Rev f rapidstill ii user manual
Technical Reference Summary
Covered Models & Document Scope
This manual covers the Labconco RapidStill II steam distillation unit, model 6520000 (Part #6544100, Rev. F). It documents the standard configuration and the optional Water Inlet Kit, part #6544200, which allows separate purified water feed to the boiler and tap water to the condenser.
Section-by-Section Technical Breakdown
The RapidStill II produces steam by heating a 1-liter boiler with a 1200 W heating element. Front-panel switches control boiler water fill, boiler heater, NaOH addition, and an infinitely adjustable 0–15 minute distillation timer. Typical distillation times are 5 to 10 minutes. The unit includes a combined Allihn and spiral condenser, a peristaltic NaOH pump, and safety interlocks.
- Boiler capacity: 1 liter; heating element: 1200 W.
- Distillation timer range: 0–15 minutes, infinitely adjustable; typical times 5–10 minutes.
- Boiler Water switch is interlocked to prevent filling during heating.
- NaOH Addition switch is interlocked to operate only when the boiler heater is on.
- NaOH cannot be dispensed when the distillation timer is in operation.
- Flow restrictor limits cooling water flow to 1.0 gallon per minute.
- 15 A fuse protects against electrical overload; never install a higher amperage fuse.
The unit must be placed on a level work surface near electrical, water, and drain connections. Electrical service must be 115 VAC, 60 Hz, 15 A, dedicated to the unit. Feed water must be municipal drinking water grade with a minimum supply of 1 gallon (3.8 L) per minute at 10 PSI. The drain hose must be routed below the drain pipe outlet and must not be restricted. The NaOH pump inlet uses a quick connector.
- Electrical outlet: 115 VAC, 60 Hz, 15 Amps, dedicated circuit.
- Feed water: municipal drinking water grade, minimum 1 gallon (3.8 L) per minute at 10 PSI.
- Drain system must be rated for a minimum capacity of 2 gallons (7.6 L) per minute.
- Drain hose and all connections must be below the level of the drain pipe outlet; gravity drainage only.
- Do not restrict the drain hose or connections; restriction may cause pressure buildup and damage.
- NaOH pump inlet quick connector locks with a click; remove by depressing the metal button.
The Optional Water Inlet Kit, part #6544200, allows two separate water supplies: ordinary tap water for the condenser and purified water for the boiler. The purified water source must deliver 1 gallon (3.8 L) per minute at a minimum of 5 PSI and must not exceed 20 PSI. Installation involves installing a bulkhead union, relocating the flow restrictor, and connecting the braided tube assembly.
- Purified water supply: 1 gallon (3.8 L) per minute at minimum 5 PSI; maximum 20 PSI.
- Upper bulkhead fitting supplies the boiler with purified water.
- Lower bulkhead fitting supplies only the condenser with cooling water.
- Do not overtighten the nut on the bulkhead union; it may damage the fitting.
- Remove the flow restrictor assembly from the T-shaped fitting and plug its outlet into the condenser inlet fitting.
The standard operating procedure describes turning on cooling water, filling the boiler to approximately 2/3 full, turning on the boiler heater, loading the digestion tube, placing the receiver flask, adding NaOH, and setting the distillation timer. For each 5 ml of concentrated sulfuric acid used in digestion, add approximately 15 ml of 50% NaOH solution. The distillate should be titrated, and if its temperature exceeds 120°F (50°C), the condenser cooling water flow rate should be checked.
- Cooling water flow rate is regulated to 1 gallon per minute (3.8 L/min).
- Fill boiler approximately 2/3 full or to the top line of the flask.
- Heat boiler for approximately 10 minutes or until steady boil before operation.
- Add approximately 15 ml of 50% NaOH per 5 ml of concentrated sulfuric acid used in digestion.
- Receiver flask outlet tube must be fully submerged.
- If distillate temperature exceeds 120°F (50°C), check condenser cooling water flow rate.
- Never turn on boiler heater when boiler is dry or low on water; do not refill when hot and dry.
Daily maintenance includes rinsing the caustic pump, delivery system, and distribution head with purified water, and cleaning exterior surfaces with mild soap and water. Weekly maintenance includes examining the boiler for deposits (removable with dilute hydrochloric acid or compatible descaling solutions) and inspecting all plumbing, steam lines, tubing connections, and the connection stopper for deterioration or cracking.
