Unichiller 210T Datasheet

Contents

  1. Overview p. 1
    Product introduction and main features of the Unichiller 210T
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 34 rows

Technical data according to DIN 12876 — Unichiller 210T (Order-No.: 3020.0029.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C21 kW
Cooling power at 0°C21 kW
Cooling power at -10°C13,5 kW
Cooling power at -20°C7,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:D3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **905x2172x1900 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 40 °C, so the process must stay within these limits.

Defines whether the unit can cover the required process temperatures.

Source
“Operating temperature range -20...40 °C”
Performance

Cooling capacity drops from 21 kW at 0 °C to 7.5 kW at -20 °C, so low-temperature duty needs a larger unit or lower load.

Cooling capacity at the actual operating temperature determines feasibility and sizing.

Source
“at -20°C 7,5 kW”
Installation

The unit requires a 400 V three-phase 50 Hz supply, so site power must be three-phase.

Electrical supply compatibility must be confirmed before ordering and installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the site plug.

Affects electrical installation scope and what must be procured on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

Environmental protection class limits where the chiller can be installed.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 and 40 °C, and cooling capacity may fall if ambient rises.

Site ambient conditions directly affect available cooling performance.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit is air-cooled and uses R-449A refrigerant with GWP 1397, so ventilation and refrigerant reporting may be required.

Refrigerant type and GWP affect room ventilation, safety and environmental compliance.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Installation

The chiller has overall dimensions 905 x 2172 x 1900 mm, so access routes and floor space must accommodate it.

Physical size determines rigging, doorway clearance and installed footprint.

Source
“Overall dimensions WxDxH ** 905x2172x1900 mm”
Installation

Space requirements must be respected, so clearance around the air-cooled unit must be planned.

Air-cooled chillers need adequate clearance for heat rejection and service access.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Installation

The pump connection is G1 1/4 male, so mating hoses and fittings must match this thread.

Ensures correct hydraulic connection hardware is ordered.

Source
“Pump connection G1 1/4 male”
Operation

Minimum filling capacity is 14 l and expansion volume is 39 l, so the external loop volume must be considered.

Loop volume affects fill quantity, expansion behaviour and heat transfer fluid selection.

Source
“min. filling capacity 14 l”
Performance

Pump delivery falls from 210 l/min at zero head to 31 l/min at 4.5 bar, so system pressure drop must be checked against the pump curve.

Available flow at the actual system pressure drop determines whether the process loop can be supplied.

Source
“max. delivery 210 l/min”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and some combinations are not allowed.

Site power quality must be verified to avoid invalid supply combinations and equipment stress.

Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance

The equipment is EMC Class A Group 1, so it may cause radio interference in residential areas.

EMC class affects installation location and potential interference with sensitive equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”