Rev c capture portable fuming systems user manual
Technical Reference Summary
Covered Models & Document Scope
Covers the Labconco CApture Portable Fuming Systems, models 3420000 (115 VAC, 60 Hz, 1 Amp) and 3420020, 3420022, 3420024 (230 VAC, 50/60 Hz, 0.5 Amp). Optional accessories include the Exhaust Connection Kit (#3425400).
Section-by-Section Technical Breakdown
Lists required supplies (CA glue, aluminum heating pans, Rain-X or Rain Block treatment) and gives electrical and exhaust requirements. Electrical data are tabulated per model, and notes cover circuit breaker ratings, power cord length, and inverter operation.
- CA glue capacity of filter is approximately 2 oz (57 g); do not use thickening or gap-filling formulations.
- Aluminum heating pans: 10 shipped; bag of 250 available as Labconco part 34252002.
- Exhaust Connection Kit optional; do not add additional duct beyond that supplied.
- Model 3420000: 115 VAC, 60 Hz, 1 Amp; models 34200-20, -22, -24: 230 VAC, 50/60 Hz, 0.5 Amp.
- Circuit breaker: 10 A for 115 V models, 5 A for 230 V models; power cords 3 m (10 ft).
- Inverter operation: 12 VDC to AC, output rating at least 100 W, protected against current overload.
Explains CA polymerization, controlled heating to generate fumes, the clear polycarbonate chamber, the two-stage filter system, and the blower. Includes temperature limits and filter efficiency data.
- CA boiling point approximately 120°F (50°C); decomposition products above 250°F (121°C); hydrogen cyanide can be produced above 400°F (204°C).
- Controller setpoint range: 120°F (50°C) to 350°F (177°C); redundant safety switch limits heater to 350°F (177°C).
- Clear chamber must be treated with Rain-X or Rain Block; protection lasts up to 20 uses if undisturbed.
- First filter stage: polyester loft particulate filter, MERV 11 (65-79% of 1.0-3.0 µm particles; >85% of 3.0-10 µm particles).
- Second filter stage: granular activated carbon (GAC) adsorbs gaseous CA.
- Filter life: minimum 100 fuming cycles (about 2 oz or 57 ml of CA).
Covers unpacking, transporting, component identification, handling hints for CA, setup, heater adjustment, fuming procedure, and humidity adjustment. Provides step-by-step operating instructions and key parameter values.
- System weight approximately 22 lbs (10 kg); transport only in transport mode with heater/filter module nested and transport lid locked.
- Setup: release clamps, remove foam insert, lower leveling foot, treat chamber, select hooks/shelves, seal chamber, plug in and switch main power ON.
- Heater adjustment: push control switch to Heat; setpoint range 120-350°F (50-177°C); temperature stabilizes within 3-5 minutes at setpoint ±15°F.
- Fuming: use 4-6 drops of CA (never exceed 8 drops); typical development 3-5 minutes with 4 drops at 350°F (177°C).
- After fuming, switch to Vent and let chamber vent at least 3-5 minutes before opening.
- Humidity: optimum 70-90% RH, best at approximately 80%; one drop of water raises RH by approximately 2%.
Provides a routine maintenance schedule and service operations including filter replacement and circuit breaker reset. Specifies cleaning intervals and treatment reapplication.
- Weekly: clean stainless steel face plate with soap/detergent or LabSolutions wipes (part #1570000); reapply Rain-X or Rain Block if chamber interior cleaned.
- Monthly: clean exterior surfaces with damp cloth or LabSolutions wipes.
- After every 20 development cycles: clean clear chamber with mild detergent or glass cleaner, then treat with Rain-X or Rain Block.
- After every 100 development cycles: replace filter cartridge.
- Filter replacement: unplug, separate module, loosen captive thumbscrew, remove panel, pull cartridge out of seal and flexible connector, install replacement, ensure seal not deformed.
- Circuit breaker reset: press switch on rear of steel module to ON (I) position.
