Contents

  1. Overview p. 1
    Product introduction and key features of TC45-F
  2. Spec p. 1
    Technical specifications and performance data table
  3. p. 1 table, 29 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range-45...100 °C
Cooling power
at 0°C0,24 kW
at -20°C0,18 kW
at -30°C0,1 kW
at -40°C0,05 kW
Safety classificationI / NFL
Refrigeration machineair-cooled, natural
refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,041 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Nominal diameter probe13 mm
Length of probe600 mm
Length flexible connection1050 mm
Overall dimensions WxDxH **190x295x360 mm
Net weight16 kg
sound pressure level +/- 4 dB(A)59 dB(A)
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current1,4 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S394802

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit's cooling capacity is only 0.05 kW at -40 °C, so verify the required heat load at the lowest operating temperature before selecting it.

Cooling capacity drops sharply at low temperatures, which determines whether the immersion cooler can hold the process setpoint.

Source
“at -40°C 0,05 kW”
Safety

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, so the installation area must be assessed for ignition sources and ventilation.

Flammable refrigerant classification affects site safety review, ventilation, and electrical zoning during installation.

Source
“R-290 (A3, H220)”
Installation

The device is rated IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

IP20 limits where the unit can be placed and what washdown or outdoor conditions it can tolerate.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 °C and 40 °C, so the room climate must be checked before siting the unit.

Exceeding the ambient range risks reduced cooling capacity and equipment protection trips.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity figures are valid at 20 °C room temperature and may fall at higher ambient temperatures, so capacity margins must account for the actual room temperature.

Sizing on nominal capacity without ambient derating can lead to insufficient cooling in warm rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Supply quality must be verified against this combined tolerance rule to avoid invalid operating conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.

The plug type must be confirmed at order time to avoid rework at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Compliance

The equipment is EMC Class A, Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Class A classification affects EMC planning and suitability for non-industrial locations.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The minimum fuse is 10 A and maximum fuse is 16 A, so the upstream protection must be sized within this window.

Correct upstream fuse rating is required for safe electrical installation and warranty compliance.

Source
“min. Fuse 10A max. Fuse 16A”
Selection

The probe is 13 mm in diameter and 600 mm long with a 1050 mm flexible connection, so vessel geometry and insertion depth must be checked.

Probe dimensions determine whether the cooler fits the vessel opening and reaches the required immersion depth.

Source
“Nominal diameter probe 13 mm Length of probe 600 mm Length flexible connection 1050 mm”
Operation

The refrigeration unit works continuously, so continuous operation and associated heat rejection to the room must be planned.

Continuous operation affects room heat load, energy consumption, and maintenance intervals.

Source
“The refrigeration unit works continiously.”
Compliance

The pressure equipment falls under Article 4.3 PED, so the applicable pressure equipment compliance route must be confirmed.

PED category determines documentation and conformity requirements for the installation.

Source
“Pressure equipment category Art. 4.3 PED”