Unistat GL 530w Technical Data Sheet

Contents

  1. Overview p. 1
    Introduces the Unistat GL 530w unit and its main features
  2. Spec pp. 1-2
    Performance, temperature range, and technical data per DIN 12876
  3. Spec p. 2
    Physical dimensions and connection details
  4. p. 2 table, 15 rows
  5. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow 0°C620 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-50...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 200°C12 kW
Cooling power with Thermooil at 100°C12 kW
Cooling power with Thermooil at 20°C12 kW
Cooling power with Ethanol at 0°C10 kW
Cooling power with Ethanol at -20°C7 kW
Cooling power with Ethanol at -40°C2,5 kW
Cooling power with Ethanol at -50°C1,5 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
Circulation pump: MK pump max. delivery196 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply, so the electrical installation must be planned for 3~ 50 Hz.

Determines the site power distribution and whether a three-phase feed is available.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar, so the site water system must be checked before installation.

Incorrect cooling water pressure can prevent rated cooling or damage the water circuit.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Installation

The unit is water-cooled and needs a cooling water connection of G3/4 male, so matching hoses and fittings must be provided.

Ensures the correct cooling water plumbing is available at the installation point.

Source
“Cooling water connection G3/4 male”
Installation

The cooling water intake temperature must be between 5 °C and 20 °C, and below-dew-point operation requires insulating the cooling water pipes.

Defines the allowable cooling water supply conditions and condensation protection measures.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The pump connection is M38x1.5 male with bore shape Y (60°) per DIN 3863, so tempering hoses must use compatible ball sockets and sleeve nuts.

Prevents mismatch of process connections and leakage at the pump outlet.

Source
“Pump connection M38x1,5 male”
Installation

The unit requires minimum 8.9 l filling capacity and has a 19.9 l expansion tank, so the external circuit volume and fluid expansion must be considered.

Avoids underfilling or overflow and ensures sufficient expansion volume for the application.

Source
“min. filling capacity 8,9 l Filling capacity expansion tank 19,9 l”
Installation

The overall dimensions are 730x804x1738 mm and space requirements must be respected, so the installation location must provide adequate clearance.

Ensures the unit fits the available floor space and access for service.

Source
“Overall dimensions WxDxH ** 730x804x1738 mm”
Operation

The ambient temperature must remain between 5 °C and 40 °C for proper operation.

Operating outside this range may affect performance, safety, and reliability.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, so the fluid must be selected to stay within this limit at operating temperature.

Too-viscous fluid can overload the pump and reduce circulation performance.

Source
“max. permissible kin. viscosity 100 mm²/s”
Performance

Cooling power with ethanol drops from 10 kW at 0 °C to 1.5 kW at -50 °C, so low-temperature applications must account for reduced capacity.

Critical for sizing the unit for low-temperature process loads.

Source
“Cooling power with Ethanol at 0°C 10 kW at -20°C 7 kW at -40°C 2,5 kW at -50°C 1,5 kW”
Performance

The circulation pump delivers 196 l/min maximum and 2.5 bar maximum pressure, with flow falling as pressure rises, so the process pressure drop must be matched to the pump curve.

Ensures the pump can meet the required flow at the actual system pressure drop.

Source
“max. delivery 196 l/min max. delivery pressure 2,5 bar”
Safety

The unit uses refrigerant R-744 (A1, H280) with GWP 1, so refrigerant safety and environmental handling requirements apply.

Affects refrigerant handling, safety classification, and environmental compliance.

Source
“Refrigerant (ASHRAE, GHS) R-744 (A1, H280) Global Warming Potential (GWP) 1”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Prevents installation on a supply that may exceed the allowed voltage/frequency combination.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% Example -5% voltage and +2% frequency -> not allowed!”
Compliance

The equipment is classified as EN55011 Class A, Group 1, so it may cause radio interference and may require EMC measures in residential or sensitive environments.

Important for EMC planning and site compatibility.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”