3938201 Rev D Purifier Filtered PCR Enclosures

Technical Reference Summary

Covered Models & Document Scope

Labconco Purifier® Filtered PCR Enclosures in 2', 3' and 4' widths, 115V and 230V models, with optional Guardian™ Airflow Monitor and optional UV light. Ordering codes listed: 3970202, 3970205, 3970222, 3970225, 3970302, 3970305, 3970322, 3970325, 3970402, 3970405, 3970422, 3970425.

Section-by-Section Technical Breakdown

pp. 9, 12, 13
Support, Vibration and Movement Requirements

The enclosure must be installed on a level, solid supporting structure, typically a base cabinet or base stand with a chemically-resistant work surface. The work surface must be at least as wide and deep as the enclosure and able to support the combined weight of the enclosure and related equipment.

  • Supporting structure: base cabinet or base stand plus chemically-resistant work surface.
  • Work surface must be smooth, rigid, durable, non-porous and resistant to materials used with the enclosure.
  • Work surface should include a dished recessed area for primary spill containment.
  • Enclosure weight is 125 to 195 lbs (55 to 85 kg) depending on model.
pp. 10
Location and Air Current Requirements

The enclosure should be placed away from air disturbances that could disrupt ISO Class 5 airflow. The manual lists high-traffic areas, HVAC diffusers, fans, radiators, doorways and openable windows as locations to avoid.

  • Avoid high-traffic areas where walking causes air disturbance.
  • Avoid overhead or wall HVAC diffusers, fans, radiators and other air-current-producing lab equipment.
  • Avoid doorways or windows that may be opened.
  • Do not position the unit so the main disconnect device is difficult to operate.
pp. 10
Airflow Specifications

An integral motorized impeller draws room air from the top and pushes it down through the HEPA filter. The enclosure is factory set to deliver 40-55 fpm downflow velocity measured 6 inches below the upper diffuser, with model-specific exhaust volume, noise and outflow face velocity data.

  • Factory downflow velocity setting: 40-55 fpm measured 6" below the upper diffuser.
  • 2' model: 30/40/50/60 fpm downflow gives 120/165/210/250 CFM exhaust, 53-63 dB(A), 85/115/145/170 fpm face velocity.
  • 3' model: 30/40/50/60 fpm downflow gives 180/250/310/370 CFM exhaust, 58-70 dB(A), 85/115/145/170 fpm face velocity.
  • 4' model: 30/40/50/60 fpm downflow gives 240/330/420/500 CFM exhaust, 58-70 dB(A), 85/115/145/170 fpm face velocity.
pp. 11
Electrical Requirements

Standard duplex electrical receptacles should be nearby. The enclosure includes iris pass-throughs for routing cords through the back without leaving a large opening. Labconco does not recommend plugging the enclosure into a GFCI outlet.

  • Standard duplex electrical receptacles should be nearby.
  • Iris pass-throughs allow electrical cords through the back of the enclosure.
  • Labconco does not recommend plugging the Filtered PCR Enclosure into a ground fault circuit interrupter (GFCI) outlet.
pp. 13, 14
Installing the Enclosure on a Supporting Structure and Work Surface

The enclosure is placed on a level work surface and secured with four mounting screws. The front of the enclosure should be aligned within 0.30 inch (8 mm) of the front of the work surface, and the work surface is secured to the base cabinet or stand with structural adhesive or silicone sealant.

  • Level the base cabinet or stand and the work surface before installation.
  • Secure the work surface to the base cabinet or stand with structural adhesive or silicone sealant.
  • Insert the supplied mounting screws in the four holes, allowing a minimum of 1/8" clearance under the screw head.
  • Align the front of the enclosure within 0.30" (8 mm) of the front of the work surface.
  • Slide the rear flange and front air foil flanges under the mounting screw heads and tighten the four screws.
pp. 15
Verifying HEPA Filter Integrity

The HEPA filter is shipped installed with the gasket on the downstream side and is leak checked at Labconco. A second leak check is recommended before use and at least annually thereafter. Replacement HEPA filter part numbers are given for each width.

  • HEPA filter is shipped installed with the gasket on the downstream side.
  • A second leak check is recommended before using the enclosure and at least annually thereafter.
  • All seams downstream of the HEPA filter are jacketed by positive pressure.
  • 2' HEPA filter part no. 3707900; 3' part no. 3707901; 4' part no. 3707902.
  • HEPA filters are rated 99.99% efficient for 0.3-micron particles.
pp. 15
Connecting the Electrical Supply Source

115V models use the supplied power cord connected to the IEC inlet on the back of the enclosure; at 50 Hz operation maximum airflow is reduced by 17%. 230V models are shipped without a plug and require a plug installed per local codes.

