Rev g xpert bulk powder enclosures and systems stations user manual
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco XPert Bulk Powder Enclosures (5', 6', 8' widths; 115V and 230V), XPert Bulk Powder Filtered Systems (5', 6'), and XPert Bulk Powder Filtered Stations (5', 6'). Ordering codes: Enclosures 3963500/3963520 (5' L), 3963501/3963521 (5' C), 3963502/3963522 (5' R), 3963600/3963620 (6' L), 3963601/3963621 (6' C), 3963602/3963622 (6' R), 3963800/3963820 (8' L), 3963801/3963821 (8' C), 3963802/3963822 (8' R); Filtered Systems 3961500/3961520 (5' L), 3961501/3961521 (5' C), 3961502/3961522 (5' R), 3961600/3961620 (6' L), 3961601/3961621 (6' C), 3961602/3961622 (6' R); Filtered Stations 3962500/3962520 (5' L), 3962501/3962521 (5' C), 3962502/3962522 (5' R), 3962600/3962620 (6' L), 3962601/3962621 (6' C), 3962602/3962622 (6' R).
Section-by-Section Technical Breakdown
Defines exhaust collar dimensions, transition adapters, and required exhaust volume, static pressure loss, FilterMate quantity, and noise level at face velocities of 60, 80, and 100 fpm for each enclosure width. Also covers filtered System/Station exhaust data and the requirement for an adjustable damper when connected to house exhaust.
- Exhaust connection accepts a 2" x 10" (5.1 cm x 25.4 cm) nominal collar.
- 5' and 6' enclosures require one exhaust connection to house exhaust; 8' requires two.
- 6" Transition Adapter 3912401 recommended for house exhaust on all 5', 6', and 8' enclosures.
- 5" Transition Adapter 3912400 required for FilterMate Portable Exhausters.
- 8' enclosure house exhaust may use 3981600 Exhaust Transition (6" x 6" x 8").
- Adjustable damper (or valve) must be installed when connected directly to house exhaust.
- Filtered Station requires 10" diameter ductwork for 5' and 6' models and a remote blower.
Specifies the standard duplex receptacle ratings, the four-plex electrical box under the work surface, and the maximum amperage for the remote blower circuit on the XPert Station. Also covers wiring requirements and the prohibition on inadequately rated detachable power cords.
- Standard duplex receptacles: 115V, 12A or 230V, 6A.
- Four-plex electrical box: four outlets rated 115V, 12A or 230V, 6A.
- XPert Station remote blower circuit: maximum 9A at 115V or 4.5A at 230V.
- All building wiring requires a licensed electrician and must conform to local codes.
- Do not use any detachable power cord not adequately rated for the unit.
Covers unpacking, moving, and installing the upper enclosure and work surface on the lift stand, including fastening hardware part numbers and quantities. Also covers sealing the upper enclosure to the work surface with silicone sealant.
- Upper enclosure with attached work surface weighs 100 to 550 lbs (45 to 248 kg); lift stands weigh 380 to 500 lbs (171 to 225 kg).
- Do not lift by the front or side air foils.
- Fastening hardware: 1880740 Hex Bolt ¼-20 x 2.5 (qty 8); 1880708 Hex Head Screw ¼-20 x 0.5 (qty 3).
- Seal upper enclosure to work surface with a clear bead of silicone sealant.
Specifies the face velocity range for weighing operations and describes factory and independent validation testing. Requires face velocity profiles and smoke testing per the organization's quality system.
- Face velocity should be adjusted from 60 to 100 fpm for all weighing operations.
- Factory tested per ASHRAE 110-1995; as-manufactured rating less than 0.05 ppm at 4 L/min.
- Independent validation performed using sodium naproxen powder.
- Face velocity profiles and smoke testing should be performed frequently.
Describes the containment-enhancing design features including barrier containment air foils, rear perforated baffle, upper dilution air supply, transfer cart, automated lift, retaining rings, safety sleeve, retaining bands, stainless steel work surface, HEPA filtration, waste chute, and four-plex electrical box.
- Automated lift: 12" travel and 350 pounds capacity; photoelectric eye stops container below work surface.
