DIN 12876 specifications table for the CC-315B unit
p. 2table, 4 rows
Technical data according to DIN 12876
Parameter
Value
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S393661
1.3/20
Engineering Facts
14 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method chosen, so the required range must be matched to water cooling or refrigerator option.
Determines whether the bath can reach the process temperatures without an external chiller or with the integrated refrigerator.
Source
“Operating temperature range 28...300 °C with water cooling 20...300 °C with refrigerator -20...300 °C”
Installation
The unit requires a 200-240 V single- or two-phase supply with up to 16 A current and a 20 A fuse, so the electrical supply and branch protection must be sized accordingly.
Prevents undersized wiring or breakers and ensures the correct mains configuration before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 16 A Fuse 20 A”
Installation
External branch circuit protection according to UL 489 is required, so the installer must provide a suitable UL-listed branch breaker.
This is a mandatory electrical installation requirement for compliance and safe operation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are limited to ±10% and must not run in opposite directions, so the site power quality must be checked before commissioning.
Out-of-tolerance supply can prevent safe operation or damage the unit and must be verified during site preparation.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug only if specified at ordering, so the plug type must be confirmed with the order.
Avoids delays and rework at installation if the wrong plug or no plug is supplied.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
Three-phase devices below 63 A are delivered with cable but without plug, so the customer must provide and fit the appropriate plug or connection.
Affects electrical installation planning and the need for a qualified electrician to terminate the supply.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The ambient temperature must stay between 5 °C and 40 °C, so the installation location must be kept within this range.
Operating outside this range can degrade performance or trip protective functions.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation
The degree of protection is IP20, so the unit must be installed in a dry indoor location protected from water and dust ingress.
Prevents unsuitable installation environments that could cause electrical hazards or failures.
Source
“Degree of Protection IP20”
Selection
The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s, so the thermofluid must be selected within this limit.
Higher viscosity can overload the pump and reduce circulation performance.
Source
“max. permissible kin. viscosity 50 mm²/s”
Selection
The minimum filling capacity is 11.5 l and the bath volume is 15.6 l, so the fluid charge and expansion volume must be planned accordingly.
Ensures sufficient fluid for safe operation and correct thermal performance.
Source
“min. filling capacity 11,5 l Bath volume 15,6 l”
Selection
Using a displacement rack reduces the available bath capacity to 8.5 l, so the usable volume for samples and fluid must account for this reduction.
Affects how much sample volume can be accommodated without overflowing or starving the pump.
Source
“Bath capacity with displacement rack 8,5 l”
Installation
The pump connection is M16x1 male, so hoses and adapters must match this thread.
Prevents mismatch of external circuit connections during installation.
Source
“Pump connection M16x1 male”
Performance
The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar, so external circuit pressure drop must be checked against these limits.
Ensures the pump can overcome the resistance of the external cooling or process loop.