 7 (BA): bit 3 Pressure too high for degas; byte 8 (BA): bit 1 Electronics/sensor warning. p.20 ↗ Degas must not be started when this warning bit is set; the pressure is too high for the degas procedure. FRG-730
Manufacturer Exception Detail 0 is set to 1 if there is an internal communication error in the gauge. p.18 ↗ This bit indicates an internal communication fault rather than a sensor or process fault. FRG-730
COUNTS-to-pressure conversion: PVmbar = 10^(COUNTS/2000) − 12.5; PVTorr = 0.75006168 × PVmbar; PVMicron = 10^-3 × PVTorr; PVPa = 100 × PVmbar. p.25 ↗ When the COUNTS unit is selected, the master must apply these formulas to obtain a physical pressure value. FRG-730

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
0 p.11 ↗ Undefined slot Check the slot number in the PKE parameter signature value. Error code table, Section 2.2.1.3
1 p.11 ↗ Parameter not changeable Verify the parameter is writable in the current device state. Error code table, Section 2.2.1.3
2 p.11 ↗ Lower or upper value range limit overflow Write a value within the permitted range. Error code table, Section 2.2.1.3
3 p.11 ↗ Index error Check the index number addressed. Error code table, Section 2.2.1.3
5 p.11 ↗ Data type error Use the data type expected by the parameter. Error code table, Section 2.2.1.3
17 p.11 ↗ Instruction not allowed in this state Bring the gauge to a state in which the instruction is permitted (e.g. IDLE for data type/unit changes). Error code table, Section 2.2.1.3
18 p.11 ↗ Other errors Check device exception status and detail records for the underlying fault. Error code table, Section 2.2.1.3
201 p.11 ↗ Already in requested state No action required; the device is already in the requested state. Error code table, Section 2.2.1.3
202 p.11 ↗ Object state conflict Resolve the conflicting device state before repeating the instruction. Error code table, Section 2.2.1.3
EPROM exception p.17 ↗ EPROM fault (Common Exception Detail Alarm bit 2) Read Exception Detail Alarm ID 27; device hardware fault indicated. Section 3.1.1.10
EEPROM exception p.17 ↗ EEPROM fault (Common Exception Detail Alarm bit 3) Read Exception Detail Alarm ID 27; device hardware fault indicated. Section 3.1.1.10
RAM exception p.17 ↗ RAM fault (Common Exception Detail Alarm bit 4) Read Exception Detail Alarm ID 27; device hardware fault indicated. Section 3.1.1.10
Electronics / sensor error p.18 ↗ Electronics or sensor error (Device Exception Detail Alarm byte 1 bit 1 for Pirani; byte 3 bit 1 for BA) Read Exception Detail Alarm ID 27; check the corresponding measuring system. Section 3.1.1.10
Sensor filament 1 Alarm p.18 ↗ BA sensor filament 1 alarm (Device Exception Detail Alarm byte 2 bit 0) Read Exception Detail Alarm ID 27; check BA filament 1. Section 3.1.1.10
Sensor filament 2 Alarm p.18 ↗ BA sensor filament 2 alarm (Device Exception Detail Alarm byte 2 bit 1) Read Exception Detail Alarm ID 27; check BA filament 2. Section 3.1.1.10
Electronics / sensor warning p.20 ↗ Electronics or sensor warning (Device Exception Detail Warning byte 1 bit 1 for Pirani; byte 8 bit 1 for BA) Read Exception Detail Warning ID 28; check the corresponding measuring system. Section 3.1.1.11
Pressure too high for degas p.20 ↗ Pressure too high for degas (Device Exception Detail Warning byte 7 bit 3, BA) Do not start degas until pressure is sufficiently low. Section 3.1.1.11
SELFTEST EXCEPTION p.23 ↗ During the self test an error has been detected; cause can be found in Device Exception Detail Alarm/Warning of the Device Block. Read Device Exception Detail Alarm/Warning attributes; the Safe State value is output as measured value. Section 3.1.2, Device statuses
ABORT p.23 ↗ Device Block is in ABORT status; gauge-specific measurement functions are not carried out. Use Device Block State Command value 3 (Recover) to bring the device from ABORT to IDLE. Sections 3.1.2 and 3.1.2.1
CRITICAL FAULT p.23 ↗ A device error has occurred; gauge-specific measurement functions are not carried out. The value defined in the Safe State is output as measured value; check device exception details. Section 3.1.2, Device statuses

Technical Data

Parameter Value Condition
Data rate up to 12 Mbaud p.6 ↗ EN 50170, automatic or fixed
Node address range 0 … 125dec (00 … 7Dhex) p.6 ↗ set via ADDRESS rotary switches, default 5Chex
Ident number 0x09AA p.6 ↗ FRG-730
Pressure measurement range 5×10-10 … 1000 mbar p.5 ↗ intended use
Pirani active pressure range 5.5×10-3 < p ≤ 1000 mbar p.24 ↗ PV Selector = Pirani
BA active pressure range 5×10-10 < p < 2×10-2 mbar p.24 ↗ PV Selector = BA
Full scale 31000 COUNTS p.26 ↗ pressure unit COUNTS
Full scale 750.06168 Torr p.26 ↗ pressure unit Torr
Full scale 750061.68 mTorr (Micron) p.26 ↗ pressure unit Micron
Full scale 1000 mbar p.26 ↗ pressure unit mbar
Full scale 100000 Pascal p.26 ↗ pressure unit Pascal
Overrange 31000 COUNTS p.27 ↗ pressure unit COUNTS
Overrange 750.06168 Torr p.27 ↗ pressure unit Torr
Overrange 750061.68 mTorr (Micron) p.27 ↗ pressure unit Micron
Overrange 1000 mbar p.27 ↗ pressure unit mbar
Overrange 100000 Pascal p.27 ↗ pressure unit Pascal
Underrange 20480 COUNTS p.27 ↗ pressure unit COUNTS
Underrange 4.12534×10-3 Torr p.27 ↗ pressure unit Torr
Underrange 4.12534 mTorr (Micron) p.27 ↗ pressure unit Micron
Underrange 5.5×10-3 mbar p.27 ↗ pressure unit mbar
Underrange 0.55 Pascal p.27 ↗ pressure unit Pascal
Data type coding 3 p.26 ↗ Integer16
Data type coding 8 p.26 ↗ Float (default)
Pressure unit coding 1001 p.26 ↗ COUNTS
Pressure unit coding 1301 p.26 ↗ Torr
Pressure unit coding 1302 p.26 ↗ mTorr (Micron)
Pressure unit coding 1308 p.26 ↗ mbar
Pressure unit coding 1309 p.26 ↗ Pascal