Selection
The thermostat covers a working temperature range of 50 to 400 °C, so it must be matched to the process temperature window.
Defines whether the unit can handle the required process temperatures at all.
Source
“Operating temperature range
50...400 °C”
Installation
The unit requires a 200–240 V single- or two-phase supply with a maximum current of 14 A and a 16 A fuse.
Determines electrical supply, breaker sizing and cabling for the installation.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
14 A
Fuse
16 A”
Installation
The cooling water connection is a G1/2 male thread, which must be provided for in the site piping.
Needed to plan the cooling water supply and return connections before installation.
Source
“Cooling water connection
G1/2 male”
Installation
The pump connections are M24x1.5 on both flow and return, so matching hoses and fittings must be ordered.
Prevents mismatch of process hoses and adapters during commissioning.
Source
“Pump connenction flow
M24x1,5
Pump connection return
M24x1,5”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.
Restricts the acceptable installation environment and enclosure requirements.
Source
“Degree of Protection
IP20”
Operation
The ambient temperature at the installation site must stay between 5 °C and 40 °C for correct operation.
Room conditions outside this range can affect performance and reliability.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The controller is connected to the machine by a cable of approximately 5 m, which affects controller placement.
Determines where the Pilot ONE controller can be mounted relative to the unit.
Source
“Length of cable between machine and controller approx. 5m.”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
Confirms built-in safety protection relevant for risk assessment and compliance.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Compliance
External branch circuit protection according to UL 489 is required for this equipment.
Mandates a specific upstream protective device in the electrical installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10 % only if they do not run in opposite directions, so supply quality must be checked.
Prevents operation outside the allowed supply envelope that could damage the unit.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order.
Ensures the correct plug type is supplied for the destination country.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Compliance
The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.
Class A equipment may require EMC measures in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”