The chiller requires a three-phase 400 V supply with a 63 A fuse per phase, so the site electrical infrastructure must be sized accordingly.
Determines upstream breaker/fuse rating and cable sizing for installation.
Source
“Fuse (3 phase) 3x63A”
| Parameter | Value |
|---|---|
| Net weight | 851 kg |
| Power supply (3 Phase) | 400V 3~ 50Hz |
| max. current (3 Phase) | 50 A |
| Fuse (3 phase) | 3x63A |
| Pressure equipment category | I |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | S433261 |
| Parameter | Value |
|---|---|
| Operating temperature range | -20...100 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Temperature stability at -10°C | 0,2 K |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Safety classification | III / FL |
| Heating power | 2 kW |
| Cooling power at 40°C | 80 kW |
| Cooling power at 15°C | 80 kW |
| Cooling power at 0°C | 60 kW |
| Cooling power at -10°C | 40 kW |
| Cooling power at -20°C | 20 kW |
| Refrigeration machine | water-cooled, CFC- and HCFC-free |
| Refrigerant (ASHRAE, GHS) | R-449A (A1, H280) |
| Global Warming Potential (GWP) | 1397 |
| Refrigerant quantity | 5 kg |
| CO2 equivalent | 7,0 t |
| Circulation pump | D3 |
| max. delivery | 196 l/min |
| max. delivery pressure | 5 bar |
| Delivery at 1,0 bar | 175 l/min |
| Delivery at 2,0 bar | 150 l/min |
| Delivery at 3,0 bar | 122 l/min |
| Delivery at 4,0 bar | 74 l/min |
| Pump connection | G1 1/4 male |
| Cooling water connection | G1 1/4 male |
| Consumption at water 15°C, flow 15°C | 3480 l/h |
| min. cooling water differential pressure | 2,5 bar |
| max. cooling water pressure | 6 bar |
| min. filling capacity | 12,7 l |
| Volume of expansion | 7 l |
| Overall dimensions WxDxH ** | 1000x1600x1620 mm |
** Overall dimensions WxDxH
The chiller requires a three-phase 400 V supply with a 63 A fuse per phase, so the site electrical infrastructure must be sized accordingly.
Determines upstream breaker/fuse rating and cable sizing for installation.
“Fuse (3 phase) 3x63A”
The unit is water-cooled and needs a cooling water differential pressure of at least 2.5 bar and no more than 6 bar at the connections.
Cooling water supply pressure must be verified before installation to avoid insufficient cooling or overpressure damage.
“min. cooling water differential pressure 2,5 bar max. cooling water pressure 6 bar”
Cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.
Defines the acceptable cooling water supply conditions and condensation protection measures.
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Cooling capacity is rated at 80 kW at 40 °C and drops to 20 kW at -20 °C, so capacity must be checked at the actual operating temperature.
Prevents oversizing or undersizing the chiller for the required process temperature.
“at 40°C 80 kW at 15°C 80 kW at 0°C 60 kW at -10°C 40 kW at -20°C 20 kW”
The circulation pump delivers 196 l/min maximum and 74 l/min at 4.0 bar, so system pressure drop must be matched to the pump curve.
Ensures the pump can overcome the external circuit pressure drop at the required flow.
“max. delivery 196 l/min max. delivery pressure 5 bar Delivery at 1,0 bar 175 l/min Delivery at 2,0 bar 150 l/min Delivery at 3,0 bar 122 l/min Delivery at 4,0 bar 74 l/min”
The unit has IP20 protection and must be installed in an ambient temperature between 5 °C and 40 °C.
Restricts installation location to indoor, dry environments within the specified ambient range.
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
The refrigerant R-449A has a GHS hazard statement H280, so refrigerant handling and ventilation precautions are required.
Informs site safety planning for refrigerant containment and emergency procedures.
“R-449A (A1, H280)”
The equipment emits Class A, Group 1 disturbances under EN55011, so it may require EMC mitigation in residential or sensitive environments.
Class A equipment is not intended for residential use without additional EMC measures.
“Classification (disturbance) to EN55011: Class A, Group 1”
Leakage current exceeds 3.5 mA, so protective earth connection and residual current device selection must account for this.
Affects earth bonding and RCD type/rating selection during electrical installation.
“Attention: leakage current > 3,5mA”
Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.
Prevents delays at commissioning due to missing power connector.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Defines acceptable supply quality for reliable operation and warranty compliance.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and treatment to avoid corrosion or scaling in the cooling circuit.
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
The minimum filling capacity is 12.7 l and expansion volume is 7 l, so the external circuit volume must be considered for proper expansion tank sizing.
Ensures the expansion vessel can accommodate thermal expansion of the filled system.
“min. filling capacity 12,7 l Volume of expansion 7 l”