Product introduction and key features of the Minichiller 600w-H
Specp. 2
Technical data according to DIN 12876 for the chiller
p. 2table, 18 rows
p. 1table, 23 rows
Technical data according to DIN 12876
Parameter
Value
max. delivery (suction)
18 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
Cooling water connection
G1/2 male
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
min. filling capacity
2,8 l
Volume of expansion
2,2 l
Overall dimensions WxDxH **
280x490x424 mm
Net weight
36 kg
Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
12 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S577946
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Minichiller 600w-H (Order-No.: 3066.0020.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Cooling power at ambient temperature +20°C at 15°C
0,6 kW
Cooling power at ambient temperature +20°C at 0°C
0,5 kW
Cooling power at ambient temperature +20°C at -10°C
0,35 kW
Cooling power at ambient temperature +20°C at -20°C
0,15 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,06 kg
CO2 equivalent
0,0 t
Circulation pump
Pressure- and suction pump
max. delivery
24 l/min
max. delivery pressure
0,7 bar
Значения приведены дословно по тексту страницы; блоки «switch for whisper mode» и «Pump data for whisper mode» в таблицу технических данных не включены.
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller requires cooling water with a minimum differential pressure of 3 bar and a maximum pressure of 6 bar, so the site water supply must be sized accordingly.
Ensures the cooling water circuit can deliver adequate flow and pressure without exceeding component ratings.
Source
“min. cooling water differential pressure 3 bar”
Installation
Cooling water inlet temperature must be between 5°C and 20°C; below the condensation point the cooling water pipes inside the unit must be insulated.
Prevents condensation and ensures rated cooling performance; dictates water supply design and insulation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power is rated at 0.6 kW at 15°C, 0.5 kW at 0°C, 0.35 kW at -10°C and 0.15 kW at -20°C, so capacity drops significantly at lower temperatures.
Allows correct sizing of the chiller for the actual process temperature and heat load.
Source
“Cooling power at ambient temperature +20°C”
Safety
The refrigerant is R-290 (propane, A3, H220) with a charge of 0.06 kg, which is flammable and requires appropriate safety measures.