Unistat TR401w HT Temperature Control System Datasheet
Contents
Overviewp. 1
Introduces the Unistat TR401w HT unit and its main features
Specpp. 1-2
Performance data, dimensions, and operating parameters per DIN 12876
p. 2table, 6 rows
p. 1table, 35 rows
Technical data according to DIN 12876
Parameter
Value
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
14,5 A
Fuse
16 A
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S413492
Technical data according to DIN 12876 — Unistat TR401w HT (Order-No.: 1028.0018.01)
Parameter
Value
Operating temperature range
50...400 °C
min. operating temperature range w. water cooling
15 °C
Resolution of display
0,01 K
Temperature stability at 70°C
0,05 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
2-point-calibration
Internal temperature sensor
Pt100
External sensor
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
3 kW
Cooling power at 100°C
10 kW
Cooling power at 200°C
10 kW
Cooling power at 300°C
10 kW
Cooling power at 400°C
10 kW
Circulation pump: max. delivery
26 l/min
max. delivery pressure
0,8 bar
Delivery at 0,2 bar
23 l/min
Delivery at 0,4 bar
18 l/min
Delivery at 0,6 bar
12 l/min
Pump connenction flow
M24x1,5
Pump connection return
M24x1,5
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
240 l/h
min. filling capacity
2,3 l
Volume of expansion
5 l
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
Overall dimensions WxDxH
288x379x890 mm
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
Net weight
54 kg
Значения приведены дословно, включая опечатку «connenction» в оригинале.
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires cooling water with a differential pressure of at least 3 bar and no more than 6 bar, so the site water supply must be verified before installation.
Insufficient or excessive cooling water pressure will prevent proper cooling and may damage the unit.
Source
“min. cooling water differential pressure 3 bar”
Installation
The cooling water inlet temperature must be between 5°C and 20°C, and pipes must be insulated if condensation occurs, which affects site utilities and pipe routing.
Cooling water outside this range reduces cooling capacity or causes condensation, so the facility water system must be checked.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The electrical supply must be 200–240 V single or two-phase at 50/60 Hz with a 16 A fuse, and the maximum current draw is 14.5 A.
The branch circuit and protective device must be sized for 14.5 A and the specified voltage/frequency.
Source
“max. current 14,5 A”
Installation
Voltage and frequency tolerances are interdependent: a −10% voltage combined with +3% frequency is not allowed, so the site power quality must be checked.
Non-compliant supply conditions can cause malfunction or damage and must be avoided during commissioning.
Source
“-10% voltage and +3% frequency -> not allowed !”
Safety
External branch circuit protection according to UL 489 is required, which dictates the type of upstream breaker to be installed.
Using the wrong breaker type may violate compliance and void safety assumptions.
Source
“External branch circuit protection according to UL 489 required.”
Operation
The cooling water discharge temperature is limited to 55°C to prevent steam generation when flow temperatures exceed 100°C, which affects cooling water piping design.
The cooling water return line must tolerate up to 55°C and be routed accordingly.
Source
“The cooling water discharge temperature is limited to 55 °C to prevent steam from being generated during cooling at flow temperatures in excess of 100 °C.”
Operation
The unit is hydraulically closed and produces no oil vapour, which is important for laboratory air quality and operator exposure.
This affects ventilation requirements and suitability for enclosed laboratory environments.
Source
“hydraulically closed: No oil vapour.”
Performance
Cooling power is 10 kW at 100°C, 200°C, 300°C and 400°C, but these ratings are valid only at 20°C room temperature, 15°C cooling water inlet and 3 bar differential pressure.
Actual cooling capacity will be lower if site conditions differ from these reference values.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 3 bar differential pressure between cooling water inlet and - outlet.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the cooling water must be compatible with these materials.
Incompatible water chemistry can cause corrosion or fouling and must be assessed before connection.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
The minimum filling capacity is 2.3 l and the expansion volume is 5 l, which determines the minimum fluid charge and expansion space.
Underfilling or insufficient expansion volume can cause pump cavitation or overflow.
Source
“min. filling capacity 2,3 l”
Installation
The pump connections are M24x1.5 for both flow and return, and the cooling water connection is G1/2 male, so matching hoses and adapters must be procured.
Wrong thread types will prevent leak-free connection and delay commissioning.
Source
“Pump connenction flow M24x1,5”
Operation
The ambient temperature must be between 5°C and 40°C for safe operation.
Operating outside this range may reduce performance or cause overheating.
Source
“max. ambient temperature 40 °C”
Compliance
The unit is classified as Class A, Group 1 for electromagnetic disturbance to EN55011, which may require additional EMC measures in sensitive environments.
Class A equipment may not be suitable for residential or sensitive laboratory areas without filtering.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The cable between the machine and controller is approximately 5 m long, which constrains the physical layout of the installation.
The controller must be located within reach of the 5 m cable, affecting bench or floor planning.
Source
“Length of cable between machine and controller approx. 5m.”