Technical Reference Summary
Covered Models & Document Scope
Covers the Labconco XPert® Nano Enclosure series in 2', 3', 4', 5' and 6' widths. 2'/3'/4' models use a 9.4" sash opening and 6" nominal exhaust duct; 5'/6' models use a 12" sash opening and 10" nominal exhaust duct. Options include factory-installed ionizer, Guardian Airflow Monitor or Guardian 1000 Digital Airflow Monitor, Canopy Exhaust Connection, remote blowers, exhaust dampers, accessory filters and utility shelves. Electrical variants: 115 V with power cord and 230 V shipped without plug.
Section-by-Section Technical Breakdown
Defines prerequisites for the installation site: level, solid construction, dedicated electrical power nearby, and a bench or stand that minimizes vibration. Temperature drift of 1–2 °C is generally tolerable per balance manufacturers. Humidity should be maintained between 45 and 60% to control static electricity, and the enclosure avoids plastics to reduce insulative surfaces.
- Avoid tubular stands or mobile benches that can move when touched.
- A bench rigidly mounted to floor or wall (not both) may be appropriate.
- Building corners typically have less vibration than the center.
- Do not place vibration-producing equipment such as shakers or pumps on the bench.
- Marble slabs with dampening pads inside the enclosure are an effective low-cost vibration control.
- Calibrate and verify the balance at least annually; best results when calibrated inside the enclosure at the desired face velocity.
- Never install balances near radiators or hot plates; avoid direct sunlight on the bench.
- Maintain humidity between 45 and 60%; a factory-installed ionizer is available to neutralize surface charges.
The enclosure must be placed to avoid cross drafts and turbulence that disrupt containment and balance readings. High-traffic areas, HVAC diffusers, fans, radiators, and openable doorways or windows should be avoided. The main disconnect device must remain accessible.
- Avoid high-traffic areas where walking causes air disturbance.
- Avoid overhead or wall HVAC diffusers, fans, radiators or other air-current-producing lab equipment.
- Avoid locations next to doorways or windows that may be opened.
- Do not position the unit so that the main disconnect device is difficult to operate.
The integral motorized impeller draws room air through the ULPA filter and exhausts filtered air from the top. Exhaust can be recirculated into the lab or ducted outside via the optional Canopy Exhaust Connection and remote blower. Tables list exhaust volume, static pressure loss, noise and maximum external static pressure for each width at 60, 80 and 100 fpm.
- 2' model at 60/80/100 fpm: 85/115/145 CFM; initial static pressure loss 0.35–0.41 / 0.48–0.57 / 0.62–0.72 in w.g.; noise 48–53 / 50–56 / 53–58 dB(A); max external static pressure 0.20 / 0.18 / 0.13 in w.g.
- 3' model at 60/80/100 fpm: 130/175/220 CFM; static pressure loss 0.37–0.43 / 0.53–0.62 / 0.68–0.79 in w.g.; noise 48–53 / 53–57 / 58–61 dB(A); max external static pressure 0.15 / 0.11 / 0.08 in w.g.
- 4' model at 60/80/100 fpm: 175/235/295 CFM; static pressure loss 0.40–0.46 / 0.57–0.66 / 0.76–0.87 in w.g.; noise 49–54 / 58–61 / 62–66 dB(A); max external static pressure 0.20 / 0.18 / 0.13 in w.g.
- 5' model at 60/80/100 fpm: 290/385/480 CFM; static pressure loss 0.38–0.46 / 0.55–0.67 / 0.72–0.85 in w.g.; noise 52–56 / 58–63 / 63–67 dB(A); max external static pressure 0.22 / 0.20 / 0.15 in w.g.
- 6' model at 60/80/100 fpm: 350/465/580 CFM; static pressure loss 0.46–0.56 / 0.67–0.81 / 0.87–1.03 in w.g.; noise 55–57 / 61–63 / 67–69 dB(A); max external static pressure 0.20 / 0.18 / 0.13 in w.g.
- Weighing and handling nano materials is best done at lower flow rates of 60–70 fpm (0.31–0.36 m/s).
- If connected directly to a house exhaust system, an adjustable damper or valve must be installed to control airflow, especially when one house system serves multiple filtered enclosures.
Standard duplex receptacles should be nearby. The 115 V model uses a supplied power cord connected to the IEC inlet on the rear; the 230 V model ships without a plug and requires an appropriate plug per local codes. All models draw less than 5 A. At 50 Hz, maximum blower airflow is reduced by 17%.
- 115 V: connect supplied power cord to IEC electrical supply plug on back of enclosure.
