Technical Reference Summary

Covered Models & Document Scope

Sponsored LCGC ebook (February 2020) covering food safety analysis workflows with the Agilent 6546 LC/Q-TOF, Agilent 1290 Infinity II LC, MassHunter Quant 10.1 and LC/Q-TOF Screener Tool; includes pesticide analysis in complex food matrices, nontargeted screening of adulterants/contaminants, and food authenticity testing with MassHunter Classifier. No instrument ordering codes or size variants are given.

Technical Data

Parameter Value Condition
Mass resolution type high mass resolution p.3 ↗ Agilent 6546 LC/Q-TOF
Mass accuracy high mass accuracy p.3 ↗ Agilent 6546 LC/Q-TOF
Mass accuracy requirement for identification less than 5 ppm p.4 ↗ SANTE/12682/2019, two ions
Minimum number of ions for identification two ions with at least one fragment ion p.4 ↗ SANTE/12682/2019
Minimum acquisition rate for quantitation 2 Hz p.5 ↗ 12 points over 6 s chromatographic peak
Data acquisition speed 6 Hz p.7 ↗ All Ions acquisition, strawberry method
Chromatographic peak width about six s p.5 ↗ typical peak
Dynamic range channels two channels (low-gain and high-gain), dual-gain 10-GHz spectrum p.5 ↗ Agilent 6546 LC/Q-TOF
LC back pressure rating 1,300 bar p.3 ↗ Agilent 1290 Infinity II LC
Detector data rate 240 Hz p.3 ↗ VWD and DAD
Autosampler carryover less than 9 ppm p.3 ↗ Agilent 1290 Infinity II LC
Autosampler sample capacity 6,144 samples in 16x384 well plates p.3 ↗ Agilent 1290 Infinity II LC
SureMass data conversion time 3–5 minutes per sample p.6 ↗ post-acquisition script
Number of pesticides quantified 180 p.6 ↗ proof of concept, four matrices
Number of pesticides screened over 200 p.6 ↗ proof of concept, four matrices
Mass error achieved less than 2 ppm p.8 ↗ almost all analytes at two concentrations
Column length 150 mm p.7 ↗ C18 column with guard column
Extraction solvent volume 10 mL p.7 ↗ acetonitrile, strawberry QuEChERS

Operating Limits

  • minimum mass accuracy for identification: less than 5 ppm p.4 ↗
  • minimum acquisition rate for quantitation: 2 Hz p.5 ↗
  • LC maximum back pressure: 1,300 bar p.3 ↗
  • autosampler carryover: less than 9 ppm p.3 ↗