Product introduction and main features of the Unichiller 200Tw
Specpp. 1-2
Technical data, performance figures and operating parameters per DIN 12876
p. 2table, 13 rows
p. 1table, 35 rows
Technical data according to DIN 12876
Parameter
Value
max. cooling water pressure
6 bar
min. filling capacity
15 l
Volume of expansion
22 l
Overall dimensions WxDxH **
760x800x1615 mm
Net weight
358 kg
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
14 A
Fuse (3 phase)
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S353269
** Please respect space requirements. See operating conditions at www.huber-online.com
Unichiller 200Tw — Technical data according to DIN 12876
Parameter
Value
Order-No.
3028.0112.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 15°C
20 kW
Cooling power at 0°C
11 kW
Cooling power at -10°C
5 kW
Cooling power at -20°C
3 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
2 kg
CO2 equivalent
2,8 t
Circulation pump
D3
max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 1,1 bar
202 l/min
Delivery at 1,5 bar
193 l/min
Delivery at 2,0 bar
179 l/min
Delivery at 2,5 bar
159 l/min
Delivery at 3,0 bar
140 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
77 l/min
Delivery at 4,5 bar
31 l/min
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 15°C
725 l/h
Consumption at water 15°C, flow 0°C
860 l/h
Consumption at water 15°C, flow -10°C
610 l/h
Consumption at water 15°C, flow -20°C
445 l/h
min. cooling water differential pressure
1 bar
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a minimum cooling water differential pressure of 1 bar to operate.
Ensures the facility's cooling water supply can provide adequate flow and pressure for heat rejection.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water inlet temperature must be between 5°C and 20°C; below the condensation point the cooling water pipes inside the unit must be insulated.
Prevents condensation and corrosion damage and defines the acceptable site cooling water conditions.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling power ratings are valid only at 20°C room temperature and 15°C cooling water inlet with 1 bar differential pressure.
Allows the engineer to compare quoted cooling capacity against actual site conditions and anticipate derating.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and - outlet.”
Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse and draws a maximum current of 14 A.
Determines electrical supply, protection device sizing and cabling for the installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide the final connection.
Affects electrical installation planning and who supplies the plug/connection hardware.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance
Voltage and frequency tolerances are ±5% and ±2% respectively, but certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.
Ensures the site power quality is within the unit's permissible operating envelope.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 2 kg (2.8 t CO2 equivalent).
Informs refrigerant safety classification, leak reporting obligations and handling requirements.
Source
“R-449A (A1, H280)”
Operation
The unit is rated IP20 and must be operated between 5°C and 40°C ambient temperature.
Defines the required indoor environment and protection against dust/water ingress at the installation site.
Source
“Degree of Protection IP20”
Installation
The unit has overall dimensions of 760x800x1615 mm and a net weight of 358 kg, so space and floor loading must be planned.
Needed for room layout, access routes and structural floor loading assessment.
Source
“Overall dimensions WxDxH ** 760x800x1615 mm”
Installation
The maximum permissible cooling water pressure is 6 bar.
Protects the cooling circuit from overpressure damage; site supply must be limited accordingly.
Source
“max. cooling water pressure 6 bar”
Operation
The pump delivers 210 l/min maximum at 4.7 bar, with flow dropping to 31 l/min at 4.5 bar.
Allows verification that the external circuit's pressure drop and required flow are within the pump curve.
Source
“max. delivery 210 l/min”
Maintenance
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality/treatment selection to avoid corrosion and fouling of the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”