- Retaining rings available in various diameters up to 19.6" dia.
- Waste chute requires hazardous waste bags minimum 8-1/2" wide x 11" long.
- Interior dimensions: 30" depth and 32" height.
- HEPA filters include true Bag-In/Bag-Out filter change system.
Provides planning, start-up, loading, work techniques, work procedure, unloading, clean-up, and shutdown steps. Includes timing requirements for purging airborne contaminants and handling of source containers with safety sleeves and retaining rings.
- Allow enclosure to operate for 1 minute prior to starting procedures.
- After loading and attaching bags, wait 15 seconds to purge airborne contaminants.
- Upon completion of work, allow enclosure to operate for two minutes undisturbed.
- Elevate large objects at least ¼" above the work surface to permit airflow underneath.
- Keep all materials behind the lower air foil and perform contaminated operations as far to the rear as possible.
Covers the routine maintenance schedule and programming of the Guardian 1000 Digital Alarm.
- Routine maintenance schedule is provided.
- Guardian 1000 Digital Alarm programming instructions are included.
Lists available accessories including safety sleeves, retaining bands, retaining rings, fluorescent light kits, exhaust transition adapters, remote blowers, exhausters dampers, FilterMate Portable Exhausters and filters, HEPA and trace odor control filters, canopy connection, transport cart, shelf kits, double HEPA filtration kit, hose kits, caster kit, and lift platform extension.
- Remote blowers and FilterMate Portable Exhausters are available for exhaust.
- HEPA and trace odor control filters are available.
- Double HEPA Filtration Kit is available.
- Canopy Connection aids in connecting to remote blower or house exhaust.
Provides a table of problems, causes, and corrective actions for contamination, blower operation, face velocity, lighting, airflow monitor, and lift malfunctions.
- Covers contamination outside enclosure, blower won't operate, low face velocity, blower and lights won't operate, lights do not work, airflow monitor malfunction, and lift won't operate or malfunctions.
- Includes service operating record log.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Face velocity should be adjusted from 60 to 100 fpm for all weighing operations. p.20 ↗ | Set exhaust blower or impeller speed to achieve average face velocity within this range for proper containment. | All XPert Bulk Powder Enclosures, Systems, and Stations |
| 5' enclosure at 60/80/100 fpm requires exhaust volumes of 290/385/480 CFM with static pressure losses of .20"/.35"/.54" H2O and 2/2/3 FilterMates. p.10 ↗ | Select blower or number of FilterMate Portable Exhausters to match required exhaust volume and static pressure at the desired face velocity. | 5' XPert Bulk Powder Enclosure |
| 6' enclosure at 60/80/100 fpm requires exhaust volumes of 350/465/580 CFM with static pressure losses of .27"/.48"/.75" H2O and 2/3/3 FilterMates. p.10 ↗ | Select blower or number of FilterMate Portable Exhausters to match required exhaust volume and static pressure at the desired face velocity. | 6' XPert Bulk Powder Enclosure |
| 8' enclosure at 60/80/100 fpm requires exhaust volumes of 465/620/775 CFM with static pressure losses of .15"/.27"/.42" H2O and 3/3/3 FilterMates. p.10 ↗ | Select blower or number of FilterMate Portable Exhausters to match required exhaust volume and static pressure at the desired face velocity. | 8' XPert Bulk Powder Enclosure |
| 5' Filtered System/Station at 60/80/100 fpm requires exhaust volumes of 290/385/480 CFM with initial static pressure loss with HEPA filter of .38"-.46"/.55"-.67"/.72"-.85" H2O. p.11 ↗ | Size remote blower for Station or verify System impeller performance against these pressure losses. | 5' XPert Bulk Powder Filtered Balance System or Station |
| 6' Filtered System/Station at 60/80/100 fpm requires exhaust volumes of 350/465/580 CFM with initial static pressure loss with HEPA filter of .46"-.56"/.67"-.81"/.87"-1.03" H2O. p.11 ↗ | Size remote blower for Station or verify System impeller performance against these pressure losses. | 6' XPert Bulk Powder Filtered Balance System or Station |
