Technical Reference Summary

Covered Models & Document Scope

This manual covers Labconco Hooded Combination (Macro) Kjeldahl Nitrogen Systems, models 21176, 21177 and 21178 Series, including the 6-place, 12-place and 18-place configurations. Electrical variants are identified by model suffix: -11 (115 V, single-phase, 60 Hz), -12 (230 V, single-phase, 60 Hz) and -13 (208/230 V, three-phase, 60 Hz). The document also covers the 12-place system 2117703 and replacement parts for the heater, blower, manifold and glassware assemblies.

Section-by-Section Technical Breakdown

pp. 4-8
Components and Component Identification

The system is shipped in one crate with two components that must be assembled: the top exhaust bonnet and the main body, plus the interior light electrical connection. The main functional components are 600 W electric heaters with infinite controls, a CPVC lower fume removal manifold with PTFE nipples, an acid fume piping exhaust system, a distillation condensation manifold with 3/4" NPT water connections, and two hood duct collars (exhaust and make-up air).

  • 600 W heaters are used in both the lower digestion rack and upper distillation rack; infinite control switches regulate each heater.
  • The lower fume removal manifold is CPVC with heat-resistant PTFE nipples that extend into the flask opening.
  • Flasks are oriented at 45 degrees to the fume removal manifold to promote condensation and refluxing during the 45+ minute digestion.
  • The 6" exhaust stack on the left side must connect to leak-tight PVC ductwork venting 10 ft (3 m) above the building roof.
  • The distillation condensation manifold has six columns per manifold with 3/4" NPT tap water inlet and outlet connections.
  • The hood has two duct collars: exhaust collar to the right of the make-up air collar; the make-up air collar can supply up to 75% of total exhaust air volume.
pp. 9-16
Installation

Installation covers uncrating, bonnet and sash assembly, electrical connection, hood exhaust and auxiliary air systems, electrical supply, tap water supply, distillation manifold plumbing, acid fume blower exhaust, airflow verification, drip shield and glassware installation. The system is shipped in two sections and is heavy; it cannot be disassembled to reduce size.

  • The bonnet is held in place by its weight on a foam gasket seal; bolting is not required.
  • Glass sashes are installed by inserting the upper edge into its track first, then positioning the lower edge over the nylon runners.
  • The hood exhaust and auxiliary air connections are 10-3/4" I.D. duct connections.
  • The acid fume exhaust duct connection on top of the blower is sized for 6" nominal (6-5/8" O.D.) vent duct.
  • The exhaust ductwork must be supported independently of the Kjeldahl system to avoid distortion of the blower housing.
  • The digestion rack drip shield is stainless steel and has a drip tube on the left side for draining acid spills.
pp. 11
Electrical Requirements

The systems are factory-wired per the model number ordered. Model suffixes define voltage: -11 is 115 V single-phase, -12 is 230 V single-phase, and -13 is 208/230 V three-phase. Amperage depends on the number of heaters (6-PL, 12-PL, 18-PL). A qualified electrician must connect the facility supply to the main breaker box, provide an earth ground, and verify connections before power-up.

  • -11: 115 V, single-phase, 60 Hz, 65 A for 6-PL.
  • -12: 230 V, single-phase, 60 Hz, 35 A for 6-PL, 66 A for 12-PL, 96 A for 18-PL.
  • -13: 208/230 V, three-phase, 60 Hz, 20 A for 6-PL, 38 A for 12-PL, 56 A for 18-PL.
  • A remote blower on/off switch can be wired to control the remote blower from the unit; the remote blower MUST be on before using the unit.
  • An earth ground must be provided and all wiring must conform to local electrical codes.
pp. 12
Tap Water Supply and Distillation Manifold

The distillation rack uses individual condensation columns connected by a common manifold (6 columns per manifold). Tap water inlet and outlet connections are 3/4" NPT, and the supply line should be minimum 1/2" ID to provide 6-8 gallons (23-30 liters) per minute. Systems with 12 or 18 heaters include two or three 6-column manifolds connected in series. The outlet must be free of back pressure.