- Rinse caustic system daily with purified water as described in Chapter 4.
- Clean exterior surfaces with a soft cloth and mild soap and water.
- Weekly: examine boiler for deposits; use dilute hydrochloric acid or glass-compatible descaling solutions.
- Weekly: examine all plumbing and steam lines for integrity and connection stopper for deterioration or cracking.
Service procedures cover fuse replacement, boiler removal, heating element replacement, condenser extension replacement, timer replacement, and NaOH system service including peristaltic pump tube replacement. Always unplug the unit before servicing. Use only proper replacement parts and never install a fuse with a higher amperage rating.
- Fuse replacement: turn slotted fuse holder ¼ turn counterclockwise, pull straight out.
- Boiler removal: pull boiler head straight up; if stuck, apply water to the joint.
- Heating element replacement requires removing the perforated screen (three screws), disconnecting two wires and ground wire, removing four nuts and lockwashers, and removing three screws on the underside of the cup.
- Condenser extension replacement: orient the drip point of the new tube to the bottom.
- Timer replacement: remove four front panel screws, loosen two set screws on timer knob, disconnect electrical connections noting terminal orientation.
- NaOH pump tube replacement: remove four small Phillips screws and pump cover; retain fittings and clamps for new tube.
The replacement parts list provides part numbers for major components including the heating element, boiling flask, boiler head, balance tee, plastic check valve, steam control valve, water fill valve, tubing, flow restrictor assembly, peristaltic pump tube, fuse, switches, connection stopper, caps/gaskets, dispensing tube, distribution head, condenser extension, condenser, ventilation valve, distillation outlet tube, and timer assembly.
- Heating Element: part #6539400.
- Boiling Flask, 1000 M/L: part #6523000.
- Boiler Head: part #6541600.
- Peristaltic Pump Tube: part #6544800.
- Fuse, 15 Amp: part #1344400.
- Condenser Extension: part #6543600.
- Condenser: part #6541000.
- Timer Assembly: part #6543400.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Electrical outlet must be rated for 115 VAC, 60 Hz, 15 Amps, and dedicated to the RapidStill II. p.11 ↗ | Ensure a dedicated 15 A circuit is available at the installation site to prevent tripping and ensure proper operation. | Model 6520000 |
| Feed water must be municipal drinking water grade with a minimum supply volume of 1 gallon (3.8 liters) per minute at 10 PSI. p.11 ↗ | Verify water supply pressure and flow rate before installation; inadequate supply may cause slow boiler filling or scaling. | Model 6520000 |
| Drain system must be rated for a minimum capacity of 2 gallons (7.6 liters) per minute, and the drain hose must be below the drain pipe outlet. p.12 ↗ | Gravity drainage is required; restricting or elevating the drain hose may cause pressure buildup and damage. | Model 6520000 |
| Flow restrictor limits cooling water flow rate to 1.0 gallon per minute. p.9 ↗ | Do not remove or bypass the flow restrictor; it minimizes water usage and prevents damage to the unit. | Model 6520000 |
| Optional Water Inlet Kit #6544200 requires purified water supply of 1 gallon (3.8 L) per minute at minimum 5 PSI and maximum 20 PSI. p.13 ↗ | Ensure purified water source meets these pressure and flow requirements to avoid damage or poor performance. | Model 6520000 with Optional Water Inlet Kit #6544200 |
| Boiler should be filled approximately 2/3 full or to the top line of the flask. p.16 ↗ | Proper water level is critical; never operate heater when boiler is dry or low on water to prevent flask breakage and element damage. | Model 6520000 |
| For each 5 ml of concentrated sulfuric acid used in digestion, add approximately 15 ml of 50% NaOH solution. p.16 ↗ | This ratio ensures the distillation solution is strongly alkaline for complete ammonia release. | Model 6520000 |
| Distillation timer is infinitely adjustable from 0 to 15 minutes; typical distillation times range from 5 to 10 minutes. p.5 ↗ | Set timer according to sample type and volume; allow enough time for distillation solution to reach boiling. | Model 6520000 |