  • 115V models: connect the supplied 115V power cord to the IEC electrical supply plug on the back of the enclosure.
  • At 50 Hz operation, blower maximum airflow is reduced by 17%.
  • 230V models are shipped without a plug; install an appropriate plug per local electrical codes.
  • Do not use any detachable power cord that is not adequately rated for the unit.
pp. 16
Setting the Downflow Velocity with the Speed Control

The speed control behind the front panel adjusts downflow velocity. Class 100 (ISO Class 5) conditions require 30-65 fpm; the enclosure is factory set at 40-55 fpm. Measurements are taken with an anemometer 6 inches below the interior upper diffuser.

  • Downflow velocity range for Class 100 (ISO Class 5): 30-65 fpm.
  • Factory setting: 40-55 fpm downflow velocity.
  • Measure downflow velocity with an anemometer 6" below the interior upper diffuser.
  • Turn the speed control counterclockwise to increase downflow velocity and clockwise to decrease it.
  • Working at the lowest appropriate downflow velocity gives the quietest operation.
pp. 16, 17
Validating the Filtered PCR Enclosure

Validation is performed by averaging anemometer readings on a model-specific downflow test grid located 6 inches below the upper diffuser. The 2', 3' and 4' models use 9, 12 and 15 test points respectively, arranged in three rows.

  • 2' model: 9 test points (3 points in 3 rows); side columns 6" from sidewalls, center column 5.32" from first/last column.
  • 3' model: 12 test points (4 points in 3 rows); side columns 6" from sidewalls, two inner columns 7.54" from first/last column.
  • 4' model: 15 test points (5 points in 3 rows); side columns 6" from sidewalls, three inner columns 8.66" from first/last column.
  • Back row of the grid is 6" from the rear of the work area; middle and front rows are 6.57" apart from back to front.
  • A certified technician should complete validation.
pp. 17
Sealing the Enclosure to the Work Surface

After the enclosure is set in place and wired, a bead of silicone sealant is recommended to seal the enclosure to the work surface and prevent materials from collecting under the walls.

  • Seal the enclosure to the work surface after it is set in place and wired.
  • A bead of silicone sealant is recommended.
pp. 18, 19, 20, 21, 22, 23
Performance Features

The enclosure provides Class 5 air conditions at downflow velocities of 30-65 fpm. Key features include a 10-degree ergonomic slope, 25-inch internal depth, safety glass sash with spring-loaded latch, iris utility ports, optional Guardian Airflow Monitor, inherently safe impeller, speed control, filter pressure gauge, HEPA filter, variable digital timed UV lamp and UV closure panel.

  • Class 5 air conditions are delivered at downflow velocities of 30 to 65 fpm.
  • Internal depth is 25 inches; ergonomic slope is 10 degrees.
  • Safety glass sash has a wiping seal and spring-loaded latch; sash must be down for normal operation.
  • Optional Guardian™ Airflow Monitor provides audio/visual alarm for low airflow.
  • Variable digital timed UV lamp has eight settings: 5, 10, 15, 30, 60, 120, 240 minutes and infinity.
  • UV closure panel must be in place for the UV lamp to work.
  • Prefilter should be replaced quarterly.
pp. 25, 26
Routine Daily Work Procedures

Daily operation includes planning to minimize disruptions, starting the blower and light, allowing one minute of purge before work, loading only required materials, working toward the rear of the work area, and purging for two to three minutes before shutdown.

  • Turn on blower and light; turn off UV light and remove UV closure panel.
  • Allow blower to operate unobstructed for 1 minute before work.
  • After loading, wait one minute to purge airborne contaminants.
  • Upon completion, allow enclosure to operate 2-3 minutes undisturbed before shutting down blower.
  • Wipe interior surfaces with mild household detergent or disinfectant; do not use abrasive cleaners, bleach or solvents.
pp. 26, 27
HEPA Filter Applications, Suitability and Guidelines

The enclosure is suitable for procedures requiring a clean Class 5 air environment and product protection from particulate matter. HEPA filters only entrap particulate matter and are fragile; damaged filters require the enclosure to be taken out of service.

