Rev m puricare open access station user manual
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco PuriCare Open Access Stations, catalog numbers 3820000 & -01 (115 VAC, 60 Hz), 3820010 & -11 (100 VAC, 50/60 Hz), and 3820020 through -38 (230 VAC, 50/60 Hz, with regional plug variants -21 US, -33 UK, -34 Schuko, -36 China/Australia, -38 India). Includes 4 FT and 5 FT models and optional accessories (side-mounted shelf, end shelf, handle kit).
Section-by-Section Technical Breakdown
Specifies site preparation for the Open Access Station, including placement away from air disturbances and minimum clearances. The station discharges HEPA-filtered exhaust air underneath the unit, requiring at least 12 inches clearance from any vertical obstruction on all sides.
- Locate away from traffic patterns, doors, fans, ventilation registers, fume hoods and other air-handling devices; keep room windows closed.
- Maintain at least 12 inches clearance between the station and any vertical obstruction on all sides.
- At least six inches (150 mm) overhead clearance from any overhead obstruction is required in final operating position.
Lists model-specific electrical supply requirements and dedicated circuit breaker ratings. A dedicated outlet with a 20 A breaker for 115 V models or 10 A breaker for 230 V models must be located within 25 ft (115 V) or 10 ft (230 V) of the station.
- 3820000 & -01: 115 VAC, 60 Hz, 12 Amps.
- 3820010 & -11: 100 VAC, 50/60 Hz, 12 Amps.
- 3820020 through -38: 230 VAC, 50/60 Hz, 6 Amps.
- Dedicated outlet with 20 A breaker for 115 V models within 25 feet; 10 A breaker for 230 V models within 10 feet.
- The station's own outlets are protected by a GFIC; Labconco does not recommend installing a GFIC on the outlet feeding the PuriCare.
Describes unpacking, moving, and verifying components of the station. The unit weighs approximately 450 lbs (205 kg) and must remain on its shipping skid until placed in its final location; it should be moved with a flat low dolly or floor jack, not tilted onto a hand truck.
- Station weight approximately 450 lbs (205 kg); carton allows lifting with mechanical lift truck or floor jack.
- Do not remove from shipping skid until ready to place in final location; do not move by tilting onto a hand truck.
- Verify components: User's Manual P/N 3792200 and one power cord (1334100 230V US, 1336100 230V Schuko, 1332600 230V UK, 1332700 230V China/Australia, 1345700 230V India).
- Assembly instructions P/N 1058400 are attached to the front panel.
Requires certification by a qualified certifier before initial use and at least annually or after service. The certifier must perform HEPA filter leak tests, downflow velocity profile test, inflow volume test, and airflow smoke patterns.
- Certify before initial use and recertify at least annually or whenever serviced.
- Required tests: HEPA Filter Leak Tests, Downflow Velocity Profile Test, Inflow Volume Test, Airflow Smoke Patterns.
- There is no tolerance on the downflow profile values.
Explains the two air-handling systems: twin motorized impellers draw room air through prefilters and force it through the upper supply HEPA filter, while a lower heavy-duty motor/blower draws air through lower prefilters and the exhaust HEPA filter. Speed controls are electronic circuits rated 10 A each, adjustable only by a qualified certifier.
- Twin motorized impellers handle supply air through the upper HEPA filter; a separate heavy-duty motor/blower handles exhaust air through the exhaust HEPA filter.
- Speed controls are rated for 10 Amps each, far in excess of normal motor current draw.
- HEPA filters are effective only against particulate matter; gases pass through.
- Never block or obstruct the grilles.
Covers sash operation, electric lift system, retractable power cord, and working technique. Both sashes must be fully closed for proper containment; the lift system is intermittent-duty and will only operate when the system power switch is ON.
- Both sashes must be completely closed for the PuriCare to contain properly.
- Lift system is for intermittent operation only; continuous operation for more than one out of each 10-minute interval may activate the thermal override. Allow at least 10 minutes cooling.
- Adjust height until the front edge of the air intake is even with the operator's beltline.
