Product name and general description of the Unichiller 500Tw
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 34 rows
Technical data according to DIN 12876 — Unichiller 500Tw (Order-No.: 3030.0011.01)
Characteristic
Value
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 15°C
50 kW
Cooling power at 0°C
30 kW
Cooling power at -10°C
24 kW
Cooling power at -20°C
14 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump
D3
max. delivery
234 l/min
max. delivery pressure
4,9 bar
Delivery at 1,3 bar
197 l/min
Delivery at 1,5 bar
192 l/min
Delivery at 2,0 bar
178 l/min
Delivery at 2,5 bar
158 l/min
Delivery at 3,0 bar
137 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
86 l/min
Delivery at 4,5 bar
40 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1 1/4 male
min. cooling water differential pressure
1,5 bar
max. cooling water pressure
6 bar
min. filling capacity
12,7 l
Volume of expansion
7 l
Overall dimensions WxDxH **
1000x1100x1635 mm
Net weight
634 kg
Power supply (3 Phase)
400V 3~ 50Hz
** см. сноску в оригинале (Overall dimensions WxDxH **)
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller requires cooling water with a differential pressure of at least 1.5 bar and no more than 6 bar, so the site cooling water supply must be verified before installation.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.
Source
“min. cooling water differential pressure 1,5 bar”
Installation
The unit is water-cooled and must be connected to a cooling water circuit; the cooling water inlet temperature must be between 5°C and 20°C.
Cooling water temperature outside this range will reduce cooling capacity or cause condensation issues.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature falls below the condensation point, the cooling water pipes inside the unit must be insulated.
Without insulation, condensation can cause corrosion or electrical problems.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
The published cooling capacity is valid only at 15°C cooling water inlet; actual performance will be lower at warmer cooling water.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Incompatible water chemistry can corrode or foul these materials and impair heat transfer.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.
The site power supply must be checked against these limits to avoid invalid operating conditions.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices with current consumption less than 63A are supplied with a cable but without a plug, so a suitable plug must be provided on site.
The installer must plan for a matching plug and connection to the site power supply.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The refrigerant is R-449A with GHS classification A1, H280, and a global warming potential of 1397.
This affects refrigerant handling, leak detection, and environmental compliance requirements.
Source
“R-449A (A1, H280)”
Operation
The unit is rated IP20 and must be operated within an ambient temperature range of 5°C to 40°C.
Ambient conditions outside this range may cause overheating or control malfunction.
Source
“max. ambient temperature 40 °C”
Installation
The unit requires a minimum filling capacity of 12.7 liters and has a 7-liter expansion volume.
The external circuit and expansion tank must be sized to accommodate this volume.
Source
“min. filling capacity 12,7 l”
Selection
The circulation pump delivers a maximum of 234 l/min at 4.9 bar, with flow dropping to 40 l/min at 4.5 bar.
The pump curve must be checked against the pressure drop of the external circuit to ensure adequate flow.
Source
“max. delivery 234 l/min”
Installation
The pump and cooling water connections are G1 1/4 male, and hose couplings for both are included.
Matching hoses and fittings must be prepared for connection to the external circuit and cooling water supply.
Source
“Pump connection G1 1/4 male”
Compliance
The equipment is classified for electromagnetic disturbance as Class A, Group 1 according to EN55011.
Class A equipment may require additional EMC precautions in residential or light-industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”