Product introduction and key features of the Minichiller 300-H OLÉ
Specp. 2
Technical data according to DIN 12876
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit uses flammable refrigerant R-290 (A3, H220), so the installation location must be assessed for flammable-refrigerant safety requirements.
Flammable refrigerant classification drives room ventilation, ignition-source and local code compliance decisions at selection and installation.
Source
“R-290 (A3, H220)”
Installation
The chiller is air-cooled with natural refrigerant and only 0.041 kg charge, so no large refrigerant volume handling is needed but airflow must be provided.
Air-cooled condensers need adequate ambient airflow; natural refrigerant limits charge and GWP obligations.
The unit has IP20 protection, so it must be kept dry and protected from water ingress.
IP20 rules out washdown or wet environments and affects placement and enclosure decisions.
Source
“Degree of Protection IP20”
Operation
The unit has an autostart function after power failure, so process restart behavior must be considered for unattended operation.
Automatic restart can be desirable or hazardous depending on the connected process and must be planned.
Source
“Autostart function for power failure”
Selection
An optional factory-fitted Pt100 sensor connection allows display (not control) of process temperature, so external monitoring must be specified at order.
If process temperature monitoring is needed, it must be ordered factory-fitted and cannot control the unit.
Source
“Option: Pt100 sensor connection #10519 to display (not control)”
Compliance
The equipment is EMC Class A Group 1, so it may cause radio interference in residential environments.
Class A equipment is intended for industrial use and may require mitigation in residential or sensitive settings.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.
The plug type must be declared at order to match the site's electrical outlet.
Source
“with power cable and country-specific plug (please specify when ordering)”