- 230 V: shipped without plug; install appropriate plug per local electrical specifications and codes.
- All models use less than 5 amps.
- At 50 Hz operation, blower maximum airflow is reduced by 17%.
- All wiring should be performed by a licensed electrician and conform to local codes.
- Do not use any detachable power cord that is not adequately rated for the unit.
Enclosures weigh 173 to 443 lbs (78.47 to 200.94 kg) and must be moved with a mechanical lift truck, floor jack or low dolly under the shipping skid; do not tilt onto a hand truck or lift by the front air foil. The base stand top surface adjusts from 27.5 to 34.50 inches (698 to 876 mm) in 1-inch (25 mm) increments. The integral stainless steel work surface mounts to the stand with four 1/4"-20 x 2.5" bolts.
- Enclosure weight range: 173 to 443 lbs (78.47 to 200.94 kg).
- Do not move the enclosure by tilting it onto a hand truck.
- Do not lift by the front air foil.
- Base stand height adjustable 27.5 to 34.50 inches (698 to 876 mm) in 1-inch (25 mm) increments.
- Leg extensions attach with 5/16"-18 x 2-1/4 inch (57 mm) bolts, flatwasher, lockwasher and acorn nut; select the same height hole for each leg.
- Work surface mounts with four 1/4"-20 x 2.5 inch (63 mm) hex head bolts, nuts and washers; front edge flush with stand front edge.
- Use a carpenter level to adjust leveling feet in both planes.
The 2', 3' and 4' models accept 6" nominal duct (6.06" ID) and the 5' and 6' models accept 10" nominal duct (10.06" ID). On 2'/3'/4' sizes the exhaust connection ships uninstalled on top and is fastened with provided screws. Ductwork weight must be supported independently of the enclosure superstructure.
- 2', 3', 4' models: 6" nominal exhaust duct, 6.06" ID.
- 5', 6' models: 10" nominal exhaust duct, 10.06" ID.
- Exhaust connection on 2'/3'/4' ships uninstalled on top; fasten with screws provided.
- The weight of the exhaust ductwork system must be supported independently of the enclosure superstructure or damage may occur.
- Exhaust system should be installed by a qualified HVAC contractor.
- Selected exhaust duct material should match the enclosure, procedures and chemical applications.
The speed control behind the front panel adjusts face velocity from 60 to 100 fpm. Average face velocity is measured with an anemometer in two rows across the sash opening: readings 6" from the ends and every 12", first row 3" down from the upper sash foil and second row 3" up from the work surface. Labconco airflow testing per ASHRAE 110-1995 achieved an as-manufactured rating of less than 0.05 ppm at 4 lpm.
- Face velocity adjustable from 60 to 100 fpm via speed control behind front panel.
- Containment is maximized within this range; lowest permissible velocity gives least turbulent, quietest operation.
- Average face velocity readings: two rows, 6" from ends, evenly spaced every 12"; first row 3" down from upper sash foil, second row 3" up from work surface.
- As-manufactured ASHRAE 110-1995 rating: less than 0.05 ppm at 4 lpm.
- Face velocity profiles and smoke testing should be performed frequently per the organization's quality system.
Describes the aerodynamic design elements: Clean-Sweep air foil, upper sash foil, upper dilution air supply, zoned rear perforated baffle, side-entry air foils, safety glass sash, utility ports, negative-pressure plenum, vibration-isolated motorized impeller, speed control, filter system pressure gauge, ULPA filter access port, control panel, ULPA filter, optional exhaust collar, canopy exhaust connection, ionizer and utility shelves.
- Upper dilution air supply introduces 5 to 7 percent of required air volume through the upper dilution air supply.
- Zoned rear perforated baffle has three zones with increasingly more open area at the bottom to form laminar airflow.
- Safety glass sash has a wiping seal and spring-loaded latch; 5' and 6' units use a gas spring assist instead of the latch.
- ULPA filter rated to remove 99.999% of all particles 0.12 micron in size.
- Plenum and ULPA filter are jacketed by negative pressure; leaks are recaptured and refiltered.
- True bag-in/bag-out ULPA filter disposal system protects worker from hazardous powders and particulates.
- Optional ionizer cannot be installed after manufacture.
- Filter system pressure gauge indicates total system pressure across filter and pressurized plenum to help predict filter loading.
Covers planning, start-up, loading, work techniques, final purging, unloading and shutdown. Allow the enclosure to operate unobstructed for 1 minute at start-up and after loading; after work, allow 2 to 3 minutes undisturbed purging before shutting down the blower.
- Turn on exhaust system and light; only raise sash for loading and cleaning.
- Check baffle air slots for obstructions.
