Rev p rapidstill i user manual
Technical Reference Summary
Covered Models & Document Scope
Covers the Labconco RapidStill I steam distillation apparatus, model 6500000, part #6504000 Rev. P. Includes 115 V configuration and a 230 V variant via transformer 1317400 (230 V to 115 V, 50/60 Hz, 300 W).
Section-by-Section Technical Breakdown
The RapidStill I is a semi-automatic steam distillation unit for sulfuric acid digests of nitrogen-bearing materials. It accepts digested samples up to 4 ml equivalent of concentrated acid and can give reproducible results down to 10 micrograms (gamma) nitrogen. It may also be used for other steam distillations that do not exceed the mixing chamber volume of approximately 50 to 55 ml.
- Designed for rapid, semi-automatic steam distillation from sulfuric acid digests.
- Sample digestion performed with Labconco Digestion Apparatus or Rapid Digestor units.
- Maximum digested sample volume: 4 ml equivalent of concentrated acid.
- Reproducible results in ranges as low as 10 micrograms (gamma) nitrogen.
- Mixing chamber volume limitation approximately 50 to 55 ml.
Identifies the main controls and glassware of the RapidStill I, including the power switch, heat control, fill valve, aspirator stopcock, sample addition funnel and stopcock, mixing chamber, cooling water connections, condenser, boiling flask, adjustable flask shelf, heater, and water bottle. The fill valve is used only when filling the boiling flask, and the aspirator uses condenser cooling water to purge spent samples.
- Power Switch switches power to the heater.
- Heat Control adjusts input to the electric immersion heater and steam output.
- Fill Valve opened only when filling the Boiling Flask; keep reservoir approximately 2/3 full during distillation.
- Aspirator Stopcock draws sample and rinse water down the drain using condenser cooling water.
- Sample Addition Stopcock is the upper PTFE stopcock for samples or pH-controlling solutions.
- Cooling water connections: 'IN' fitting to cold water source, 'OUT' fitting to drain.
- Flask Shelf is adjustable in height to maintain relationship between condenser outlet and receiving flask liquid level.
- Heater is energized when sufficient water is in the Boiling Flask and may be unplugged from the housing.
- Use distilled water only in the still; Water Bottle is squeezed to introduce water via the Fill Valve.
Describes unpacking, assembly of the glass still onto the base housing, connection of cooling water tubing, and electrical connection. Plastic tubing and valve for condenser water and drain are not included; 3/8" ID x 1/16" wall flexible plastic or silicone tubing is required. The unit requires a convenient 115 V outlet and cold tap water (<25 °C) with fine flow control.
- Carton contains Base Housing, Glass Still, Heater, Stainless Steel Flask Shelf, one large and two small U-clamps, six clamp nuts, Fill Valve and Tube, two pinch clamps, and Power Cord.
- Plastic tubing and valve for condenser water and drain are NOT included.
- Use flexible plastic or silicone tubing 3/8" ID x 1/16" wall for water connections.
- Connect inlet to cold water source <25 °C with fine flow control through a valve or faucet.
- Do not add clamps or any lubricant other than water to the glass barb fittings.
- Insert heater element into Boiling Flask and press stopper firmly; heater must not contact glass.
- Do not connect to electrical power until set-up is complete.
Provides the start-up procedure to verify correct operation before sample analysis. The boiling flask is filled 2/3 full with ammonia-free distilled or deionized water, cooling water is turned on, and the heat control is set to 9 until boiling begins in about 5 to 6 minutes. The heat control is then reduced to obtain a distillate rate of 5 ml/min and cooling water adjusted to keep distillate at 25 °C or lower.
- Never energize the heater unless the heater element is properly immersed in water.
- Fill Boiling Flask 2/3 full with ammonia-free distilled or deionized water through the Fill Valve; keep Fill Valve closed.
- Sample Addition Stopcock and Aspirator Stopcock must be closed before start-up.
- Set Heat Control dial to 9; water begins boiling in about 5 to 6 minutes.
- Reduce Heat Control to obtain a distilling rate of 5 ml/min at the output.
- Adjust cooling water to maintain distillate temperature of 25 °C or lower.
- For low-level nitrogen determinations, allow steam distillation until distillate is nitrogen free; may require 30 minutes at initial start-up.
- Add approximately 15 ml ammonia-free distilled water to Mixing Chamber to check aspirator siphon operation.