| XPert Station remote blower circuit maximum amperage: 9 amps at 115V or 4.5 amps at 230V. p.12 ↗ | Ensure remote blower electrical draw does not exceed these limits when connected to the enclosure blower switch. | XPert Filtered Balance Station |
| Four-plex electrical box outlets rated at 115V, 12A or 230V, 6A. p.12 ↗ | Total connected load to the convenience outlets must not exceed these ratings. | All models |
| Automated lift has 12" travel and 350 pounds capacity. p.22 ↗ | Do not exceed 350 lb source container weight; lift height adjustment limited to 12 inches. | All models with automated lift |
| Retaining rings available in various diameters up to 19.6" dia. p.22 ↗ | Select retaining ring slightly larger than source container diameter; maximum container diameter approximately 19.6 inches. | All models |
| Waste chute requires hazardous waste bags minimum 8-1/2" wide x 11" long. p.23 ↗ | Procure bags meeting minimum dimensions for proper waste chute use. | All models |
| Interior dimensions: 30" depth and 32" height. p.22 ↗ | Verify equipment fits within these interior clearances. | All XPert Bulk Powder Enclosures |
| Maintain humidity level between 45 and 60% to minimize electrostatic issues. p.9 ↗ | HVAC or humidification may be required to sustain this range for accurate weighing and reduced powder attraction. | All models |
| Leave 6-12" around the enclosure for proper airflow. p.12 ↗ | Maintain clearance from walls and other equipment during installation. | All models |
| Factory as-manufactured containment rating less than 0.05 ppm at 4 L/min per ASHRAE 110-1995. p.20 ↗ | Baseline containment performance; consult Labconco for individual ratings. | All XPert Bulk Powder Enclosures |
| Allow enclosure to operate for 1 minute prior to starting procedures; wait 15 seconds after loading; operate for two minutes undisturbed upon completion. p.25 ↗ | Follow these purge times to ensure airborne contaminants are cleared before and after work. | All models |
| Elevate large objects at least ¼" above the work surface to permit airflow underneath. p.27 ↗ | Use stands or supports to allow air to sweep under equipment and into rear baffle slots. | All models |
| Upper enclosure with attached work surface weighs 100 to 550 lbs (45 to 248 kg); lift stands weigh 380 to 500 lbs (171 to 225 kg). p.13 ↗ | Use hydraulic lift, electric lift, or forklift for moving; do not lift by front or side air foils. | All models |
| Fastening hardware for work surface to lift stand: 1880740 Hex Bolt ¼-20 x 2.5 (qty 8); 1880708 Hex Head Screw ¼-20 x 0.5 (qty 3). p.15 ↗ | Use specified bolts and screws for secure assembly; long bolts pass through 2" sq. tubing. | All models |
| 6" Transition Adapter 3912401 recommended for house exhaust on all 5', 6', and 8' enclosures; 5" Transition Adapter 3912400 required for FilterMate Portable Exhausters. p.17 ↗ | Order correct adapter for exhaust configuration. | All XPert Bulk Powder Enclosures |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Contamination outside of enclosure p.39 ↗ | Improper user techniques for the enclosure. | See “Using your Enclosure” Chapter 5 and “Safety Precautions” Chapter 4 sections in the manual. | Chapter 8 |
| Contamination outside of enclosure p.39 ↗ | Restriction of the baffle air slots or blockage of the exhaust outlet. | Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. | Chapter 8 |
| Contamination outside of enclosure p.39 ↗ | External factors are disrupting the enclosure airflow patterns or acting as a source of contamination. | See “Location Requirements” Chapter 2, “Certifying the Enclosure” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. | Chapter 8 |
| Contamination outside of enclosure p.39 ↗ | Enclosure has improper face velocity. | Have enclosure certified and check exhaust system. Check HEPA filters for loading. Adjust enclosure speed control. Enclosure should have an average face velocity of 60-100 fpm for weighing procedures. | Chapter 8 |
