Unichiller 210Tw-H Recirculating Chiller Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller 210Tw-H chiller and its main features
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 10 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Volume of expansion22 l
Overall dimensions WxDxH **760x800x1615 mm
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)24 A
Fuse (3 phase)3x25A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S455148

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 210Tw-H (Order-No.: 3020.0047.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C21 kW
Cooling power at 0°C21 kW
Cooling power at -10°C15,5 kW
Cooling power at -20°C9,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity2,7 kg
CO2 equivalent3,8 t
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity15 l

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 3x25 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and breaker requirements before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

The chiller is water-cooled and needs a cooling water supply with at least 1 bar differential pressure and no more than 6 bar.

Cooling water pressure must be verified on site to ensure adequate heat rejection and avoid overpressure.

Source
“min. cooling water differential pressure 1 bar”
Installation

Cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if it drops below the condensation point.

Out-of-range cooling water temperature reduces cooling capacity and can cause condensation and corrosion.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops from 21 kW at 0 °C to 9.5 kW at -20 °C, so the cooling capacity must be checked at the actual operating temperature.

Selecting the chiller based only on nominal capacity can lead to insufficient cooling at low temperatures.

Source
“at -20°C 9,5 kW”
Performance

The circulation pump delivers 210 l/min at zero head but only 31 l/min at 4.5 bar, so system pressure drop must be matched to the pump curve.

Ensures the pump can overcome the external circuit pressure drop at the required flow rate.

Source
“max. delivery 210 l/min”
Installation

The unit needs a minimum filling capacity of 15 l and has a 22 l expansion volume, which must be considered when sizing the external loop.

Prevents underfilling and ensures the expansion tank can accommodate thermal volume changes.

Source
“min. filling capacity 15 l”
Safety

The refrigerant R-449A is classified A1/H280 with GWP 1397 and a CO2 equivalent of 3.8 t, which affects leak detection and environmental compliance.

Requires appropriate refrigerant handling, leak monitoring and F-gas reporting obligations.

Source
“R-449A (A1, H280)”
Installation

The unit is IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C.

IP20 offers no water protection, and ambient limits affect reliable operation and cooling performance.

Source
“Degree of Protection IP20”
Installation

Three-phase devices below 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.

Avoids delays at commissioning because the customer must supply and fit the plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance

Voltage and frequency tolerances are ±5% and ±2% respectively, but certain combinations such as -5% voltage with +2% frequency are not allowed.

Ensures the site power quality is within the permitted envelope before connecting the chiller.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

The unit is classified as pressure equipment category I and uses an atmospheric open expansion tank, so it is intended for externally closed applications.

Prevents applying the chiller to open or unsuitable circuits that could cause overflow or contamination.

Source
“For externally closed applications.”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to avoid corrosion and fouling of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”