  • All plumbing connections and components must be free of foreign material before final connections.
  • Supply line minimum 1/2" ID; flow 6-8 gal/min (23-30 L/min).
  • 12- and 18-heater systems use two or three 6-column manifolds connected in series.
  • The distillation manifold outlet must be free of back pressure to avoid flow restriction.
  • Water discharged from the distillation manifold is not acid contaminated and may use standard drain lines.
pp. 13-14
Acid Fume Blower Exhaust System

The acid fume exhaust system uses a remote blower connected to 6" nominal (6-5/8" O.D.) acid-resistant ductwork. The exhaust piping assembly includes a short 6" PVC pipe and a flexible coupling that must be sealed air-tight. A condensate drain at the bottom of the piping assembly collects concentrated acid and must be handled as acid waste. The blower airflow requirement is 25 CFM at a measurement point 48 inches above the flexible coupling.

  • Exhaust duct connection on top of the blower is for 6" nominal (6-5/8" O.D.) vent duct.
  • All exhaust ductwork must be acid-resistant and sealed air-tight; use only acid-resistant sealant on joints.
  • Condensate from the piping assembly is concentrated acid; collect in a closed acid-proof container or plumb to an acid-resistant drain.
  • Blower airflow requirement: 25 CFM; average airflow velocity 117 feet per minute.
  • Airflow test uses the Pitot Traverse in Inlet Duct method per ANSI/ASHRAE 51-1999, with a 6-point duct traverse at 48 inches from the blower.
  • If the remote blower pulls more than 25 CFM at the measurement point, close the damper to reduce airflow to 25 CFM.
pp. 17
Normal Operation Start-Up

Start-up covers flask sizing, preheating, starting the acid fume removal blower or water aspirator, verifying manifold suction with tissue paper, starting distillation cooling water, and rinsing the condenser tubes. The apparatus accepts 500 or 800 ml Kjeldahl flasks, and the digestion heaters move on runways to accommodate either size.

  • The apparatus is designed for 500 or 800 ml Kjeldahl flasks; digestion heaters move forward or backward on the runway.
  • Turn heat controllers on both digester and distillation racks and wait approximately 5 to 10 minutes for preheating.
  • Start the acid fume removal blower or water aspirator; suction should hold a 2" x 2" (51 mm x 51 mm) tissue paper over the digestion manifold nipple.
  • Distillation cooling water temperature should not exceed 80°F (27°C); regulate flow with the control valve and watch the thermometer.
  • Rinse the inner stainless steel condensing tubes by boiling nitrogen-free water in Kjeldahl flasks attached to the condensation columns before distillation.
pp. 18-19
Clean-Up and Routine Maintenance

Routine maintenance includes cleaning with weak sodium hydroxide solution, neutralizing sulfate buildup between the channel support and fume removal manifold, inspecting the remote blower annually, and clearing obstructions in the exhaust system by boiling water in flasks to draw steam through the fume pipe. The drip shield requires prompt acid spill wash-down and periodic stainless steel cleaning.

  • Access the blower exhaust or water aspirator system by removing the lower left front access panel below the electrical panel.
  • Wash equipment with a weak sodium hydroxide solution and rinse with clear water.
  • Neutralize sulfate buildup between the channel support and fume removal manifold with weak sodium hydroxide periodically.
  • Inspect the remote blower annually for debris, proper operation, and adequate airflow as specified on pages 11-12.
  • Clear exhaust system obstructions by boiling water in flasks on the digestion heaters with the remote blower on to draw steam through the fume pipe.
  • For the drip shield: promptly wash down major acid spills, clean with residential stainless steel sink cleaner, and renew appearance by lightly sanding with 220 grade emery paper.
pp. 20
Replacement Parts

Appendix A lists replacement parts for electrical heaters, motor/blower components, and miscellaneous items. Key parts include infinite heater controls for 115 V and 230 V, 600 W heater elements in 115 V and 230 V versions, a 6-place stainless steel distillation manifold, PTFE nipples, ceramic nipples, and a replacement fume pipe kit.

