Product introduction and general description of the Unistat P615w unit
Specpp. 1-2
Technical performance data and operating specifications per DIN 12876
p. 2table, 21 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 2,0 bar
49 l/min
Delivery at 2,5 bar
36 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
426 l/h
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
min. filling capacity
5,65 l
Filling capacity expansion tank
26,5 l
Overall dimensions WxDxH **
630x704x1520 mm
Net weight
358 kg
sound pressure level +/- 4 dB(A)
62 dB(A)
Power supply requirement
400V 3~ 50Hz
max. current
25,5 A
Fuse
3x32A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S432760
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P615w — Technical data according to DIN 12876
Characteristic
Value
Order-No.
1074.0011.01
Operating temperature range
-60...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Cooling power with Thermooil at 200°C
9,5 kW
Cooling power with Thermooil at 100°C
9,5 kW
Cooling power with Thermooil at 20°C
9,5 kW
Cooling power with Ethanol at 0°C
9,5 kW
Cooling power with Ethanol at -20°C
8 kW
Cooling power with Ethanol at -40°C
4,6 kW
Cooling power with Ethanol at -60°C
1,2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
2,55 kg
CO2 equivalent
5,5 t
Circulation pump
MK pump
max. delivery
82 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
75 l/min
Delivery at 1,0 bar
68 l/min
Delivery at 1,5 bar
59 l/min
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 3x32 A fuse and draws up to 25.5 A, so the site electrical infrastructure must be sized accordingly.
Determines the electrical supply, breaker and cabling needed at the installation site.
Source
“Power supply requirement 400V 3~ 50Hz max. current 25,5 A Fuse 3x32A”
Installation
The cooling water circuit needs a differential pressure of at least 3 bar and must not exceed 6 bar, with a flow of about 426 l/h at 15 °C water.
Sizes the cooling water supply and protects the water-cooled condenser from under- or over-pressure.
Source
“Consumption at water 15°C, flow 0°C 426 l/h min. cooling water differential pressure 3 bar max. cooling water pressure 6 bar”
Installation
Cooling water intake must stay between 5 °C and 20 °C, and pipes must be insulated if water can drop below the condensation point.
Prevents condensation, corrosion and loss of cooling performance in the water circuit.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity falls sharply at low temperatures, from 9.5 kW at 0 °C with ethanol down to 1.2 kW at -60 °C, so low-temperature loads must be checked against the curve.
Ensures the selected unit can actually remove the process heat at the required low setpoint.
Source
“at -60°C 1,2 kW”
Performance
The circulation pump delivers 82 l/min at zero head but only 36 l/min at 2.5 bar, so high-pressure-drop loops must be evaluated against the pump curve.
Confirms whether the pump can overcome the pressure drop of the external application circuit.
The refrigerant R-452A is classified A1 with GHS hazard H280 and has a GWP of 2141 with 5.5 t CO2 equivalent, which affects refrigerant handling and environmental reporting.
Needed for refrigerant safety classification, leak response and CO2-equivalent reporting obligations.
Source
“R-452A (A1, H280)”
Safety
Leakage current exceeds 3.5 mA, so protective earth and RCD selection must account for this higher leakage.
Prevents nuisance tripping and ensures proper earthing and residual-current protection.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the site plug.
Avoids surprises at commissioning when no mains plug is present on the delivered unit.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are ±5 % and ±2 % respectively, but the combination of -5 % voltage with +2 % frequency is not allowed.
Guides assessment of site power quality before connecting the unit.
Source
“-5% voltage and +2% frequency -> not allowed!”
Operation
The unit is rated for ambient temperatures from 5 °C to 40 °C, so the installation room must stay within this range.
Out-of-range ambient conditions degrade performance and may void reliable operation.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation
The unit is IP20 and weighs 358 kg with dimensions 630x704x1520 mm, so it needs a dry indoor location and adequate floor space and access.
Determines floor loading, transport route and clearance for installation and service.
Source
“Overall dimensions WxDxH ** 630x704x1520 mm Net weight 358 kg”
Operation
The pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863, requiring matching ball-socket fittings and sleeve nuts per DIN 3870.
Ensures correct mating hoses and fittings are ordered for leak-free connection.
Source
“Pump connection M30x1,5 male”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Prevents corrosion and fouling of the mixed-material water circuit.
Source
“Please use suitable cooling water.”
Compliance
The unit is EMC class A group 1 per EN55011, which is intended for industrial environments and may cause interference in residential areas.
Affects EMC planning and suitability of the installation location.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”