- Allow enclosure to operate unobstructed for 1 minute before work.
- After loading, wait one minute to purge airborne contaminants.
- Upon completion, allow enclosure to operate for two to three minutes undisturbed before shutting down blower.
- Surface decontaminate objects in contact with contaminated material before removal; cover open trays, weigh vessels or containers before removal.
The ULPA filter is suitable for weighing and handling powders, nonvolatile particulates, hazardous powders and biohazardous materials. It provides personnel and environmental protection but not product protection from environmental contamination. Gases or vapors require ventilation to the outside. The filter surface is fragile and must not be touched or punctured.
- Suitable for weighing and handling of powders and release of nonvolatile particulates.
- Provides personal and environmental protection from particulate matter, hazardous powders and biohazardous materials.
- Not for operations requiring sample protection from environmental contamination.
- ULPA filters are only effective for entrapment of particulate matter; gases or vapors require outside ventilation.
- If a ULPA filter is suspected damaged, do not use the enclosure; contact a certification agency or Labconco.
- For powders and particulates follow the ULPA Filter Bag-In/Bag-Out procedure in Chapter 6; for biohazardous materials follow the Decontamination Procedure.
The factory-installed ionizer controls static charge on surfaces within the enclosure. Its discharge is located at the front face of the work surface, safe from contaminants and cleaning. The emitters have a finite life and may require replacement.
- Ionizer floods the work area with positive and negative ions to neutralize static charges on non-conductive items.
- Discharge located at the front face of the work surface.
- Ionizer cannot be installed after manufacture.
- Emitters have a finite life and may require replacement.
Provides a schedule for routine maintenance including decontamination, determination of when to replace ULPA filters, bag-in/bag-out filter installation, filter leak testing, setting inflow face velocity, and calibrating and operating airflow monitors.
- Routine maintenance schedule covers decontamination, filter replacement determination, bag-in/bag-out filter installation, filter leak test, face velocity setting and airflow monitor calibration.
- Initial certification and re-certification are required.
- Fluorescent and LED light replacement, motorized impeller replacement, speed control replacement, ionizer maintenance and trace odor carbon filter installation are covered.
Describes determination of when to replace ULPA filters, installation of a new ULPA filter using the bag-in/bag-out procedure, and filter leak testing. The filter clamping bolt evenly seals the filter to the frame. Exhaust ULPA filters are disposed as hazardous waste using the disposal bag.
- Replace ULPA filter when face velocity falls below acceptable limits established by Safety Officer.
- Use true bag-in/bag-out procedure for harmful powders and particulates.
- Filter clamping bolt evenly seals filter to enclosure frame.
- Exhaust ULPA filters are disposed as hazardous waste; user is responsible for recording material and disposing properly.
- Filter leak test is part of maintenance.
Covers the Guardian Digital 1000 Airflow Monitor and Guardian Airflow Monitor, including programming the Guardian 1000 Digital Alarm. The Guardian 1000 displays face velocity, provides RS232 output, night setback mode and auxiliary relay ports. Calibration steps are outlined in Chapter 6 or the supplied Digital 1000 Monitor Manual.
- Guardian 1000 Digital Airflow Monitor displays face velocity value, provides RS232 output, night setback mode and several auxiliary relay ports.
- Guardian Airflow Monitor continuously monitors airflow with audio/visual alarm for low airflow.
- Programming the Guardian 1000 Digital Alarm is covered.
- Calibration steps are in Chapter 6 or the supplied Digital 1000 Monitor Manual.
Covers fluorescent and LED light replacement, motorized impeller replacement, speed control replacement, maintenance on the optional ionizer, and trace odor carbon filter installation.
- Fluorescent light replacement and LED light replacement procedures.
- Motorized impeller replacement procedure.
- Speed control replacement procedure.
- Maintenance on the optional ionizer.
- Trace odor carbon filter installation.
Lists accessory options: Guardian Airflow Monitors, Canopy Exhaust Connection, remote blowers, exhaust dampers and accessory filters. The Canopy Exhaust Connection includes an epoxy-coated steel exhaust transition adapter for 6" nominal duct on 2'/3'/4' sizes and 10" nominal duct on 5'/6' sizes.
- Guardian Airflow Monitors available as accessories.
- Canopy Exhaust Connection includes epoxy-coated steel exhaust transition adapter.
- 2', 3', 4' sizes use 6" nominal duct; 5', 6' sizes use 10" nominal duct.
- Remote blowers and exhaust dampers available.
- Accessory filters available.
Appendix A lists replacement parts, Appendix B provides dimensions for all models, Appendix C provides electrical and environmental specifications, and Appendix D is a quick chart.