  • Heater controls: 1315400 (115 V infinite), 1315500 (230 V infinite).
  • Heater elements: 2033100 (115 V, 600 W, package of 6), 2033200 (230 V, 600 W, package of 6).
  • Distillation manifold: 2152402 (6-place stainless steel).
  • PTFE nipple 2165800 with plastic screws 1880128; ceramic nipple 2056600.
  • Replacement fume pipe kit 2078800; connector bulb/caustic trap 2081300 (package of 6); delivery tube 2128800 (package of 6).
  • Inlet and outlet manifold assembly 2146400 for 12-unit only.
pp. 22-26
Total Kjeldahl Nitrogen Method (Appendix C)

Appendix C provides a titrimetric Total Kjeldahl Nitrogen method covering scope, apparatus, reagents, digestion, distillation, titration and calculation. The method uses 500-800 ml Macro flasks, 600 W burners, sulfuric acid digestion with copper sulfate catalyst and potassium sulfate, sodium hydroxide distillation, boric acid receiving solution, and standardized sulfuric acid titration.

  • Sample size 0.250-1.000 g weighed to nearest 0.1 mg; add 20 ml H2SO4, 0.04 g CuSO4, 15 g K2SO4, 8-10 boiling stones.
  • Digestion total time approximately 1 hour; continue boiling for an equal length of time after clearing.
  • Distillation: add 100 ml of 45% NaOH without mixing, connect flask immediately, distill >8-10 minutes until all NH3 transferred.
  • Titrate with 0.02 N or 0.10 N H2SO4; titrations should require at least 15 ml of titrant to be accurate.
  • Calculation: %N = (ml H2SO4 sample - ml H2SO4 blank) x normality x 1.4007 / weight in grams.
  • Protein factors: 5.7 bread/wheat flour, 6.25 other grains, 6.38 milk, 6.25 unknown source.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Heaters are 600 W in both the lower digestion rack and upper distillation rack, regulated by infinite control switches. p.6 ↗ Electrical supply must be sized for the total heater load; see amperage table for 6-, 12- and 18-place configurations. All models
Model suffix -11 is 115 V single-phase 60 Hz; -12 is 230 V single-phase 60 Hz; -13 is 208/230 V three-phase 60 Hz. p.11 ↗ Verify facility voltage and phase match the model suffix before ordering or installation. All models
Amperage requirements: -11 6-PL 65 A; -12 6-PL 35 A, 12-PL 66 A, 18-PL 96 A; -13 6-PL 20 A, 12-PL 38 A, 18-PL 56 A. p.11 ↗ Size circuit breakers and supply wiring accordingly; a qualified electrician must connect the facility supply. All models
The remote blower MUST be on before using the unit; a remote blower on/off switch can be wired to control it from the unit. p.11 ↗ Interlock or operating procedure must ensure blower operation before heaters are used. All models
Distillation cooling water temperature should not exceed 80°F (27°C); for ground water above 80°F a recirculation water chiller system is required. p.7 ↗ Facilities with warm ground water must provide chilled recirculation to achieve sufficient condensation. All models
Tap water connections are 3/4" NPT; supply line minimum 1/2" ID; flow 6-8 gallons (23-30 liters) per minute. p.12 ↗ Plumbing must be sized to deliver this flow with no back pressure on the outlet. All models
The acid fume exhaust stack is 6" and must connect to leak-tight PVC ductwork venting 10 ft (3 meters) above the building roof. p.7 ↗ Duct routing and roof penetration must meet this height and material requirement. All models
Blower airflow requirement is 25 CFM at a measurement point 48 inches above the flexible coupling; average velocity 117 feet per minute. p.13 ↗ After installation, verify airflow using the Pitot traverse method and adjust the damper if above 25 CFM. All models
The exhaust duct connection on top of the blower is sized for 6" nominal (6-5/8" O.D.) vent duct; the flexible coupling must be sealed air-tight. p.13 ↗ Use compatible 6" acid-resistant duct and seal all joints with acid-resistant sealant. All models
The hood exhaust and auxiliary air connections are 10-3/4" I.D. duct connections; the make-up air collar can supply up to 75% of total exhaust air volume. p.10 ↗ Auxiliary air duct sizing should not exceed 75% of the hood exhaust volume. All models
The acid fume removal exhaust ductwork must not be connected to the hood heat exhaust ductwork; each must be routed independently to the building exterior. p.7 ↗ Two separate duct systems are required; combining them is prohibited. All models
The apparatus accepts 500 or 800 ml Kjeldahl flasks; digestion heaters move on runways to accommodate either size. p.17 ↗ Flask size selection is flexible within this range without changing heater hardware. All models
Suction at the digestion manifold should hold a 2" x 2" (51 mm x 51 mm) tissue paper over the nipple. p.17 ↗ This is the field check for adequate fume removal suction before digestion. All models
Preheat time is approximately 5 to 10 minutes before use. p.17 ↗ Allow this warm-up before placing samples or starting distillation. All models
The digestion process typically runs 45+ minutes; flasks are oriented at 45 degrees to the fume removal manifold to promote refluxing and prevent boiling dry. p.6 ↗ Flask orientation and manifold position are critical to prevent sample loss. All models
The remote blower should be inspected annually for debris, proper operation, and adequate airflow. p.18 ↗ Include blower inspection in the annual maintenance schedule. All models
Exhaust system obstructions can usually be cleared by boiling water in flasks on the digestion heaters with the remote blower on to draw steam through the fume pipe. p.18 ↗ This is the recommended field remedy for reduced nipple suction. All models
The drip shield is stainless steel; clean periodically with residential stainless steel sink cleaner and renew appearance by lightly sanding with 220 grade emery paper. p.19 ↗ Prompt spill wash-down prevents discoloration and marring. All models
The method uses 0.250-1.000 g sample, 20 ml H2SO4, 0.04 g CuSO4, 15 g K2SO4, 8-10 boiling stones; digestion total time approximately 1 hour. p.23 ↗ These are the standard reagent quantities and timing for the Macro method. All models
Titration should require at least 15 ml of titrant to be accurate; use 0.02 N or 0.10 N standardized H2SO4 depending on expected nitrogen level. p.25 ↗ Select titrant normality so that titration volume is at least 15 ml. All models

