KISS K25 Immersion Circulator

Contents

  1. Overview p. 1
    Product name and cover of KISS K25 datasheet
  2. Spec p. 2
    Technical data table according to DIN 12876
  3. p. 2 table, 7 rows

Technical data according to DIN 12876

CharacteristicValue
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S412560

Engineering Facts

13 facts · every claim linked to a page
Selection

The unit covers a working temperature range from -30 to 200 °C, so it suits both sub-ambient and high-temperature bath applications.

Defines whether the bath can reach the process temperatures required by the application.

Source
“Operating temperature range -30...200 °C”
Performance

Cooling capacity drops sharply at low temperatures, from 0.4 kW at 20 °C to only 0.065 kW at -30 °C, so heat loads must be sized against the actual set point.

Prevents oversizing the process heat load at low-temperature operation.

Source
“at -30°C 0,065 kW”
Safety

The refrigerant is R-290 (propane, A3 flammable, H220), which imposes flammable-refrigerant handling and ventilation requirements.

Affects site safety classification, ventilation and service procedures.

Source
“R-290 (A3, H220)”
Installation

The power supply must be 220-240 V single or two-phase at 50/60 Hz, and the refrigerated bath draws up to 2 A.

Determines the electrical supply and breaker sizing at the installation site.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz”
Installation

The circuit must be protected with a fuse between 10 A and 16 A.

Sets the acceptable upstream overcurrent protection for safe operation.

Source
“min. Fuse 10A max. Fuse 16A”
Installation

The enclosure is rated IP20, so it must be installed in a dry, protected location away from water spray.

Restricts where the unit can be placed in the laboratory or plant.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity falls as ambient temperature rises.

Requires climate control or derating of cooling performance in warm rooms.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Operation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Prevents operation on unstable mains that could damage the compressor or electronics.

Source
“-5% voltage and +2% frequency -> not allowed!”
Selection

The plastic pump is not resistant to acetone or polyglycols, so alternative fluids or a more resistant pump must be specified.

Avoids chemical attack on wetted pump parts when choosing the bath fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.

Prevents freezing and damage to the bath at sub-zero set points.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Ensures the correct plug is supplied for the site's electrical system.

Source
“with power cable and country-specific plug (please specify when ordering)”
Compliance

The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit provides adjustable overtemperature protection according to DIN 12876 and is rated safety class III/FL.

Confirms the built-in safety function required for unattended operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”