Describes the sample distillation procedure for samples digested with a Micro Digestor, conforming to standard Micro Kjeldahl distillation techniques. The procedure covers cooling water, heater warm-up, sample addition, alkali addition, distillation timing, and rinsing between samples. Average distillation time is 5 to 10 minutes.
- Set Heat Control at 9 to bring steam reservoir to a roll, then reduce to a 4 to 5 ml/min distilling rate.
- Place entire 4 ml maximum acid digestion sample in the Sample Addition Funnel.
- Rinse Sample Addition Funnel with 3 to 4 ml ammonia-free distilled water, leaving a liquid seal.
- Position Flask Shelf so condensate outlet is slightly below the receiving liquid surface level.
- Add strong alkali slowly; stop flow if neutralizing action becomes too vigorous or causes receiving solution to siphon backward.
- If reaction is uncontrollable, reduce subsequent acid digest sample volumes and recalculate alkali strength so inner chamber contents have a normality of 4 to 8.
- If foaming occurs, add 1 or 2 drops of anti-foam reagent.
- Average distillation time is 5 to 10 minutes; lower Flask Shelf and continue approximately one minute.
- Close Aspirator Stopcock before addition of each rinse.
- If long delays between sample runs, rinse Mixing Chamber with distilled or deionized water before new sample distillation.
Covers shutdown steps and the formulas for standard acid addition, percent nitrogen, and percent protein. The mixing chamber is left about half full (10 to 12 ml) after rinsing, and cooling water is turned off. Protein factors are given for wheat, milk, rice, and all other materials.
- Turn OFF heater with Power Switch, remove Receiving Flask, rinse Mixing Chamber leaving it about 1/2 full (10 to 12 ml), then turn off cooling water.
- Standard acid to add (ml) = [(% nitrogen expected) x (sample aliquot x g sample weight) + 2] / [(normality of standard acid) x 1.4007 x 250].
- %Nitrogen = [(ml std acid x normality) - (ml std base x normality)] x 1.4007 / effective weight.
- %Protein = [(ml std acid x normality) - (ml std base x normality)] x 1.4007 x factor / effective sample weight.
- Effective sample weight = weight of sample in grams x sample aliquot / 250.
- Protein factors: wheat 5.70, milk 6.38, rice 5.95, all other 6.25.
Describes replacement of the glass still assembly and maintenance of the two PTFE stopcocks. PTFE stopcocks require no lubricant and may be cleaned with acetone. Alkalis or other glass-corrosive liquids require thorough water rinsing after each use to prevent scoring of the PTFE plug.
- Disconnect power cord before performing any service work.
- Remove heater assembly by unplugging it and remove stopper from flask.
- Disconnect all tubing connections on glass still, lay Base Housing on its back, and remove three U-clamp nuts and bolts.
- PTFE stopcocks do not require lubricants; clean plug and glass barrel with acetone to remove solid particles.
- Solid particles lodged between plug and glass barrel can score the PTFE plug and cause a leaky plug.
- Rinse all valve components thoroughly with water after each use with alkalis or other glass-corrosive liquids.
Lists replacement part numbers for the RapidStill I, including glass still assembly, PTFE plugs, heat controller, power switch, power cord, circuit breaker, shelf, heater assembly, U-clamp hardware, tubing, fill valve, transformer, and water bottle.
- Glass Still Assembly with PTFE stopcocks: 6500100.
- PTFE plug replacements for PTFE stopcock: 1352500.
- Heat Controller: 6507700; Knob for Heat Controller: 1876500.
- Power Switch: 1327000; Power Cord, 115V: 1331300; Circuit Breaker, 3 amp: 1327200.
- Shelf for Receiving Flask: 6506900.
- Heater assembly, complete with stopper and male plug; 115 VAC, 280 Watts: 6504201.
- Large U-Clamp bolt (1 required): 6507200; Small U-Clamp bolt (2 required): 6507100; Nuts for U-Clamps (6 required): 6507500.
- Tubing for U-Clamps (1/4" ID): 1624400; Fill Valve: 1355400.
- Transformer (230 V to 115 V, 50/60 Hz, 300 Watt): 1317400.
- Water Bottle: 6507600; Tubing for water outlet (3/8" ID, 6" required): 1625000; Tubing for water inlet (12mm ID, 9" required): 6523100.