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Suction in the nipples is reduced p.18 ↗ Obstruction within the exhaust system Wash obstructions from the fume pipe by boiling water in a number of flasks placed on the digestion heaters with the remote blower on to draw steam through the fume pipe. Exhaust Systems Maintenance
Sulphate buildup between the channel support and the fume removal manifold p.18 ↗ Salt of sulfuric acid accumulating during use Neutralize with a weak solution of sodium hydroxide periodically to keep the unit clean. Clean-Up and Cosmetic Guidelines
Acid spills on the drip shield causing discoloration and eventual marring p.19 ↗ Acid attack on stainless steel surface Promptly wash down all major acid spills; periodically clean with residential stainless steel sink cleaning compound; renew appearance by lightly sanding with 220 grade emery paper. Drip Shield
Sample foaming up into neck of flask during digestion p.24 ↗ Vigorous digestion reaction Lower the burner heat setting to prevent foaming; once sample comes to a smooth rolling boil return heat setting to maximum. Appendix C, 5.2.2
Black specs in the flask neck during digestion p.24 ↗ Sample matrix not fully refluxed Rotate the flask 180 degrees to allow refluxing acid to rinse the internal flask surface. Appendix C, 5.2.2
K2SO4 salt solids forming after digestion p.24 ↗ Digestion mixture cooling without dilution Carefully add 50 ml of distilled water; all salt crystals must be dissolved before proceeding. If necessary partially reheat sample and agitate to dissolve crystals. Appendix C, 5.2.2