Introduces the Unichiller 070Tw chiller unit and its main features
Specpp. 1-2
Lists performance, temperature, and physical specifications per DIN 12876
p. 2table, 15 rows
p. 1table, 35 rows
Technical data according to DIN 12876
Parameter
Value
Consumption at water 15°C, flow -10°C
264 l/h
min. cooling water differential pressure
1,5 bar
max. cooling water pressure
6 bar
min. filling capacity
5,9 l
Volume of expansion
7,5 l
Overall dimensions WxDxH **
600x600x1450 mm
Net weight
206 kg
Power supply (3 Phase)
400V 3~ 50Hz
max. current
9 A
Fuse (3 phase)
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S433366
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Characteristic
Unichiller 070Tw
Order-No.
3016.0030.01
Operating temperature range
-10...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 15°C
7 kW
Cooling power at 0°C
4,2 kW
Cooling power at -10°C
2,5 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,35 kg
CO2 equivalent
1,9 t
Circulation pump
C3
max. delivery
84 l/min
max. delivery pressure
5,6 bar
Delivery at 0,33 bar
81 l/min
Delivery at 0,5 bar
79 l/min
Delivery at 1,0 bar
74 l/min
Delivery at 1,5 bar
69 l/min
Delivery at 2,0 bar
64 l/min
Delivery at 2,5 bar
58 l/min
Delivery at 3,0 bar
52 l/min
Delivery at 3,5 bar
46 l/min
Delivery at 4,0 bar
39 l/min
Delivery at 4,5 bar
30 l/min
Delivery at 5,0 bar
21 l/min
Delivery at 5,5 bar
7 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
474 l/h
Consumption at water 15°C, flow 0°C
230 l/h
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water inlet temperature between 5°C and 20°C, and if it drops below the condensation point the cooling water pipes inside the unit must be insulated.
Defines the permissible cooling water supply conditions and mandatory insulation to prevent condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power ratings are valid only at 20°C room temperature, 15°C cooling water inlet and 1.5 bar differential pressure, so actual capacity will differ under other conditions.
Prevents overestimating cooling capacity when site conditions deviate from the reference point.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1,5 bar differential pressure between cooling water inlet and -outlet.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility to avoid corrosion or damage.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.
Defines acceptable power supply quality for reliable operation and warranty compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices with current consumption less than 63A are delivered with cable but without plug, so the customer must provide a suitable plug.
Affects electrical installation planning and procurement of the mating plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The unit uses refrigerant R-449A (A1, H280) with GWP 1397 and 1.35 kg charge, giving 1.9 t CO2 equivalent, which matters for refrigerant handling and leak reporting.
Needed for refrigerant safety classification, environmental reporting and service planning.
Source
“R-449A (A1, H280)”
Installation
The chiller is water-cooled and requires a minimum cooling water differential pressure of 1.5 bar and maximum cooling water pressure of 6 bar.
Ensures the site cooling water supply can meet the hydraulic requirements.
Source
“min. cooling water differential pressure”
Operation
The unit is IP20 and must be operated within an ambient temperature range of 5°C to 40°C.
Defines the environmental limits for installation location and enclosure protection.
Source
“min. ambient temperature”
Installation
The unit is intended for externally closed applications, which constrains how the hydraulic circuit may be connected.
Prevents misapplication in open-loop or unsuitable external circuits.
Source
“For externally closed applications.”
Maintenance
The expansion tank is atmospheric open and the unit has a digital level indicator, so the filling capacity and expansion volume must be respected.
Affects fluid filling, expansion volume planning and maintenance checks.
Source
“Housing, atmospheric open expansion tank and copper soldered evaporator made of stainless steel.”
Performance
Cooling water consumption depends strongly on flow temperature, ranging from 474 l/h at 15°C flow to 264 l/h at -10°C flow.
Needed to size cooling water supply and estimate water consumption.
Source
“Consumption at water 15°C, flow 15°C”
Compliance
The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1, which has implications for the installation environment.
Class A equipment may require additional EMC precautions in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”