Installation
The chiller requires a three-phase 400 V supply with a 3x10 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and protection device rating before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
The unit is water-cooled and needs a cooling water supply with at least 0.5 bar differential pressure and no more than 6 bar.
Cooling water pressure must be verified on site to ensure sufficient flow and avoid overpressure damage.
Source
“min. cooling water differential pressure 0,5 bar”
Installation
Cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.
Defines the acceptable cooling water source temperature and the need for pipe insulation to prevent condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Performance must be derated if the available cooling water is warmer than the rated 15 °C.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Prevents corrosion or damage by selecting suitable water quality and avoiding incompatible additives.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Ensures the site power quality is within the allowed envelope before commissioning.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and the bill of materials for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit is rated for ambient temperatures from 5 °C to 40 °C and IP20 protection, so it must be installed in a dry indoor environment.
Defines the permissible room conditions and excludes wet or dusty locations.
Source
“max. ambient temperature 40 °C”
Selection
The refrigerant is R-449A with GWP 1397 and a charge of 0.95 kg, giving a CO2 equivalent of 1.3 t.
Relevant for refrigerant regulations, leak checks and environmental compliance.
Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Selection
The pump delivers up to 130 l/min and 5.7 bar, but flow falls sharply at high pressure, so the system pressure drop must be checked.
Ensures the pump can overcome the external circuit pressure drop at the required flow.
Source
“max. delivery pressure 5,7 bar”
Installation
The unit needs a minimum filling capacity of 18 l and has a 48 l expansion volume, which affects the external loop volume and expansion tank sizing.
Determines the minimum heat transfer fluid volume and expansion volume for safe operation.
Source
“min. filling capacity 18 l”
Installation
The overall dimensions are 740x1160x1050 mm and the net weight is 301 kg, so space and floor loading must be planned.
Needed for site layout, access routes and structural floor loading.
Source
“Overall dimensions WxDxH ** 740x1160x1050 mm”
Safety
The safety classification is I / NFL according to DIN 12876, which defines the permitted application and safety requirements.
Confirms the unit is suitable for the intended temperature control application under DIN 12876.
Source
“Safety classification I / NFL”
Operation
The unit is designed for externally closed applications, so the external circuit must be closed.
Prevents misapplication with open baths or unsealed external circuits.
Source
“For externally closed applications.”