  1. Overview p. 1
    Product introduction and general description of the Unistat 610w
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 17 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C400 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity6,5 l
Filling capacity expansion tank26 l
Overall dimensions WxDxH **630x704x1520 mm
Net weight360 kg
sound pressure level +/- 4 dB(A)62 dB(A)
Power supply requirement400V 3~ 50Hz
max. current15,5 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S395645

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterUnistat 610w
Order-No.1052.0005.01
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 200°C7 kW
Cooling power with Thermooil at 100°C7 kW
Cooling power with Ethanol at 0°C7 kW
Cooling power with Ethanol at -20°C6,4 kW
Cooling power with Ethanol at -40°C2,6 kW
Cooling power with Ethanol at -60°C0,05 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2,55 kg
CO2 equivalent5,5 t
Circulation pump:MK pump
max. delivery82 l/min
max. delivery pressure1,5 bar
Delivery at 0,25 bar75 l/min
Delivery at 0,5 bar67 l/min
Delivery at 0,75 bar59 l/min
Delivery at 1,0 bar47 l/min
Delivery at 1,25 bar29 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and breaker rating before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 15,5 A Fuse 3x16A”
Installation

Cooling water must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and the pipes must be insulated if condensation can occur.

Defines the cooling water utility requirements and condensation protection needed at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

The published cooling capacities are only valid at 20 °C room temperature and 15 °C cooling water inlet with 1 bar differential pressure, so actual capacity will differ at other conditions.

Prevents overestimating cooling performance when site conditions differ from the rating basis.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Performance

Cooling capacity drops sharply at low temperatures, from 2.6 kW at -40 °C to only 0.05 kW at -60 °C, which limits usable load near the lower end of the range.

Critical for selecting the unit when the process needs meaningful heat removal near -60 °C.

Source
“at -40°C 2,6 kW at -60°C 0,05 kW”
Installation

The pump connection is M30x1.5 male with a Y-shaped bore per DIN 3863, so mating hoses must use ball socket and sleeve nut per DIN 3863/3870.

Ensures correct fittings and hoses are ordered for leak-free connection.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The pump delivers 82 l/min at zero head but only 29 l/min at 1.25 bar, so external circuit pressure drop directly reduces available flow.

Needed to check whether the external loop resistance will still provide adequate flow.

Source
“max. delivery 82 l/min max. delivery pressure 1,5 bar Delivery at 1,25 bar 29 l/min”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which constrains fluid choice at low temperatures.

Prevents selecting a fluid that is too viscous at low temperature and would overload the pump.

Source
“max. permissible kin. viscosity 50 mm²/s”
Safety

The refrigerant R-452A has a GWP of 2141 and a charge of 2.55 kg giving 5.5 t CO2 equivalent, which matters for leak monitoring and F-gas obligations.

Affects regulatory compliance and leak-check requirements for the refrigeration circuit.

Source
“Global Warming Potential (GWP) 2141 Refrigerant quantity 2,55 kg CO2 equivalent 5,5 t”
Installation

The unit is IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in a wet or hot environment.

Sets the permissible location and environmental conditions for installation.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not combine in the same direction, e.g. -5% voltage with +2% frequency is not allowed.

Guides assessment of site power quality before connecting the unit.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.

Avoids delays at commissioning because the plug is not included.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Prevents corrosion or scaling from unsuitable cooling water that would damage the circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL, relevant for unattended operation.

Confirms built-in safety protection for the temperature control loop.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection

The unit is suitable for both externally closed and externally open applications, which broadens the possible process configurations.

Helps decide whether the unit fits an open bath or only closed-loop process.

Source
“As well as for externally closed and also externally open applications.”