| If distillate temperature exceeds 120°F (50°C), check the flow rate of the condenser cooling water. p.17 ↗ | High distillate temperature indicates insufficient cooling, which may affect recovery and safety. | Model 6520000 |
| NaOH pump lines must be rinsed with purified water at the end of every work day. p.6 ↗ | Regular rinsing prevents caustic crystallization and damage to the peristaltic pump and fittings. | Model 6520000 |
| Fuse is 15 Amp; never install a fuse with a higher amperage rating. p.9 ↗ | Use only proper replacement part #1344400 to maintain electrical protection. | Model 6520000 |
| Heating element is 1200 W and heats a 1-liter boiler. p.5 ↗ | Ensure electrical supply can handle the load; dedicated 15 A circuit is required. | Model 6520000 |
| Boiler Water switch is interlocked to prevent operation during heating of the boiler. p.6 ↗ | Do not attempt to fill boiler while heater is on; allow boiler to cool before adding water. | Model 6520000 |
| NaOH Addition switch is interlocked to operate only when the boiler heater is on. p.6 ↗ | Ensure boiler heater is on before attempting to add NaOH. | Model 6520000 |
| NaOH cannot be dispensed into the digestion tube when the distillation timer is in operation. p.7 ↗ | Add NaOH before starting the distillation timer. | Model 6520000 |
| Condenser extension replacement requires orienting the drip point of the tube to the bottom. p.23 ↗ | Incorrect orientation may cause poor distillation performance or damage. | Model 6520000 |
| Weekly maintenance includes examining boiler for deposits and cleaning with dilute hydrochloric acid or glass-compatible descaling solutions. p.20 ↗ | Regular descaling prevents scaling and maintains heat transfer efficiency. | Model 6520000 |
| Connection stopper should be inspected and cleaned after every operating session. p.6 ↗ | Prevents deterioration and ensures proper seal between digestion tube and distribution head. | Model 6520000 |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Unit will not operate at all p.25 ↗ | Tripped circuit breaker | Check facility’s circuit breaker | |
| Unit will not operate at all p.25 ↗ | Blown fuse | Replace fuse | |
| Unit will not operate at all p.25 ↗ | Power cord not plugged in | Plug in cord | |
| Boiler doesn’t fill p.25 ↗ | ‘Boiler Heater’ switch is on | Turn off boiler heater before filling | |
| Boiler doesn’t fill p.25 ↗ | Cooling water off or flow rate insufficient | Check flow rate of condenser cooling water | |
| Boiler doesn’t fill p.25 ↗ | Water fill valve plugged or defective | Clean inlet screen of filter; replace valve | |
| Boiler doesn’t fill p.25 ↗ | ‘Boiler Water’ switch defective | Replace switch | |
| Heating element doesn’t work p.25 ↗ | Heating element is defective | Replace heating element | |
| Heating element doesn’t work p.25 ↗ | ‘Boiler Heater’ switch is defective | Replace switch | |
| NaOH pump runs, but doesn’t pump p.26 ↗ | NaOH tube is kinked or restricted | Check all lines for kinks or blockage | |
| NaOH pump runs, but doesn’t pump p.26 ↗ | Peristaltic pump tube damaged | Replace peristaltic pump tube | |
| NaOH pump runs, but doesn’t pump p.26 ↗ | Peristaltic pump is damaged | Replace peristaltic pump | |
| Water is noisy when flowing through the unit p.26 ↗ | Excessive water flow rate | Reduce the water flow rate slightly | |
| Steam drains erratically; steam valve is noisy during standby p.26 ↗ | Drain hose is kinked or elevated above the drain pipe | Drain hose must flow unrestricted to a drain below the drain pipe | |
| Boiling flask breaks p.26 ↗ | Flask has gone dry | Do not let flask go dry | |
| Boiling flask breaks p.26 ↗ | Water added when boiler is dry or low on water and hot | Do not add water to the boiler when it is hot and dry, or low on water | |
| Recoveries are low p.26 ↗ | Seal is leaking | Clean/replace the seal | |
| Recoveries are low p.26 ↗ | Receiving tube is not completely submerged | The receiving tube must be completely submerged in the receiving vessel | |
| Recoveries are low p.26 ↗ | Not collecting enough distillate | Increase distillation time | |
| Recoveries are too high p.26 ↗ | Excessive NH3 in feed water | Connect to a different water source; install Optional Water Inlet Kit #6544200 and feed purified water to the boiler |