- Appendix A: Replacement Parts.
- Appendix B: Dimensions for different models.
- Appendix C: Specifications for electrical and environmental conditions.
- Appendix D: Quick Chart.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| ULPA filter rated at least 99.999% efficient for 0.12-micron particles. p.5 ↗ | Suitable for containment of hazardous powders and biohazardous particulates; not for gases or vapors. | whole series |
| Face velocity adjustable from 60 to 100 fpm via speed control behind front panel. p.18 ↗ | Set within this range for containment; lowest permissible velocity gives least turbulent, quietest operation. | whole series |
| For weighing procedures, face velocity should be 60–100 fpm; for biohazardous procedures, 75–105 fpm. p.54 ↗ | Select face velocity according to application; verify with anemometer. | whole series |
| Weighing and handling nano materials is best done at lower flow rates of 60–70 fpm (0.31–0.36 m/s). p.10 ↗ | Use lower face velocity for nanoparticle work to optimize containment. | whole series |
| 2', 3', 4' models accept 6" nominal exhaust duct (6.06" ID); 5', 6' models accept 10" nominal duct (10.06" ID). p.15 ↗ | Ductwork and canopy exhaust connection must match model width. | 2'/3'/4' vs 5'/6' models |
| 2' model at 60 fpm: 85 CFM, initial static pressure loss 0.35–0.41 in w.g., noise 48–53 dB(A), max external static pressure 0.20 in w.g. p.10 ↗ | Use for blower sizing and exhaust system design. | 2' model |
| 6' model at 100 fpm: 580 CFM, initial static pressure loss 0.87–1.03 in w.g., noise 67–69 dB(A), max external static pressure 0.13 in w.g. p.10 ↗ | Highest airflow and pressure loss of the series; blower must be sized accordingly. | 6' model |
| All models use less than 5 amps. p.18 ↗ | Standard duplex receptacle circuit is adequate; verify local codes. | whole series |
| At 50 Hz operation, blower maximum airflow is reduced by 17%. p.18 ↗ | Account for reduced airflow when sizing blower and setting face velocity in 50 Hz regions. | whole series |
| Enclosure weight range: 173 to 443 lbs (78.47 to 200.94 kg). p.12 ↗ | Use mechanical lift truck, floor jack or low dolly under shipping skid; do not tilt onto hand truck or lift by front air foil. | whole series |
| Base stand top surface adjustable from 27.5 to 34.50 inches (698 to 876 mm) in 1-inch (25 mm) increments. p.14 ↗ | Select optimum working height; use same height hole for each leg. | models with base stand |
| Work surface mounts with four 1/4"-20 x 2.5 inch (63 mm) hex head bolts, nuts and washers; leg extensions use 5/16"-18 x 2-1/4 inch (57 mm) bolts. p.14 ↗ | Use specified hardware for assembly; tighten after all four bolts started. | models with base stand |
| Upper dilution air supply introduces 5 to 7 percent of required air volume. p.23 ↗ | Part of total airflow; account for when measuring face velocity. | whole series |
| Average face velocity measurement: two rows, readings 6" from ends and every 12", first row 3" down from upper sash foil, second row 3" up from work surface. p.18 ↗ | Standardized method for certifying face velocity. | whole series |
| As-manufactured ASHRAE 110-1995 rating: less than 0.05 ppm at 4 lpm. p.18 ↗ | Baseline containment performance; field testing available through Labconco certified contractor. | whole series |
| Maintain humidity between 45 and 60% to control static electricity. p.8 ↗ | Static charges can cause weighing errors and particulate attraction; ionizer available as factory option. | whole series |
| Temperature drift of 1–2 °C is generally tolerable per balance manufacturers. p.8 ↗ | Avoid heating sources and direct sunlight; validate through Operational Qualification protocol. | whole series |
| Leave blower on for at least one minute after work; allow 2 to 3 minutes undisturbed purging before shutdown. p.26 ↗ | Ensures airborne contaminants are purged before blower is turned off. | whole series |
| Enclosure should be recertified whenever serviced or at least annually thereafter. p.25 ↗ | Maintain certification schedule for continued safe operation. | whole series |
| If connected directly to a house exhaust system, an adjustable damper or valve must be installed to control airflow. p.10 ↗ | Required especially when one house system serves multiple filtered enclosures. | models with house exhaust connection |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Contaminants outside of enclosure. p.54 ↗ | Improper user techniques for the enclosure. | See "Certifying the Enclosure" Chapter 3 and "Safety Precautions" Chapter 4 sections in the manual. (Ref. Appendix D). | Chapter 8 |