- Retractable power cord is 26 feet long with a safety brake; pull until ratcheting stops to release the brake.
- Keep all materials at least 4 inches inside of the sash and perform contaminated operations near the center of the work area.
- Allow the station to operate unobstructed for 5 minutes before work and 2 to 3 minutes after work for final purging.
Defines weekly, monthly, annual, and biannual maintenance tasks including disinfection, prefilter inspection/replacement, recertification, and fluorescent lamp replacement. Prefilters are disposable and cannot be cleaned.
- Weekly: disinfect interior and work surface with 70% ethanol or suitable disinfectant; clean sash with optical plastic cleaner; log pressure readings; lift work surface and inspect prefilters.
- Monthly: clean exterior surfaces; disinfect lower plenum; replace prefilters as needed.
- Annually: recertify by qualified certification technician.
- Biannually: replace fluorescent lamps.
- Prefilters may be contaminated with allergenic material; take appropriate precautions and dispose of them (do not clean).
Describes removal of work surface, grille, drain system access, prefilter removal, fluorescent lamp replacement, and circuit breaker reset. The station has a drain in the catch pan under each grille connected to a single drain line.
- Work surface removal: lift front edge and pull out.
- Grille removal: loosen two black thumbscrews, pull spring-loaded locking pins toward center, pivot front edge up ~1/4 inch, pull grille straight away.
- Drain line plug is on the rear panel; pull plug straight out, drain into container, reinstall.
- Upper and lower prefilters must be installed with airflow direction arrow pointing down.
- Fluorescent lamps: unplug station, swing up both sashes, rotate lamps 90 degrees and pull straight down.
- Circuit breakers are on top of unit near power cord outlet; depress white button until it sets.
Specifies downflow velocity setpoints, test grid patterns, and exhaust volume calculation method. Average downflow velocity is 55 ±5 FPM for 100/115 V models and 50-55 FPM for 230 V models.
- Models 3820000, -01, -10 & -11: average downflow velocity 55 ±5 FPM.
- Models 3820020-38: average downflow velocity 50-55 FPM.
- 5 FT model downflow test grid: 3 rows of 8 points (24 total); 4 FT model: 3 rows of 6 points (18 total).
- Exhaust volume calculation: 5 velocity readings along centerline of bottom diffuser, 2 inches below diffuser; average multiplied by open filter area (3.77 sq ft on 4 FT, 5.11 sq ft on 5 FT).
- Target exhaust volumes: 4 FT 100/115 V 509-528 CFM; 4 FT 230 V 434-452 CFM; 5 FT 100/115 V 639-664 CFM; 5 FT 230 V 537-562 CFM.
Describes adjustment of the two independent speed controls for supply impellers and exhaust motor/blower. Adjustment must be performed only by a qualified certification technician.
- Two plugged access holes on lower rear panel; left screw controls lower motor/blower, right screw controls upper impellers.
- Turn counterclockwise to increase speed, clockwise to decrease.
- Turning the screw fully counterclockwise shuts off the speed control circuit and stops the motor; turn clockwise until motor restarts.
- After adjustment, measure downflow velocities or inflow volume as needed.
Provides procedures for supply and exhaust HEPA filter replacement, which must be performed by a qualified certification technician. The station must be recertified after replacement.
- Supply HEPA filter: remove upper prefilters and front panel; loosen four locking bolts at least 1/2 inch; tighten four plenum lift bolts until ~1/4 inch clearance between filter top and plenum bottom; slide filter straight out.
- Exhaust HEPA filter: lower station to lowest position and unplug; remove lower front panel, work surface, lower prefilters, and front panel flange; loosen four locking bolts at least 1/2 inch; tighten lift bolts for ~1/4 inch clearance; remove two filter retainers and slide filter out.
- Replacement filter must match original performance specifications, particularly efficiency and flow rate at a given pressure drop.
- Station must be recertified before use after any HEPA filter replacement.
States that blower motor and impeller bearings are sealed and require no lubrication. Service should be performed by a qualified certification technician.
- Blower motor and impeller bearings are sealed; DO NOT attempt to lubricate them.
