Induztr
Engineering Digest Document Viewer Related Models

Rev j micro kjeldahl digestors user manual

Technical Reference Summary

Covered Models & Document Scope

Covers Labconco Micro Kjeldahl Digestors, catalog numbers 6030000 (115 V AC, with glass manifold) and 6030001 (230 V AC, with glass manifold). Replacement-parts chapters additionally cover legacy models 1200 & 1200-A, 1250 & 1250-A (manufactured before 1968), old-style units from 1968 to 1972 (serial numbers 1001 to 21858), new-style units from January 1972 to September 1972, and models 60300C, 60300-01C, 60301C & 60301-01C (January 1972 to present). Wiring diagrams are given for 6030000 & 6030100 and 6030001 & 6030101.

Section-by-Section Technical Breakdown

pp. 6
General Description and Detailed Function of Components

The digestor is intended for digestion of nitrogen-containing material from micro samples and accommodates six Micro Kjeldahl flasks of 30 ml or 100 ml. Each flask station has an individual 200 W heater assembly with a tubular element and stainless-steel cup, controlled by an infinite heat thermostat with its own pilot lamp. The cabinet is epoxy coated and the top is 1-1/2 inch thick Maronite I insulating board; the unit is factory wired with a 5 ft, 3-wire cord and plug.

  • Capacity: six Micro Kjeldahl flasks, 30 ml or 100 ml sizes.
  • Each station: individual 200 W heater assembly (tubular heater element plus stainless-steel cup).
  • Infinite heat controls provide variable input from 20-100% of capacity.
  • Individual pilot lamp per heater indicates 'HEATER ON'.
  • Cabinet epoxy coated; top is 1-1/2 inch thick Maronite I insulating board.
  • Factory wired including a 5 ft, 3-wire cord and plug; only an electrical receptacle is needed for installation.
pp. 8
Installation

Installation covers electrical requirements, unpacking, siting, mounting of the glass manifold support brackets and mounting/adjustment of the glass manifold. Electrical operating voltage is 115 or 230 V, maximum current 12 A at 115 V or 6 A at 230 V, at 50 or 60 Hz. 230 V models require permanent wiring connections by an electrician.

  • Operating volts: 115 or 230; current max: 12 A at 115 V, 6 A at 230 V; frequency 50 or 60 Hz.
  • Remove all fillers carefully and check for separately packed parts; 6030000 and 6030001 include a glass manifold, 2 support brackets and related hardware.
  • Place unit in a location having a convenient 120 V AC 3-wire receptacle.
  • 230 V AC models require permanent wiring connections; contact an electrician for initial electrical hook-up in all cases.
  • Mount glass manifold support brackets into the glide rod clamps on each side panel; loosen the thumb screw on the bottom glide clamp, slide the support rods in place and tighten the thumb screw.
  • Connect manifold outlet to a suitable water aspirator; adjust manifold to provide a minimum 1/2 inch of the flask neck within the manifold.
pp. 10
Normal Operation

Operation refers to applicable A.O.A.C. methods or other technical procedures for protein (nitrogen) determination by the Kjeldahl method. The infinite heat control uses a bi-metallic element to proportion heater on/off time: in the 'high' position the heater is on 100% of the time and in the 'low' position 20% of the time. Very small samples may require less heat than the lowest setting, in which case a voltage-reducing device should be used.

  • Refer to A.O.A.C. methods or other technical procedures for nitrogen determination by the Kjeldahl method.
  • Infinite heat control uses a bi-metallic element to proportion on/off time of the heating element.
  • 'High' position: heater on 100% of the time; 'low' position: heater on 20% of the time.
  • For extremely small samples requiring less heat than the lowest control setting, use a voltage-reducing device to reduce power input.
  • CAUTION: DO NOT boil flasks dry; breakage of the flask may occur.
pp. 10
Maintenance

The maintenance section directs users to contact Labconco Customer Service if a problem arises, and to include the catalog and serial numbers on all phone or letter inquiries for prompt action.

  • If a problem arises, contact Labconco Customer Service Department at 800-821-5525 or 816-333-8811 between 7:00 a.m. and 6:00 p.m. Central Standard Time.
  • Include Labconco Catalog and Serial Numbers on all phone or letter inquiries.
pp. 11-13
Replacement Parts

Replacement parts are listed by production era and model: old style before 1968 (models 1200 & 1200-A, 1250 & 1250-A), old style from 1968 to 1972 (serial numbers 1001 to 21858), new style from January 1972 to September 1972, and models 60300C, 60300-01C, 60301C & 60301-01C from January 1972 to present. Parts include glass manifolds, support rods and brackets, infinite controls, knobs, heating elements and heater assemblies, pilot lamps, flask support cones and assemblies, power cord sets, top plates, vent louvers, heater conversion kits and flask kits.