  • Suitable for procedures traditionally performed on an open bench that now require a clean Class 5 air environment.
  • Provides product protection from particulate matter only.
  • HEPA filters are only effective for entrapment of particulate matter.
  • Do not touch the fragile HEPA filter surface; avoid puncturing during maintenance.
  • If a HEPA filter is suspected damaged, do not use the enclosure; contact a certification agency or Labconco.
pp. 23, 29
Routine Maintenance Schedule

Weekly maintenance includes wiping interior surfaces with disinfectant or cleaner and cleaning exterior surfaces, particularly the front and top. The prefilter should be replaced quarterly.

  • Weekly: wipe down interior surfaces with disinfectant or cleaner and allow to dry.
  • Weekly: clean exterior surfaces, particularly the front and top, with a damp cloth.
  • Prefilter should be replaced quarterly.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Factory downflow velocity setting is 40-55 fpm measured 6 inches below the upper diffuser. p.10 ↗ Verify actual downflow with an anemometer at the specified measurement plane before putting the enclosure into service. All Purifier Filtered PCR Enclosure models
Class 100 (ISO Class 5) conditions require downflow velocity of 30-65 fpm. p.16 ↗ The speed control must be adjusted so the average downflow velocity remains within this range for the application. All models
HEPA filters are rated at least 99.99% efficient for 0.3-micron particles. p.5 ↗ The enclosure meets ISO Class 5 / Class 100 particulate cleanliness when airflow is correctly set and the filter is intact. All models
HEPA filter part numbers: 2' = 3707900, 3' = 3707901, 4' = 3707902. p.15 ↗ Order the correct replacement filter by enclosure width. 2', 3' and 4' models respectively
At 50 Hz operation, blower maximum airflow is reduced by 17%. p.15 ↗ Account for reduced airflow when sizing or validating 115V units on 50 Hz power. 115V models
230V models are shipped without a plug and require an appropriate plug installed per local codes. p.15 ↗ Plan for a qualified electrician to fit the plug before commissioning. 230V models
Labconco does not recommend plugging the enclosure into a ground fault circuit interrupter (GFCI) outlet. p.11 ↗ Provide a standard duplex receptacle rather than a GFCI-protected outlet for the enclosure. All models
Enclosure weight is 125 to 195 lbs (55 to 85 kg) depending on model. p.12 ↗ The supporting structure and work surface must safely support the enclosure plus related equipment. All models
The front of the enclosure should be aligned within 0.30 inch (8 mm) of the front of the work surface. p.13 ↗ Position the enclosure accurately on the work surface before tightening the four mounting screws. All models
Mounting screws require a minimum of 1/8 inch clearance under the head for positioning the enclosure. p.14 ↗ Leave the screws loose enough to slide the rear and front air foil flanges under the heads before final tightening. All models
Downflow velocity is measured 6 inches below the interior upper diffuser using an anemometer. p.16 ↗ Use the specified measurement plane and a calibrated anemometer for validation and routine checks. All models
Downflow test grid: 2' model has 9 points (3x3), 3' model has 12 points (4x3), 4' model has 15 points (5x3). p.16 ↗ Average readings across the correct grid for the model to determine average downflow velocity. 2', 3' and 4' models respectively
The variable digital timed UV lamp has eight settings: 5, 10, 15, 30, 60, 120, 240 minutes and infinity. p.22 ↗ Select the UV decontamination time appropriate to the PCR protocol. Models with UV light option
The UV closure panel must be in place for the UV lamp to work. p.23 ↗ Install the UV closure panel before turning on the UV light; the magnetic switch prevents operation without it. Models with UV light option
The prefilter should be replaced quarterly. p.23 ↗ Include prefilter replacement in the quarterly maintenance schedule to extend HEPA filter life. All models
With normal usage, HEPA filters last two to five years before requiring replacement. p.24 ↗ Plan HEPA filter replacement based on filter pressure gauge readings and annual certification results. All models
A second HEPA filter leak check is recommended before use and at least annually thereafter. p.15 ↗ Schedule initial and annual leak testing in addition to the factory leak check. All models
The enclosure should be validated annually and whenever it is relocated. p.23 ↗ Re-certify after any move or significant change in location or airflow. All models
Allow the blower to operate unobstructed for 1 minute before work and 2-3 minutes after work before shutdown. p.25 ↗ Include purge times in the daily operating procedure to maintain Class 5 conditions. All models
The sash must remain in the down position while using the enclosure. p.23 ↗ Operate only with the sash closed; raise it only for loading and cleaning. All models

