The chiller is intended only for externally closed applications, so an open-loop or vented process circuit is not permitted.
Determines whether the unit can be applied to the process loop at all.
Source
“For externally closed applications.”
| Parameter | Value |
|---|---|
| min. filling capacity | 18 l |
| Volume of expansion | 48 l |
| Overall dimensions WxDxH ** | 740x1160x1165 mm |
| Net weight | 300 kg |
| sound pressure level +/- 4 dB(A) | 62 dB(A) |
| Power supply (3 Phase) | 400V 3~ 50Hz |
| max. current (3 Phase) | 12 A |
| Fuse (3 phase) | 3x16A |
| Pressure equipment category | Art. 4.3 PED |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | S315756 |
** Please respect space requirements. See operating conditions at www.huber-online.com
| Parameter | Unichiller P050-H |
|---|---|
| Order-No. | 3038.0005.01 |
| Operating temperature range | -20...100 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Temperature stability at -10°C | 0,2 K |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Safety classification | III / FL |
| Heating power | 2 kW |
| Cooling power at ambient temperature +20°C | |
| at 100°C | 5 kW |
| at 20°C | 5 kW |
| at 0°C | 3,4 kW |
| at -10°C | 2,1 kW |
| at -20°C | 0,7 kW |
| Refrigeration machine | air-cooled, CFC- and HCFC-free |
| Refrigerant (ASHRAE, GHS) | R-449A (A1, H280) |
| Global Warming Potential (GWP) | 1397 |
| Refrigerant quantity | 1,03 kg |
| CO2 equivalent | 1,4 t |
| Circulation pump: | E2 |
| max. delivery | 130 l/min |
| max. delivery pressure | 5,7 bar |
| Delivery at 2,4 bar | 116 l/min |
| Delivery at 2,5 bar | 115 l/min |
| Delivery at 3,0 bar | 108 l/min |
| Delivery at 3,5 bar | 100 l/min |
| Delivery at 4,0 bar | 92 l/min |
| Delivery at 4,5 bar | 80 l/min |
| Delivery at 5,0 bar | 59 l/min |
| Delivery at 5,5 bar | 20 l/min |
| Pump connection | G1 1/4 male |
The chiller is intended only for externally closed applications, so an open-loop or vented process circuit is not permitted.
Determines whether the unit can be applied to the process loop at all.
“For externally closed applications.”
Cooling capacity falls sharply at low temperature, from 5 kW at 20 °C to 0.7 kW at -20 °C, so the load must be checked at the actual set point.
Sizing on nominal capacity would undersize the unit at low-temperature duty.
“at -20°C 0,7 kW”
Cooling capacity is rated at 20 °C room temperature and drops if ambient rises, so the installed ambient must be considered.
Prevents overestimating available cooling in a warm plant room.
“If the ambient temperature rises, the cooling capacity may drop.”
The unit requires a 400 V three-phase supply with 12 A max current and 3x16 A fusing.
Defines the electrical supply and protection needed at the installation site.
“max. current (3 Phase) 12 A”
Three-phase devices below 63 A are delivered with cable but without a plug, so a plug must be sourced separately.
Affects electrical installation planning and bill of materials.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Voltage and frequency tolerances are linked: -5% voltage with +2% frequency is not allowed.
Site power quality must be checked against the combined tolerance rule.
“-5% voltage and +2% frequency -> not allowed!”
The unit needs 740x1160x1165 mm of space and weighs 300 kg, so floor loading and access must be verified.
Determines rigging, floor loading and room layout.
“740x1160x1165 mm”
Ambient temperature must stay between 5 °C and 40 °C for correct operation.
Sets the allowable plant-room conditions for installation.
“max. ambient temperature 40 °C”
The refrigerant R-449A is classified A1 but carries GHS hazard H280 (pressurised gas), with 1.4 t CO2 equivalent.
Affects refrigerant handling, labelling and leak-response procedures.
“R-449A (A1, H280)”
The unit has adjustable overtemperature protection per DIN 12876 and safety classification III/FL.
Confirms the safety cut-out required for unattended operation.
“With adjustable overtemperature protection according to DIN 12876.”
The pump delivers 130 l/min max at 5.7 bar, but flow collapses to 20 l/min at 5.5 bar, so system pressure drop must be matched.
High-pressure-drop circuits may starve the pump despite the high nominal head.
“max. delivery pressure 5,7 bar”
The expansion tank is atmospheric and open, with 48 l expansion volume and 18 l minimum filling.
Sets the minimum fill and expansion allowance for the loop.
“Volume of expansion 48 l”
The unit is EMC Class A Group 1, which is industrial-grade and may not suit residential environments.
Determines EMC suitability of the installation location.
“Classification (disturbance) to EN55011: Class A, Group 1”
The pressure equipment falls under PED Article 4.3, so no CE pressure marking applies but documentation is still required.
Affects compliance documentation and acceptance at handover.
“Pressure equipment category Art. 4.3 PED”