  • Old style (before 1968): heating elements 6013200 (115 V, 150 W), 6013300 (115 V, 200 W), 6020500 (230 V, 250 W); infinite controls 1315400 (120 V, 15 A) and 1315500 (240 V, 15 A); pilot lights 6011900 (115 V) and 6020300 (230 V); flask support cones 6011000 (30 ml) and 6000700 (100 ml).
  • Old style (1968-1972, serial 1001 to 21858): heater assemblies 6030700 (115 V, 200 W) and 6030701 (230 V, 200 W); infinite controls 1315400 (115 V, 15 A) and 1315500 (230 V, 15 A); power cords 6010300 (115 V) and 6020700 (230 V); heater conversion kits 6035700 (115 V) and 6035800 (230 V); flask kit 6037500 (6 @ 100 ml).
  • New style (Jan-Sep 1972): heater assemblies 6035000 (115 V, 200 W) and 6035100 (230 V, 200 W); top plate 6033600 (Marinite I); upright supports 6034200 and 6034300.
  • Models 60300C, 60300-01C, 60301C & 60301-01C (Jan 1972 to present): same as new style except heater assemblies 6034600 (115 V, 200 W with heater cup) and 6034700 (230 V, 200 W with heater cup).
  • Common parts across eras include glass manifold 6001100, knob 1870200, guide rod clamp 6032100, pilot lamps 6011900/6020300, and power cords 6010300/6020700.
pp. 14-15
Wiring Diagrams

Wiring diagrams are provided for catalog numbers 6030000 & 6030100 and for 6030001 & 6030101.

  • Wiring diagram for Catalog No. 6030000 & 6030100 models.
  • Wiring diagram for Catalog No. 6030001 & 6030101 models.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Operating voltage is 115 or 230 V AC; maximum current is 12 A at 115 V and 6 A at 230 V; frequency 50 or 60 Hz. p.8 ↗ Size the branch circuit and receptacle for at least 12 A at 115 V or 6 A at 230 V; verify supply frequency compatibility. 6030000 (115 V) and 6030001 (230 V)
230 V AC models require permanent wiring connections; contact an electrician for initial electrical hook-up in all cases. p.8 ↗ Plan for hard-wired installation of 230 V units rather than a cord-and-plug connection. 6030001 (230 V AC)
The unit is factory wired including a 5 ft, 3-wire cord and plug; an electrical receptacle is all that is needed for installation. p.6 ↗ For 115 V models, provide a nearby 120 V AC 3-wire receptacle within reach of the 5 ft cord. 6030000 (115 V AC)
All Micro Digestors accommodate six Micro Kjeldahl flasks of 30 ml or 100 ml sizes. p.6 ↗ Select flask size and matching flask support cones/assemblies accordingly; six stations are available per unit. Whole series
Each flask station has an individual 200 W heater assembly consisting of a tubular heater element and a stainless-steel cup. p.6 ↗ Total connected heater load is six 200 W elements; the concave cup holds the flask securely and ensures even heat distribution. Whole series
Infinite heat controls have variable input from 20-100% of capacity; in the 'high' position the heater is on 100% of the time and in the 'low' position 20% of the time. p.6 ↗ Use the control range to proportion heating; the lowest setting still delivers 20% duty cycle. Whole series
A digestion of extremely small samples may require less heat than available from the lowest control setting; in that event a voltage-reducing device should be used to reduce power input. p.10 ↗ For micro-scale digestions, plan for an external voltage-reducing device to avoid overheating. Whole series
DO NOT boil flasks dry; breakage of the flask may occur. p.10 ↗ Monitor digestion to prevent dry boiling and flask breakage. Whole series
The glass manifold should be adjusted to provide a minimum 1/2 inch of the flask neck within the manifold, and the manifold outlet connected to a suitable water aspirator. p.8 ↗ Set manifold height so at least 1/2 inch of flask neck is inside the manifold; provide a water aspirator for exhaust. 6030000 and 6030001 (with glass manifold)
Catalog #6030000 and 6030001 Digestors ship with a glass manifold, 2 support brackets and related hardware packed in the carton. p.8 ↗ Verify these loose parts are present before installation; remove fillers carefully and check for separately packed parts. 6030000 and 6030001
The cabinet top is 1-1/2 inch thick Maronite I insulating board to protect wiring and controls from heat and chemicals; the cabinet is epoxy coated. p.6 ↗ The top plate provides thermal and chemical protection; replacement top plates are listed as 6030900 (Marinite #23) and 6033600 (Marinite I). Whole series
Replacement heater assemblies for current models 60300C, 60300-01C, 60301C & 60301-01C are 6034600 (115 V, 200 W with heater cup) and 6034700 (230 V, 200 W with heater cup). p.13 ↗ Order the heater-cup version for these models; earlier new-style units use 6035000/6035100 without the cup. 60300C, 60300-01C, 60301C & 60301-01C
Heater conversion kits are available as 6035700 (115 V) and 6035800 (230 V). p.12 ↗ Voltage conversion of old-style units (1968-1972) is supported via these kits. Old style, serial numbers 1001 to 21858
Infinite controls are rated 15 A: 1315400 (120/115 V) and 1315500 (240/230 V). p.11 ↗ Replacement controls must match the unit voltage; both are 15 A rated. Legacy models listed in Chapter 5
Flask support cones are available as 6011000 (30 ml) and 6000700 (100 ml); flask kit 6037500 contains 6 @ 100 ml flasks. p.11 ↗ Match support cones to the flask size used; 100 ml flasks are available as a six-pack kit. Old style models (before 1968 and 1968-1972)
Pilot lamps are voltage-specific: 6011900 (115 V) and 6020300 (230 V). p.11 ↗ Select the pilot lamp matching the unit's operating voltage. Whole series
Power cord sets are voltage-specific: 6010300 (115 V) and 6020700 (230 V). p.11 ↗ Order the correct cord set for the unit voltage when replacing. Whole series
Applicable methods for nitrogen determination are those outlined by the A.O.A.C. or other technical procedures for protein (nitrogen) determination by the Kjeldahl method. p.10 ↗ Follow A.O.A.C. or equivalent Kjeldahl procedures for digestion; the digestor provides controlled heating only. Whole series