Unistat 410w Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat 410w unit
  2. Spec pp. 1-2
    Technical performance data and operating parameters per DIN 12876
  3. p. 2 table, 22 rows
  4. p. 1 table, 28 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 0,4 bar42 l/min
Delivery at 0,6 bar33 l/min
Delivery at 0,8 bar20 l/min
Pump connectionM24x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C280 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity1,5 l
Filling capacity expansion tank2,8 l
Overall dimensions WxDxH **426x360x631 mm
Net weight68 kg
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current16 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S577520

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat 410w (Order-No.: 1066.0013.01)

ParameterValue
Operating temperature range-45...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power3 kW
Cooling power with Thermooil at 250°C1,3 kW
Cooling power with Thermooil at 200°C1,5 kW
Cooling power with Thermooil at 100°C2,5 kW
Cooling power with Thermooil at 20°C2,5 kW
Cooling power with Ethanol at 0°C1,5 kW
Cooling power with Ethanol at -20°C0,8 kW
Cooling power with Ethanol at -40°C0,17 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity0,098 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump: max. delivery56 l/min
Circulation pump: max. delivery pressure0,9 bar
Circulation pump: Delivery at 0,2 bar50 l/min

Таблица транспонирована: характеристики в строках, значения в колонке Value. Cooling power сгруппирована по средам (Thermooil / Ethanol) и температурам, как напечатано.

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is water-cooled and requires a cooling water supply, so a suitable water circuit must be available at the installation site.

Determines site requirements and rules out air-cooled installation.

Source
“Refrigeration machine water-cooled, natural refrigerant”
Safety

The refrigerant is flammable (A3, H220), so the installation must respect flammable-refrigerant safety requirements.

Affects risk assessment, ventilation and compliance for the installation location.

Source
“R-1270 (A3, H220)”
Installation

The cooling water inlet must be between 5 °C and 20 °C, so the available cooling water temperature must be checked before installation.

Out-of-range cooling water will degrade cooling performance or prevent proper operation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water is colder than the condensation point, the cooling water pipes must be insulated.

Prevents condensation and moisture damage in the cooling circuit.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops as cooling water temperature rises, so performance must be derated for warmer site water.

Needed to correctly size the unit for the actual site cooling water conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit requires at least 1 bar differential pressure and must not exceed 6 bar.

Ensures adequate flow and protects the unit from overpressure.

Source
“min. cooling water differential pressure 1 bar”
Installation

The cooling water connection is G1/2 male, so matching hoses/fittings must be provided.

Prevents delays and leaks during commissioning.

Source
“Cooling water connection G1/2 male”
Installation

The pump connection is M24x1.5 male, so external hoses or adapters must match this thread.

Required for correct connection of external circuits.

Source
“Pump connection M24x1,5 male”
Installation

The unit requires a 220–240 V single/two-phase supply with a 16 A fuse, so the electrical supply must be sized accordingly.

Determines electrical installation and breaker sizing.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not be combined in the unfavorable direction, so supply quality must be verified.

Prevents operation outside the allowed supply envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

The unit is IP20 and must be installed in an environment between 5 °C and 40 °C ambient temperature.

Defines the acceptable operating environment and protection level.

Source
“max. ambient temperature 40 °C”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, so the fluid must be selected accordingly.

Ensures pump performance and prevents excessive load.

Source
“max. permissible kin. viscosity 50 mm²/s”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Prevents corrosion and premature failure of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

The equipment is EMC Class A Group 1, so it may cause radio interference in residential environments.

Affects installation location and EMC compliance planning.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”