Contents

  1. Overview p. 1
    Introduces the Unichiller 025T model and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and physical specifications per DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S591844

Technical data according to DIN 12876 — Unichiller 025T (Order-No.: 3054.0012.01)

ParameterValue
Operating temperature range-10...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C2,5 kW
Cooling power at 0°C1,2 kW
Cooling power at -10°C0,6 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,2 kg
CO2 equivalent1,7 t
UN-numberUN 2857
Circulation pump:B
max. delivery25 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar20 l/min
Delivery at 1,0 bar17 l/min
Delivery at 2,0 bar10 l/min
Delivery at 3,0 bar5 l/min
Pump connectionG3/4 male
min. filling capacity2,5 l
Volume of expansion2 l
Overall dimensions WxDxH **450x510x1230 mm
Net weight144 kg
sound pressure level +/- 4 dB(A)74 dB(A)
Power supply requirement230V 1~ 50Hz
max. current11 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -10 to 40 °C, so process requirements outside this window cannot be met.

Defines the process temperature envelope for which the unit can be selected.

Source
“Operating temperature range -10...40 °C”
Performance

Cooling capacity drops from 2.5 kW at 15 °C to 0.6 kW at -10 °C, so the required heat load must be checked at the actual operating temperature.

Prevents oversizing assumptions based on a single capacity figure.

Source
“Cooling power at 15°C 2,5 kW at 0°C 1,2 kW at -10°C 0,6 kW”
Installation

The unit requires a 230 V single-phase 50 Hz supply with a 16 A fuse and draws up to 11 A, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device rating.

Source
“Power supply requirement 230V 1~ 50Hz max. current 11 A Fuse 16 A”
Installation

The chiller is air-cooled and its cooling capacity is valid at 20 °C room temperature, so adequate ventilation and ambient conditions must be provided.

Ambient temperature directly affects achievable cooling performance and must be considered during siting.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The permitted ambient temperature range is 5 to 40 °C, so the installation location must stay within these limits.

Defines the acceptable room conditions for reliable operation.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Voltage and frequency tolerances are ±5 % and ±2 % respectively, but not simultaneously in the same direction, so supply quality must be verified.

Avoids invalid supply combinations that could damage the unit or void compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The refrigerant is R-449A with GHS classification A1/H280 and GWP 1397, so leak detection and refrigerant handling procedures are required.

Affects safety measures, environmental reporting and service planning.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Compliance

The unit is classified as pressure equipment category Art. 4.3 PED, so its installation must respect the applicable pressure equipment directive requirements.

Determines the regulatory framework for installation and commissioning.

Source
“Pressure equipment category Art. 4.3 PED”
Installation

The pump connection is G3/4 male, so mating hoses and fittings must match this thread.

Ensures correct hydraulic connection without adapters.

Source
“Pump connection G3/4 male”
Operation

The minimum filling capacity is 2.5 l and expansion volume is 2 l, so the external circuit must be filled and sized within these limits.

Prevents underfilling or overfilling of the expansion system.

Source
“min. filling capacity 2,5 l Volume of expansion 2 l”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts the installation environment and enclosure requirements.

Source
“Degree of Protection IP20”
Operation

The sound pressure level is 74 dB(A) ±4 dB(A), so noise exposure and acoustic planning must be considered at the installation site.

Affects workplace noise limits and location selection.

Source
“sound pressure level +/- 4 dB(A) 74 dB(A)”
Compliance

The equipment is EMC class A group 1 to EN55011, so it is intended for industrial environments and may cause interference in residential areas.

Determines suitability for industrial versus residential installation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are delivered with a country-specific plug that must be specified at ordering, so the correct plug type must be confirmed.

Avoids delivery of an incompatible mains plug.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”