Presents a troubleshooting table with problems, causes, and corrective actions for common system failures. Covers power, print development, fume generation, heater, fan, and controller errors.
- Blower and controller won't turn on: check outlet and circuit breaker.
- Prints not developing properly: insufficient humidity; add 1-2 drops of water in a heating pan.
- Little or no fumes: heater temperature too low or pan bottom dented/bent.
- Heater temperature not changing: possible over-temperature switch failure or defective controller.
- Fan does not run: fan or 12 V power supply failure.
- Controller displays 'Er' or 'oo' or '-': controller error.
Lists replacement parts and accessories with part numbers and quantities. Includes clear chamber, heating pans, flexible connector, filter cartridge, shelf, hook, and foam storage insert.
- Clear chamber: part 3421901.
- Heating pans, package of 250: part 3425202.
- Flexible connector: part 3424500.
- Filter cartridge: part 3421801.
- Shelf: part 3422203; Hook: part 3422202.
- Foam storage insert: part 3425500.
Provides the interior dimensions of the clear chamber.
- Clear chamber interior: 16" (406 mm) H x 8" (203 mm) W x 8" (203 mm) D.
Lists electrical data and environmental operating conditions for the system.
- Electrical: 3420000 115 VAC, 60 Hz, 1 Amp, 75 W; 34200-20, -22, -24 230 VAC, 50/60 Hz, 0.5 Amp, 75 W.
- Indoor use only; maximum altitude 6562 ft (2000 m).
- Ambient temperature range: 41°F to 104°F (5°C to 40°C).
- Maximum relative humidity: 80% up to 88°F (31°C), decreasing linearly to 50% at 104°F (40°C).
- Main supply voltage fluctuations not to exceed ±10% of nominal.
- Transient overvoltages per Installation Category II: up to 1500 V for 115 V models, 2500 V for 230 V models.
- Pollution degree 2 environment.
Lists optional and replacement accessories with part numbers, descriptions, and weights.
- Clear Chamber (#3421901): replacement clear chamber with latches, 2 lb (0.9 kg).
- Heating Pans (#3425202): package of 250 aluminum disposable pans, 1 lb (0.5 kg).
- Filter Cartridge (#3421801): replacement particulate/GAC filter, lasts 100 cycles, 5 lbs (2.3 kg).
- Exhaust Connection Kit (#3425400): allows remote exhaust or fume hood connection; includes 10 ft (1.2 m) extended, 4" (100 mm) diameter collapsible duct, 1 lb (0.5 kg).
- Replacement Hook (#3422202): for hanging evidence, 1 lb (0.5 kg).
- Replacement Shelf (#3422203): for supporting small evidence, 1 lb (0.5 kg).
- Perforated Shelf Cover (#3426101): stainless steel cover fits over Shelf #3426103, 1 lb (0.5 kg).
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Model 3420000 requires 115 VAC, 60 Hz, 1 Amp; models 34200-20, -22, -24 require 230 VAC, 50/60 Hz, 0.5 Amp. p.7 ↗ | Select model according to available mains voltage; use a circuit breaker rated 10 A for 115 V models or 5 A for 230 V models. | whole series |
| Power cords supplied are 3 meters (10 feet) long. p.7 ↗ | Locate the system within 3 m of a suitable outlet; use an extension cord only if adequately rated. | whole series |
| The system can be operated on an inverter converting 12 VDC to AC, with output rating at least 100 W and current overload protection. p.7 ↗ | Enables mobile operation from a vehicle 12 V supply; ensure inverter meets minimum rating. | whole series |
| CA boiling point is approximately 120°F (50°C); decomposition products appear above 250°F (121°C); hydrogen cyanide can be produced above 400°F (204°C). p.9 ↗ | Never exceed 350°F (177°C) setpoint; the redundant safety switch limits heater to 350°F (177°C). | whole series |
| Controller setpoint range is 120°F (50°C) to 350°F (177°C). p.9 ↗ | Program setpoint within this range; temperatures stabilize within 3-5 minutes at setpoint ±15°F. | whole series |
