Technical data table according to DIN 12876 for the chiller
p. 2table, 14 rows
Technical data according to DIN 12876
Parameter
Value
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
min. filling capacity
2,8 l
Volume of expansion
2,2 l
Overall dimensions WxDxH **
280x490x424 mm
Net weight
39 kg
Power supply requirement
110-120V 1~ 60Hz
max. current
8,5 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S582910
** Please respect space requirements. See operating conditions at www.huber-online.com
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires cooling water supply pressure between 3 bar minimum differential and 6 bar maximum, so the facility water system must be sized accordingly.
To verify the site cooling water loop can meet both the minimum differential pressure for adequate flow and the maximum pressure rating of the unit.
Source
“min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar”
Installation
The unit is designed for cooling water intake between 5°C and 20°C, and pipes must be insulated if water temperature can fall below the condensation point.
To confirm the available cooling water temperature range and plan insulation to prevent condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power ratings are valid only at 20°C room temperature, 15°C cooling water inlet and 3 bar differential pressure, so actual capacity will differ at other conditions.
To avoid overestimating cooling capacity when site conditions deviate from the reference values.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 3 bar differential pressure between cooling water inlet and -”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump option ordered.
To select a compatible pump material or alternative fluid before commissioning to prevent pump damage.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
To prepare a suitable heat transfer fluid that prevents freezing at low operating temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Compliance
Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not combine in the same direction (e.g., -5% voltage with +2% frequency is not allowed).
To assess whether the site power quality falls within the permitted combined tolerance envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects handling and ventilation requirements.
To apply appropriate safety measures for a flammable refrigerant during installation and service.
Source
“R-290 (A3, H220)”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.
To ensure the installation environment matches the enclosure protection rating.
Source
“Degree of Protection
IP20”
Installation
Ambient temperature must remain between 5°C and 40°C for proper operation.
To verify the room conditions where the chiller will be installed.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The power supply requirement is 110-120V 1~ 60Hz with a maximum current of 8.5 A and a 16 A fuse.
To confirm the electrical supply and breaker sizing at the installation site.
Source
“Power supply requirement
110-120V 1~ 60Hz
max. current
8,5 A
Fuse
16 A”
Installation
The pump and cooling water connections are M16x1 male and G1/2 male respectively, so matching fittings must be provided.
To procure correct adapters and hoses before installation.
Source
“Pump connection
M16x1 male
Cooling water connection
G1/2 male”
Operation
Whisper mode reduces sound pressure to 51 dB(A) ±4 dB(A) but also lowers pump delivery to 14 l/min and 0.2 bar.
To understand the trade-off between noise reduction and hydraulic performance when selecting the operating mode.
Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the cooling water must be compatible with these.
To specify suitable cooling water quality and avoid corrosion or deposits.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection
The permissible return line temperature is 80°C, which defines the upper limit for the return flow.
To ensure the return line temperature from the application does not exceed the unit's limit.