** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller P007-H (Order-No.: 3012.0300.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Cooling power at ambient temperature +20°C
at 15°C
0,7 kW
at 0°C
0,55 kW
at -10°C
0,4 kW
at -20°C
0,2 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,14 kg
CO2 equivalent
0,0 t
Circulation pump:
B
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
21 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 1,5 bar
11 l/min
Delivery at 2,0 bar
6 l/min
Pump connection
G3/4 male
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Строка «Cooling power at ambient temperature +20°C» напечатана с точкой-заполнителем без значения; далее следуют подстроки при 15°C, 0°C, -10°C, -20°C.
Engineering Facts
14 facts · every claim linked to a page
Selection
The chiller is rated for an operating temperature range of -20 to 100 °C, so it must be matched to the process temperature window.
Defines whether the unit can cover the required process temperatures before selection.
Source
“Operating temperature range
-20...100 °C”
Selection
Cooling capacity drops with lower temperature and with rising ambient temperature, so sizing must use the capacity at the actual operating point.
Prevents undersizing when the chiller runs in a warm room or at low temperatures.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects installation and safety planning.
Flammable refrigerant classification drives ventilation, ignition-source and safety requirements.
Source
“Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)”
Installation
The unit requires a 208-240 V single/two-phase 50/60 Hz supply with a maximum current of 14 A and a 16 A fuse.
Determines the electrical supply, breaker and cabling needed at the installation site.
Source
“Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
14 A
Fuse
16 A”
Installation
Voltage and frequency tolerances are coupled: ±5% voltage only with ±2% frequency, and certain combinations are not allowed.
Prevents connecting the chiller to a supply whose combined voltage/frequency deviation is outside the permitted envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The chiller must be installed where ambient temperature stays between 5 °C and 40 °C.
Ambient limits define the acceptable room conditions and affect cooling performance.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The pump connection is G3/4 male, so hoses and couplings must match this thread.
Ensures correct hose coupling selection and leak-free connection at installation.
Source
“Pump connection
G3/4 male”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump ordered.
Prevents chemical attack and pump failure when choosing the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Guides correct heat transfer fluid preparation for sub-zero operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
Avoids a missing or wrong plug on delivery and ensures site power compatibility.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Compliance
The equipment is EMC classified as Class A, Group 1 to EN55011, which has implications for the electromagnetic environment.
Class A equipment may require EMC considerations in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The unit has a level indicator with sight glass, allowing the fill level to be checked visually.
Supports routine monitoring of heat transfer fluid level during operation.
Source
“Level indicator with sight glass.”
Safety
The unit provides adjustable overtemperature protection according to DIN 12876.
Confirms a safety cut-out is present and must be set correctly for the application.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance
The circulation pump delivers up to 25 l/min and 2.5 bar, with flow falling as pressure rises.
Allows checking whether the pump can overcome the pressure drop of the external circuit.