| Blower won’t operate p.39 ↗ | Unit not plugged into outlet. | Plug the enclosure into appropriate electrical service. | Chapter 8 |
| Blower won’t operate p.39 ↗ | Circuit breaker(s) or Ground Fault Interrupter. | Reset circuit breaker. | Chapter 8 |
| Blower won’t operate p.39 ↗ | Blower wiring is disconnected. | Inspect blower wiring. | Chapter 8 |
| Blower won’t operate p.39 ↗ | Blower switch is defective. | Replace switch. | Chapter 8 |
| Blower won’t operate p.39 ↗ | Motorized impeller or blower is defective. | Replace motorized impeller or blower. | Chapter 8 |
| Low face velocity or poor containment of contaminants p.39 ↗ | Enclosure sash not closed. | Close sash to the lowest position. | Chapter 8 |
| Low face velocity or poor containment of contaminants p.39 ↗ | HEPA filter clogged. | Replace HEPA filter or increase speed. | Chapter 8 |
| Low face velocity or poor containment of contaminants p.39 ↗ | Carbon filter loaded with chemicals. | Replace carbon filter. | Chapter 8 |
| Blower and lights won’t operate p.39 ↗ | Unit not plugged into outlet. | Plug enclosure into appropriate electrical service. | Chapter 8 |
| Blower and lights won’t operate p.39 ↗ | Circuit breaker(s) tripped. | Reset or replace circuit breaker. | Chapter 8 |
| Lights do not work p.40 ↗ | Lamp not installed properly. | Inspect lamp installation. | Chapter 8 |
| Lights do not work p.40 ↗ | Lamp wiring disconnected. | Inspect lamp wiring. | Chapter 8 |
| Lights do not work p.40 ↗ | Defective lamp. | Replace lamp. | Chapter 8 |
| Lights do not work p.40 ↗ | Light switch is defective. | Replace light switch. | Chapter 8 |
| Lights do not work p.40 ↗ | Defective electronic ballast. | Replace ballast. | Chapter 8 |
| Airflow monitor malfunction: No power. No lights. No display. p.40 ↗ | Power supply is not plugged into proper voltage; plug in power supply. Verify that airflow monitor interface cables are connected. Check fuses on FilterMate or accessory light. | Plug in power supply; verify interface cables; check fuses. | Chapter 8 |
| Airflow monitor malfunction: No audible alarm. p.40 ↗ | Alarm has been temporarily silenced using “test/reset” or “enter” buttons. | Press “test/reset” or “enter” to re-enable audible alarm. | Chapter 8 |
| Airflow monitor malfunction: Wrong alarm set point. p.40 ↗ | Airflow monitor was not properly adjusted. | Repeat calibration steps outlined in the User’s Manual. | Chapter 8 |
| Airflow monitor malfunction: Constant audible alarm. p.40 ↗ | Check airflow and calibration of airflow monitor. | Check airflow and calibration of airflow monitor. | Chapter 8 |
| Airflow monitor malfunction: Continuous alarm. p.40 ↗ | Check the face velocity of the enclosure as the airflow of the system may have changed. If face velocity is correct, calibrate the airflow monitor as outlined in the User’s Manual. | Check face velocity; calibrate airflow monitor. | Chapter 8 |
| Airflow monitor malfunction: Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. p.40 ↗ | Air way to airflow monitor sensor is blocked by insects, dust or debris. | Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. | Chapter 8 |
| Airflow monitor malfunction: Audible disable will not stay ON. p.40 ↗ | An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. | Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower. | Chapter 8 |
| Lift won’t operate or malfunctions: No power. p.40 ↗ | Lift motor is not connected to a proper outlet. | Connect the plug into the four-plex electrical box located under work surface. | Chapter 8 |
| Lift won’t operate or malfunctions: Intermittent operation. p.40 ↗ | Thermal override activated. | Allow lift system’s thermal override to cool for 10 minutes then adjust height. | Chapter 8 |
| Lift won’t operate or malfunctions: Photoelectric eye malfunctions. p.40 ↗ | Photoelectric eye malfunction. | Either adjust photoelectric eye sensitivity or replace photoelectrical eye if it malfunctions. See replacement parts. | Chapter 8 |
| Lift won’t operate or malfunctions: Hydraulic lines pinched. p.40 ↗ | Hydraulic lines pinched. | Replace entire lift system. See replacement parts. | Chapter 8 |