Unichiller 210Tw Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller 210Tw chiller model
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

Technical data according to DIN 12876 — Unichiller 210Tw (Order-No.: 3020.0046.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 15°C21 kW
Cooling power at 0°C21 kW
Cooling power at -10°C15,5 kW
Cooling power at -20°C9,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity15 l
Volume of expansion22 l
Overall dimensions WxDxH **760x800x1615 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI

** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and must not exceed 6 bar, so the site cooling water supply must be verified against these limits.

Insufficient differential pressure will prevent proper cooling; excessive pressure may damage the cooling circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is water-cooled and requires a cooling water connection of G3/4 male, so the site must provide a matching connection.

Mismatched connections require adapters and can delay commissioning.

Source
“Cooling water connection G3/4 male”
Operation

The cooling water inlet temperature must be between 5°C and 20°C; below the condensation point the cooling water pipes in the unit must be insulated.

Operating outside this range reduces cooling capacity or causes condensation and corrosion risk.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power ratings are valid only at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure, so actual performance may differ on site.

Selection must account for site conditions; higher cooling water temperature reduces capacity and increases flow demand.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and - outlet.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or foul the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations (e.g., -5% voltage with +2% frequency) are not allowed.

Site power quality must be checked to avoid invalid supply conditions and equipment damage.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices with current consumption less than 63A are supplied with cable but without plug, so a suitable plug must be provided on site.

Electrical installation planning must include a compatible plug and connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The refrigerant is R-449A with GHS classification H280 (pressurized gas), so handling and ventilation precautions apply.

Personnel must be aware of pressurized gas hazards during installation and maintenance.

Source
“R-449A (A1, H280)”
Installation

The minimum filling capacity is 15 l and expansion volume is 22 l, so the external circuit volume and expansion vessel sizing must be considered.

Undersized or oversized external circuits affect temperature stability and pump performance.

Source
“min. filling capacity 15 l”
Operation

The unit has IP20 protection and must operate between 5°C and 40°C ambient temperature, so it must be installed in a dry, temperature-controlled environment.

IP20 offers no water protection; ambient limits affect reliability and warranty.

Source
“Degree of Protection IP20”
Installation

Space requirements must be respected as indicated in the operating conditions, so clearance around the unit must be planned.

Insufficient clearance can impede maintenance and cooling air flow.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Compliance

The equipment is classified as EN55011 Class A, Group 1 for electromagnetic disturbance, so it is intended for industrial environments.

Class A equipment may cause interference in residential areas and requires appropriate EMC planning.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”