Unichiller 350T-H Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the chiller
  2. Spec pp. 1-2
    Technical data according to DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/19

Technical data according to DIN 12876 — Unichiller 350T-H (Order-No.: 3021.0008.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C35 kW
Cooling power at 0°C23 kW
Cooling power at -10°C14 kW
Cooling power at -20°C8 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:D3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **905x2172x1900 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Selection

The chiller's cooling capacity drops as ambient temperature rises, so selection must account for the actual room temperature at the installation site.

Cooling capacity is rated at 20°C ambient; higher ambient reduces available cooling and may require a larger unit.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so the installer must provide a suitable plug or hard-wired connection.

Determines what electrical termination hardware must be procured and planned for before commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances must be met simultaneously, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Ensures the site power quality is compatible with the compressor and electronics before installation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit requires respect for space requirements and operating conditions, which must be checked before siting.

Air-cooled units need adequate clearance and ambient conditions; ignoring this can derate or damage the chiller.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Operation

The chiller is intended only for externally closed applications, so the process loop must be closed.

Using it on an open loop would be outside the intended design and could cause overflow or contamination.

Source
“For externally closed applications.”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.

Informs safety planning for refrigerant leak scenarios and service access.

Source
“R-449A (A1, H280)”
Compliance

The equipment emits Class A, Group 1 disturbances per EN55011, so it is intended for industrial electromagnetic environments.

Class A equipment may not be suitable for residential or light-commercial EMC environments without mitigation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.

Proper setting of the safety limiter protects the process and the chiller from overheating.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation

The pump connection is G1 1/4 male, and a matching hose coupling is included, so process piping must match this thread.

Determines the fittings and hoses needed for connecting the process loop.

Source
“Pump connection G1 1/4 male”
Selection

The minimum filling capacity is 14 l and expansion volume is 39 l, which sets the minimum heat-transfer fluid charge and expansion allowance.

Ensures the external loop and expansion vessel are sized so the pump and level monitoring work correctly.

Source
“min. filling capacity 14 l”
Operation

The unit is rated for ambient temperatures from 5 °C to 40 °C, so the installation room must stay within this range.

Operating outside the ambient range can reduce cooling capacity or trip protections.

Source
“max. ambient temperature 40 °C”
Selection

The chiller uses an air-cooled refrigerating machine that is CFC- and HCFC-free, which matters for environmental compliance and service.

Confirms refrigerant type and cooling method for site planning and environmental requirements.

Source
“air-cooled, CFC- and HCFC-free”