HB960 Immersion Circulator

Contents

  1. Overview p. 1
    Product introduction and main features of HB960
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 20 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range60...250 °C
Resolution of display0,01 K
Temperature stability at 70°C0,1 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationII / FL
Heating power96 kW
Circulation pump:MK pump
max. delivery200 l/min
max. delivery pressure5,5 bar
Pump connectionM38x1,5 male
Overall dimensions WxDxH800x1007x1607 mm
Degree of ProtectionIP20
Power supply (3 Phase)460V 3~ 60Hz
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is designed exclusively for external open operation and has no expansion tank, so the connected application must accommodate fluid filling and thermal volume changes.

Determines whether the unit can be used with a closed loop or requires an external open bath/reservoir with volume compensation.

Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Installation

Three-phase devices drawing less than 63 A are supplied with a cable but no plug, so the customer must provide and install a suitable plug or hard-wire connection.

Affects electrical installation planning, procurement of a plug, and compliance with local wiring codes.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

External branch circuit protection according to UL 489 is required, so the installer must provide a UL 489 listed breaker in the supply circuit.

Necessary for a compliant and safe electrical installation; affects panel design and component selection.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10% only if they do not run in opposite directions, so supply quality must be checked before commissioning.

Prevents misapplication on weak or unstable grids and defines acceptable supply conditions for warranty and reliable operation.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Operation

The minimum temperature of 60 °C depends on ambient temperature being below 60 °C and on the application heat loss, which must be between 0.1 kW and 2.8 kW depending on model.

Critical for verifying that the process load and ambient conditions allow the specified minimum temperature to be reached.

Source
“The heat loss must be between 0.1kW and 2.8kW, depending on the HB model, to reach the minimum temperature of 60°C.”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set correctly for the application to protect the fluid and process.

Affects safety configuration during commissioning and protects against overheating of the thermofluid and connected system.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation

The pump connection is M38x1.5 male, so mating hoses and fittings must match this thread.

Determines the adapters and hoses needed for connecting the external circuit and avoids delays during installation.

Source
“Pump connection M38x1,5 male”
Operation

The unit is rated IP20, so it must be installed in a dry, protected location and not exposed to water or dust ingress.

Constrains the installation environment and enclosure requirements to maintain safety and reliability.

Source
“Degree of Protection IP20”
Operation

The permissible ambient temperature range is 5 °C to 40 °C, so the installation site must be kept within these limits.

Affects site selection and room ventilation/cooling requirements for reliable operation.

Source
“min. ambient temperature 5 °C”
Compliance

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, so it is intended for industrial environments and may cause interference in residential areas.

Important for EMC planning and for deciding whether additional filtering or separation from sensitive equipment is needed.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Performance

The circulation pump delivers up to 200 l/min at up to 5.5 bar, which defines the achievable external circuit flow and pressure drop budget.

Used to check whether the external application's pressure drop and required flow can be met by the pump.

Source
“max. delivery 200 l/min”
Selection

The unit provides 96 kW of heating power, which must be matched to the thermal load and heat-up requirements of the application.

Determines heat-up time and whether the unit can hold setpoint under process load.

Source
“Heating power 96 kW”