| Clear chamber must be treated with Rain-X or Rain Block; protection lasts up to 20 uses if undisturbed. p.9 ↗ | Reapply treatment after every 20 development cycles or whenever chamber interior is cleaned. | whole series |
| First filter stage is MERV 11, removing 65-79% of 1.0-3.0 µm particles and >85% of 3.0-10 µm particles. p.9 ↗ | Provides particulate filtration; replace filter cartridge after 100 development cycles or when odor is detected. | whole series |
| Filter life is a minimum of 100 fuming cycles using about 2 oz (57 ml) of CA. p.9 ↗ | Plan filter replacement after 100 cycles; part number 3421801. | whole series |
| Never exceed 8 drops of CA per development cycle; typical procedure uses 4-6 drops. p.11 ↗ | Exceeding 8 drops may overload filter and produce excessive fumes; use 4 drops for typical 3-5 minute development at 350°F (177°C). | whole series |
| Optimum relative humidity for fingerprint development is 70-90%, best at approximately 80%; one drop of water raises RH by approximately 2%. p.20 ↗ | Use the humidity chart to add water drops: e.g., to raise RH by 30%, add 15 drops. | whole series |
| Typical development time is 3-5 minutes with 4 drops of CA at 350°F (177°C). p.21 ↗ | Monitor development and switch to Vent when prints are developed as desired. | whole series |
| After fuming, let chamber vent for at least 3-5 minutes before opening. p.22 ↗ | Ensures CA fumes are removed before retrieving evidence. | whole series |
| System weight is approximately 22 lbs (10 kg). p.11 ↗ | Follow safe-lifting guidelines when transporting; use transport mode with heater/filter module nested and transport lid locked. | whole series |
| Clear chamber interior dimensions: 16" (406 mm) H x 8" (203 mm) W x 8" (203 mm) D. p.28 ↗ | Limits evidence size; ensure evidence fits within chamber. | whole series |
| Environmental limits: indoor use, max altitude 6562 ft (2000 m), ambient 41-104°F (5-40°C), max RH 80% up to 88°F (31°C) decreasing to 50% at 104°F (40°C). p.29 ↗ | Install in a location meeting these conditions; voltage fluctuations not to exceed ±10%. | whole series |
| Exhaust Connection Kit (#3425400) includes 10 ft (1.2 m) extended, 4" (100 mm) diameter collapsible duct; do not add additional duct. p.6 ↗ | Additional duct reduces exhaust flow and may cause incomplete venting. | whole series |
| Heating pans should be discarded after the second use; do not use pans with dented or warped bottoms. p.16 ↗ | Residue buildup affects heat transfer; damaged pans heat CA inefficiently. | whole series |
| Store opened CA glue in a freezer or airtight container with silica gel to prevent setting. p.16 ↗ | Extends usable life of CA by limiting moisture exposure. | whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Blower and controller won’t turn on p.26 ↗ | Unit not plugged into outlet | Plug the cabinet into appropriate electrical service. | |
| Blower and controller won’t turn on p.26 ↗ | Circuit breaker tripped | Reset circuit breaker. | |
| Prints are not developing properly p.26 ↗ | Insufficient humidity in the clear chamber | Place 1-2 drops of water in a heating pan, and place on the heater. | |
| Little or no fumes are being produced p.26 ↗ | Heater temperature is too low. | Increase heater temperature | |
| Little or no fumes are being produced p.26 ↗ | Bottom of heater pan is dented/bent | Discard damaged pan, and use one with flat bottom. | |
| Heater temperature not changing on controller (“out” indicator is on) p.26 ↗ | Heater /Over temperature switch failure | Contact Labconco Product Service Department. | |
| Heater temperature not changing on controller (out indicator is off) p.26 ↗ | Controller is defective | Contact Labconco Product Service Department. | |
| Fan does not run p.26 ↗ | Fan or 12 volt power supply failure | Contact Labconco Product Service Department. | |
| Controller displays “Er” or “oo” or “-” p.26 ↗ | Controller error | Contact Labconco Product Service Department. |