- Motor/blower should be serviced by a qualified certification technician.
Covers installation of optional side-mounted shelf, end shelf, and handle kit.
- Installing a side mounted shelf.
- Installing an end shelf.
- Installing the handle kit.
Provides a table of problems, probable causes, and corrective actions for the Open Access Station. Covers electrical, blower, lamp, pressure, and contamination issues.
- Problems include station blower and lights not turning on, blower not turning on but lights working, lights not working but blower on, slight increase in pressure reading, and contamination of work.
- Corrective actions include checking power, resetting circuit breakers, inspecting wiring, replacing motors/lamps/ballasts, checking prefilters and grilles, and recertifying the station.
Appendices provide component diagrams, dimensions, electrical data, impeller and motor specifications, accessory part numbers, and a quick chart of operating specifications.
- Appendix A: PuriCare Components (labeled diagrams).
- Appendix B: Open Access Station Dimensions.
- Appendix C: Specifications including electrical data, impeller specifications, and motor specifications.
- Appendix D: Accessories including Shelf Kit and Handle Kit.
- Appendix E: Quick Chart for operating specifications.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Model 3820000 & -01 requires 115 VAC, 60 Hz, 12 Amps; models 3820010 & -11 require 100 VAC, 50/60 Hz, 12 Amps; models 3820020 through -38 require 230 VAC, 50/60 Hz, 6 Amps. p.9 ↗ | Verify site electrical supply matches the model before installation; use the correct regional plug variant. | All models |
| A dedicated outlet with a circuit breaker rated at 20 Amps for 115 V models and 10 Amps for 230 V models should be located within 25 feet (115 V) or 10 feet (230 V) of the station. p.9 ↗ | Plan electrical service location accordingly; longer runs or shared circuits may cause voltage drop or nuisance trips. | All models |
| The station discharges HEPA-filtered exhaust air underneath the unit and must be placed at least 12 inches away from any vertical obstruction on all sides. p.9 ↗ | Maintain clearance to avoid restricting exhaust airflow and degrading containment. | All models |
| At least six inches (150 mm) clearance from any overhead obstructions is required when the station is in its final operating position. p.10 ↗ | Account for overhead clearance in site planning, especially when the lift system raises the station. | All models |
| The PuriCare models weigh approximately 450 lbs (205 kg). p.11 ↗ | Use mechanical lifting equipment (lift truck, floor jack, or flat low dolly under the shipping skid); do not tilt onto a hand truck. | All models |
| Average downflow velocity should be set at 55 ±5 FPM for models 3820000, -01, -10 & -11, and 50-55 FPM for models 3820020-38. p.30 ↗ | Certifier must adjust speed controls to achieve these setpoints; deviation affects containment. | Model-specific |
| Exhaust volume targets: 4 FT 100/115 V units 509-528 CFM (135-140 FPM); 4 FT 230 V units 434-452 CFM (115-120 FPM); 5 FT 100/115 V units 639-664 CFM (125-130 FPM); 5 FT 230 V units 537-562 CFM (105-110 FPM). p.30 ↗ | Use these values to verify exhaust airflow during certification; open filter area is 3.77 sq ft (4 FT) and 5.11 sq ft (5 FT). | Model-specific |
| The lift system is designed for intermittent operation only; continuous operation for more than one out of each 10-minute interval may activate the thermal override. p.21 ↗ | If thermal override activates, allow the lift system to cool for at least 10 minutes before adjusting height. | All models |
| The retractable power cord is 26 feet long and equipped with a safety brake. p.22 ↗ | To release the brake, pull the cord out until the ratcheting sound stops, then hold horizontally to guide it back. | All models |
| Both sashes must be completely closed for the PuriCare to contain properly. p.20 ↗ | Operating with sashes open compromises containment; ensure sashes are closed before starting work. | All models |
| Keep all materials at least 4 inches inside of the sash and perform contaminated operations as close to the center of the work area as possible. p.23 ↗ | Improper technique can disrupt airflow and cause contamination; follow work technique guidelines. | All models |
