Product name and cover page for the Minichiller unit
Specp. 2
Technical data according to DIN 12876 standard
p. 2table, 5 rows
Technical data according to DIN 12876
Parameter
Value
Pressure equipment category Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S438237
Engineering Facts
12 facts · every claim linked to a page
Installation
The unit requires cooling water supply pressure between 3 and 6 bar, so the site water system must be verified before installation.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the water circuit.
Source
“min. cooling water differential pressure 3 bar”
Installation
Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if condensation conditions occur.
Out-of-range cooling water temperature reduces cooling capacity and can cause condensation and corrosion.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly as water temperature falls, from 0.3 kW at 15°C to 0.07 kW at -20°C, so sizing must account for the actual operating point.
Selecting the chiller based only on nominal capacity will underperform at low temperatures.
Source
“at -20°C 0,07 kW”
Safety
The refrigerant is R-290 (propane, A3 flammable), so the installation area must be assessed for flammable refrigerant handling and ventilation.
Flammable refrigerant classification affects site safety requirements and installation approval.
Source
“R-290 (A3, H220)”
Operation
The plastic pump is not resistant to acetone and polyglycols, so these fluids must not be used as heat transfer media.
Using incompatible fluids will damage the pump and void reliable operation.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Proper antifreeze mixture prevents freezing and damage in the cooling circuit at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Site power quality must be checked to avoid invalid supply combinations that could damage the unit.
Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance
The unit is classified as EN55011 Class A Group 1, meaning it is intended for industrial electromagnetic environments, not residential.
Class A equipment may cause interference in residential areas and affects where the unit can be installed.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.
IP20 limits installation to clean, dry environments and rules out washdown or outdoor placement.
Source
“Degree of Protection IP20”
Operation
Ambient temperature must remain between 5°C and 40°C for specified operation.
Exceeding ambient limits reduces cooling performance and may trip alarms or damage components.
Source
“max. ambient temperature 40 °C”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so cooling water must be compatible with these materials.
Incompatible water chemistry will cause corrosion or fouling in the water circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection
The unit is water-cooled with a natural refrigerant and very low GWP, which matters for sites with refrigerant regulations.
Water cooling avoids air-cooled condenser heat rejection and the natural refrigerant simplifies compliance.