Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is classified as safety class III / FL per DIN 12876, which determines the permitted flammable-liquid use and protective measures.
Safety class governs whether and how flammable bath fluids may be used and what over-temperature protection is required.
Source
“Safety classification
III / FL”
Selection
Cooling capacity drops from 0.4 kW at 20 °C to 0.065 kW at -30 °C, so the low-temperature working point must be checked against the actual heat load.
Sizing at the coldest set point prevents undersized cooling and long pull-down times.
Source
“at -30°C
0,065 kW”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which imposes ventilation and ignition-source restrictions on the installation site.
A3 flammability affects room ventilation, electrical zoning and site risk assessment.
Source
“R-290 (A3, H220)”
Installation
The unit is air-cooled, so sufficient clearance and cool intake air must be provided at the installation location.
Air-cooled condensers need unobstructed airflow; poor clearance reduces cooling capacity and can trip the compressor.