| The speed controls are rated for 10 Amps current each, far in excess of the motor's normal current draw. p.17 ↗ | Provides reliability margin; adjustment should only be performed by a qualified certifier. | All models |
| Turning the speed control adjustment screw all the way counterclockwise will shut the speed control circuit off and the motor(s) will stop. p.31 ↗ | If motor stops during adjustment, turn screw clockwise until motor starts running again. | All models |
| HEPA filter replacement requires loosening four locking bolts at least 1/2 inch and tightening four plenum lift bolts until approximately 1/4 inch clearance exists between the top of the HEPA filter and the bottom of the plenum. p.32 ↗ | Follow this clearance to avoid damaging the filter during removal/installation; recertify after replacement. | All models |
| The station's electrical outlets are protected by a Ground Fault Interrupter Circuit (GFIC); Labconco does not recommend installing a GFIC on the outlet that the PuriCare plugs into. p.10 ↗ | Avoid double GFIC protection which may cause nuisance tripping. | All models |
| Prefilters are disposable and cannot be cleaned; they may be contaminated with allergenic material. p.27 ↗ | Take appropriate precautions when removing and handling prefilters; replace as needed. | All models |
| Fluorescent lamps contain mercury and must be disposed of as hazardous waste according to local laws. p.29 ↗ | Do not place lamps in trash; consult www.lamprecycle.org for recycling information. | All models |
| The PuriCare Open Access Station is NOT a biosafety cabinet and must not be used for containment of biohazardous, toxic, flammable, or explosive materials. p.17 ↗ | Do not use for applications requiring biosafety containment; HEPA filters are only effective against particulate matter. | All models |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Station blower and lights won’t turn on p.41 ↗ | Unit not plugged into outlet | Plug the PuriCare into appropriate electrical service. | Chapter 8 |
| Station blower and lights won’t turn on p.41 ↗ | Circuit breakers tripped | Reset circuit breakers. | Chapter 8 |
| Blower(s) won’t turn on but lights work p.41 ↗ | Blower(s) wiring is disconnected | Inspect blower(s) wiring. | Chapter 8 |
| Blower(s) won’t turn on but lights work p.41 ↗ | Blower(s) motor is defective | Replace blower(s) motor. | Chapter 8 |
| Station blower turns on but lights don’t work p.41 ↗ | Lamp(s) not installed correctly | Inspect lamp installation. | Chapter 8 |
| Station blower turns on but lights don’t work p.41 ↗ | Lamp(s) is defective | Replace lamp(s). | Chapter 8 |
| Station blower turns on but lights don’t work p.41 ↗ | Lamp wiring is disconnected | Inspect lamp wiring. | Chapter 8 |
| Station blower turns on but lights don’t work p.41 ↗ | Defective lamp ballast | Replace lamp ballast. | Chapter 8 |
| Slight increase in pressure reading p.42 ↗ | HEPA filter loading | The pressure reading will steadily increase as the unit is used. | Chapter 8 |
| Slight increase in pressure reading p.42 ↗ | Prefilter loading | Check the prefilters for obstruction(s). Clean if necessary. | Chapter 8 |
| Slight increase in pressure reading p.42 ↗ | Blockage or restriction under the work surface | Ensure that the plenum beneath the work surface is unobstructed. | Chapter 8 |
| Slight increase in pressure reading p.42 ↗ | Blockage of the exhaust outlet | Ensure that the exhaust outlet is not blocked or restricted. | Chapter 8 |
| Contamination of work in the station p.42 ↗ | Improper technique or procedure for the Open Access Station | See “Use of the Open Access Station” section in the manual. | Chapter 8 |
| Contamination of work in the station p.42 ↗ | Restriction of the grille – blockage of the exhaust outlet | Ensure that all return air slots, grilles and the exhaust outlet are unobstructed. | Chapter 8 |
| Contamination of work in the station p.42 ↗ | External factors are disrupting the station’s airflow patterns or acting as a source of contamination | See “Installation” section of this manual. | Chapter 8 |
| Contamination of work in the station p.42 ↗ | Station is out of adjustment/HEPA filter(s) are defective | Have station recertified. | Chapter 8 |