Unichiller P022-H Chiller Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller P022-H chiller model and its main features
  2. Spec pp. 1-2
    Performance, temperature, cooling capacity and electrical data per DIN 12876
  3. p. 2 table, 11 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Overall dimensions WxDxH **460x590x743 mm
Net weight83 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current15 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S546853

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-10...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Resolution of display0,1 K
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C
at 15°C2,2 kW
at 0°C1,6 kW
at -10°C1 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,58 kg
CO2 equivalent0,8 t
UN-numberUN 2857
Circulation pump:B
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l

Order-No.: 3010.0105.01; Unichiller P022-H

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -10 to 100 °C, so the process must stay within these limits.

Defines the application window and excludes processes requiring lower or higher temperatures.

Source
“Operating temperature range -10...100 °C”
Selection

Cooling capacity drops with lower evaporating temperature, from 2.2 kW at 15 °C to 1 kW at -10 °C, so size the unit for the coldest duty point.

Prevents undersizing when the process requires low-temperature cooling.

Source
“at 15°C 2,2 kW at 0°C 1,6 kW at -10°C 1 kW”
Installation

The unit requires a 220-240 V single-phase 50 Hz supply with up to 15 A current and a 16 A fuse.

Determines electrical supply, breaker rating and cabling for the installation.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 15 A Fuse 16 A”
Installation

The chiller is air-cooled and must be installed where ambient temperature stays between 5 °C and 40 °C.

Ambient conditions affect cooling performance and reliable operation of the air-cooled refrigeration machine.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is declared at 20 °C room temperature and may fall if ambient temperature rises.

Installation in a hot room can reduce available cooling capacity below the datasheet values.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The pump connection is G3/4 male, so hoses and fittings must match this thread.

Ensures correct plumbing connection and avoids adapters or leaks.

Source
“Pump connection G3/4 male”
Operation

The pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with flow falling as pressure rises.

Needed to check whether the pump can overcome the pressure drop of the external circuit.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”
Operation

The minimum filling capacity is 3.8 l and expansion volume is 1.7 l, so the loop must be filled and expansion accommodated accordingly.

Prevents insufficient filling and overpressure from thermal expansion.

Source
“min. filling capacity 3,8 l Volume of expansion 1,7 l”
Safety

The refrigerant is R-449A with GHS hazard H280, UN 2857, and GWP 1397, so refrigerant handling and leak response must follow safety rules.

Affects safety classification, ventilation requirements and environmental reporting.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Safety

The plastic pump is not resistant to acetone and polyglycols, so these media must not be used without a more resistant pump.

Prevents chemical attack and pump failure when selecting heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used alone.

Avoids freezing and damage when operating near or below 0 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Compliance

The unit is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which matters for the electrical environment.

Class A equipment may require EMC precautions in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but not all combinations are allowed.

Ensures the supply quality is within the allowed envelope before commissioning.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The degree of protection is IP20, so the unit must be protected from water and dust ingress.

Restricts installation location to dry, clean indoor areas.

Source
“Degree of Protection IP20”