Chapter 6 is designated as the wiring diagram section. The extracted text contains only the chapter heading and no diagram content or textual wiring details.
- Wiring diagram chapter present; no textual wiring details available in the extracted document.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Maximum digested sample volume is 4 ml equivalent of concentrated acid. p.5 ↗ | Do not exceed 4 ml acid digest per distillation; larger volumes may cause uncontrollable reactions or exceed mixing chamber capacity. | RapidStill I model 6500000 |
| Mixing chamber volume is approximately 50 to 55 ml. p.5 ↗ | Other steam distillations must not exceed this volume limitation. | RapidStill I model 6500000 |
| Reproducible results in ranges as low as 10 micrograms (gamma) nitrogen. p.5 ↗ | Suitable for low-level nitrogen determinations; allow steam distillation until distillate is nitrogen free, which may require 30 minutes at initial start-up. | RapidStill I model 6500000 |
| Boiling Flask should be filled 2/3 full with ammonia-free distilled or deionized water. p.11 ↗ | Maintain sufficient water level at all times; never energize heater unless element is properly immersed. | RapidStill I model 6500000 |
| Water in the Boiling Flask begins boiling in about 5 to 6 minutes with Heat Control at 9. p.13 ↗ | Use Heat Control setting 9 for warm-up, then reduce to achieve the desired distillation rate. | RapidStill I model 6500000 |
| Target distilling rate is 5 ml/min during initial test; normal operation uses 4 to 5 ml/min. p.13 ↗ | Adjust Heat Control to maintain this rate for reproducible distillations. | RapidStill I model 6500000 |
| Distillate temperature must be maintained at 25 °C or lower by adjusting cooling water input. p.13 ↗ | Insufficient cooling can cause loss of nitrogen; adjust tap water flow accordingly. | RapidStill I model 6500000 |
| Cooling water inlet must be cold water <25 °C with fine flow control through a valve or faucet. p.10 ↗ | Installation requires accessible cold tap water and drain; use 3/8" ID x 1/16" wall tubing. | RapidStill I model 6500000 |
| Inner chamber contents should have a normality of 4 to 8 during distillation. p.15 ↗ | If reaction is uncontrollable, reduce subsequent acid digest sample volumes and recalculate required alkali solution strengths. | RapidStill I model 6500000 |
| Average distillation time is 5 to 10 minutes. p.15 ↗ | Plan sample throughput accordingly; lower Flask Shelf and continue approximately one minute after distillation. | RapidStill I model 6500000 |
| After shutdown, rinse Mixing Chamber leaving it about 1/2 full (10 to 12 ml). p.16 ↗ | Prevents residue buildup and keeps stopcock and chamber ready for next use. | RapidStill I model 6500000 |
| Protein conversion factors: wheat 5.70, milk 6.38, rice 5.95, all other 6.25. p.16 ↗ | Use the appropriate factor in the %Protein calculation. | RapidStill I model 6500000 |
| Heater assembly is 115 VAC, 280 Watts; circuit breaker is 3 amp. p.19 ↗ | Electrical supply and replacement parts must match these ratings; 230 V operation requires transformer 1317400 (230 V to 115 V, 50/60 Hz, 300 Watt). | RapidStill I model 6500000 |
| PTFE stopcocks require no lubricant and may be cleaned with acetone. p.18 ↗ | Do not lubricate; clean to remove solid particles that can score the plug and cause leaks. | RapidStill I model 6500000 |
| Rinse all valve components thoroughly with water after each use with alkalis or other glass-corrosive liquids. p.18 ↗ | Prevents liquid concentrate formation on evaporation, which can score the PTFE plug. | RapidStill I model 6500000 |
| Use 3/8" ID x 1/16" wall flexible plastic or silicone tubing for water connections; tubing is not furnished. p.10 ↗ | Procure tubing separately from laboratory supply dealers; two pinch clamps are provided. | RapidStill I model 6500000 |
| Heater element must not contact any glass and the stopper must be firmly sealed for distillation. p.10 ↗ | Improper seating can cause glass breakage or loss of steam pressure. | RapidStill I model 6500000 |
| If foaming is a problem during distillation, add 1 or 2 drops of anti-foam reagent. p.15 ↗ | Prevents foam from carrying over into the receiving flask. | RapidStill I model 6500000 |