| Contaminants outside of enclosure. p.54 ↗ | Restriction of the baffle air slots or blockage of the exhaust outlet. | Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. | Chapter 8 |
| Contaminants outside of enclosure. p.54 ↗ | External factors are disrupting the enclosure airflow patterns or acting as a source of contamination. | See "Location Requirements" Chapter 2, "Certifying the Enclosure" Chapter 3, and "Safety Precautions" Chapter 4 sections of this manual. (Ref. Appendix D). | Chapter 8 |
| Contaminants outside of enclosure. p.54 ↗ | Enclosure has improper face velocity. | Have enclosure certified and check exhaust system. Check filters for loading. Adjust enclosure speed control. Enclosure should have an average face velocity of 60-100 fpm for weighing procedures and 75-105 fpm for biohazardous procedures. | Chapter 8 |
| Blower won't operate. p.54 ↗ | Unit not plugged into outlet. | Plug the enclosure into appropriate electrical service. | Chapter 8 |
| Blower won't operate. p.54 ↗ | Circuit breaker(s) or Ground Fault Interrupter. | Reset circuit breaker. | Chapter 8 |
| Blower won't operate. p.54 ↗ | Blower wiring is disconnected. | Inspect blower wiring. | Chapter 8 |
| Blower won't operate. p.54 ↗ | Blower switch is defective. | Replace switch. | Chapter 8 |
| Blower won't operate. p.54 ↗ | Motorized impeller or blower is defective. | Replace motorized impeller or blower. See Chapter 6. | Chapter 8 |
| Low face velocity or poor containment of contaminants. p.54 ↗ | Enclosure sash not closed. | Close sash to the lowest position. | Chapter 8 |
| Low face velocity or poor containment of contaminants. p.54 ↗ | ULPA filter clogged. | Replace ULPA filter or increase speed. | Chapter 8 |
| Low face velocity or poor containment of contaminants. p.54 ↗ | Carbon filter loaded with chemicals. | Replace carbon filter. | Chapter 8 |
| Blower and lights won't operate. p.54 ↗ | Unit not plugged into outlet. | Plug enclosure into appropriate electrical service. | Chapter 8 |
| Blower and lights won't operate. p.54 ↗ | Circuit breaker(s) tripped. | Reset or replace circuit breaker. | Chapter 8 |
| Lights do not work. p.55 ↗ | Lamp not installed properly. | Inspect lamp installation. | Chapter 8 |
| Lights do not work. p.55 ↗ | Lamp wiring disconnected. | Inspect lamp wiring. | Chapter 8 |
| Lights do not work. p.55 ↗ | Defective lamp. | Replace lamp. | Chapter 8 |
| Lights do not work. p.55 ↗ | Light switch is defective. | Replace light switch. | Chapter 8 |
| Lights do not work. p.55 ↗ | Defective electronic ballast. | Replace ballast. | Chapter 8 |
| Airflow monitor malfunction. No power. No lights. No display. p.55 ↗ | Power cable to circuit board is disconnected and needs to be connected. Sensor cable needs to be connected. Power supply is not plugged into proper voltage; plug in power supply. Verify that all airflow monitor interface cables are connected. Check fuses on enclosure. | Connect power cable to circuit board; connect sensor cable; plug power supply into proper voltage; verify all airflow monitor interface cables are connected; check fuses on enclosure. | Chapter 8 |
| No audible alarm. p.55 ↗ | Alarm has been temporarily silenced using "SILENCE ALARM" or "enter" buttons. | Check alarm status; refer to Chapter 6 or Digital 1000 Monitor Manual. | Chapter 8 |
| Wrong alarm set point. p.55 ↗ | Airflow monitor was not properly adjusted. | Repeat calibration steps outlined in this manual in Chapter 6 or in the supplied Digital 1000 Monitor Manual. | Chapter 8 |
| Constant audible alarm. p.55 ↗ | Airflow monitor calibration or airflow issue. | Check airflow and calibration of airflow monitor. See Chapter 6. | Chapter 8 |
| Continuous alarm. p.55 ↗ | Face velocity of the enclosure may have changed; airflow of the system may have changed; ULPA filter may have become loaded. | Check the face velocity of the enclosure. If incorrect, adjust the speed control to increase face velocity. If face velocity is correct, calibrate the airflow monitor outlined in this manual in Chapter 6. | Chapter 8 |
| Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. p.55 ↗ | Airway to airflow monitor sensor is blocked by insects, dust or debris. | Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. | Chapter 8 |
| Audible disable will not stay operational. p.55 ↗ | An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. | Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower. | Chapter 8 |