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Class 100 (ISO Class 5) air conditions are not present inside the enclosure. p.44 ↗ Improper user techniques for the enclosure. See “Certifying the Enclosure” Chapter 3 and “Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix D). Chapter 8
Class 100 (ISO Class 5) air conditions are not present inside the enclosure. p.44 ↗ External factors are disrupting the enclosure airflow patterns. See “Location Requirements” Chapter 2, “Certifying the Enclosure” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix D). Chapter 8
Class 100 (ISO Class 5) air conditions are not present inside the enclosure. p.44 ↗ Enclosure has improper downflow velocity. Have enclosure certified and check exhaust system. Check HEPA filters for loading. Adjust enclosure speed control. Enclosure should have an average downflow velocity of 30-65 fpm. Chapter 8
Blower won’t operate. p.44 ↗ Unit not plugged into outlet. Plug the enclosure into appropriate electrical service. Chapter 8
Blower won’t operate. p.45 ↗ Circuit breaker(s) or Ground Fault Interrupter. Reset circuit breaker. Chapter 8
Blower won’t operate. p.45 ↗ Blower wiring is disconnected. Inspect blower wiring. Chapter 8
Blower won’t operate. p.45 ↗ Blower switch is defective. Replace switch. Chapter 8
Blower won’t operate. p.45 ↗ Motorized impeller or blower is defective. Replace motorized impeller or blower. See Chapter 6. Chapter 8
Poor internal Class 5 air conditions. p.45 ↗ Enclosure sash not closed. Close sash. Chapter 8
Poor internal Class 5 air conditions. p.45 ↗ HEPA filter clogged. Replace HEPA filter or increase speed. Chapter 8
Blower and lights won’t operate. p.45 ↗ Unit not plugged into outlet. Plug enclosure into appropriate electrical service. Chapter 8
Blower and lights won’t operate. p.45 ↗ Circuit breaker(s) tripped. Reset or replace circuit breaker. Chapter 8
Light does not work. p.45 ↗ Lamp not installed properly. Inspect lamp installation. Chapter 8
Light does not work. p.45 ↗ Lamp wiring disconnected. Inspect lamp wiring. Chapter 8
Light does not work. p.45 ↗ Defective lamp. Replace lamp. Chapter 8
Light does not work. p.45 ↗ Light switch is defective. Replace light switch. Chapter 8
Light does not work. p.45 ↗ Defective electronic ballast. Replace ballast. Chapter 8
Light does not work. p.45 ↗ UV closure panel not installed. Install UV closure panel (UV lights only). Chapter 8
Light does not work. p.45 ↗ Magnetic switch for UV closure panel defective. Replace switch (UV lights only). Chapter 8
Light does not work. p.45 ↗ UV timer circuit board defective. Replace circuit board (UV lights only). Chapter 8
UV light timer does not work p.45 ↗ No power. No lights. No display Power cable to UV light timer circuit board is disconnected. Verify connection. Replace circuit board for UV light timer if it malfunctions. Chapter 8
Airflow monitor malfunction p.45 ↗ No power. No lights. No display. Power cable to circuit board is disconnected and needs to be connected. Sensor cable needs to be connected. Verify that all airflow monitor interface cables are connected. Check fuses on enclosure. Chapter 8
Airflow monitor malfunction p.45 ↗ No audible alarm. Alarm has been temporarily silenced using “SILENCE ALARM.” Chapter 8
Airflow monitor malfunction p.46 ↗ Wrong alarm set point. Airflow monitor was not properly adjusted. Repeat calibration steps outlined in this manual in Chapter 6. Chapter 8
Airflow monitor malfunction p.46 ↗ Constant audible alarm. Check airflow and calibration of airflow monitor. See Chapter 6. Chapter 8
Airflow monitor malfunction p.46 ↗ Continuous alarm. Check the downflow velocity of the enclosure as the airflow of the system may have changed. If incorrect, adjust the speed control to increase downflow velocity. The HEPA filter may have become loaded. If downflow velocity is correct, calibrate the airflow monitor outlined in this manual in Chapter 6. Chapter 8
Airflow monitor malfunction p.46 ↗ Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. Chapter 8
Airflow monitor malfunction p.46 ↗ Audible disable will not stay operational. An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower. Chapter 8