Rev v xpert rxpert filtered balance systems stations
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco XPert® Filtered Balance System, XPert® Filtered Balance Station, and RXPert™ Filtered Balance System in 2', 3', 4', 5' and 6' widths, with 9.4" or 12" sash openings, 115V or 230V electrical models, single or double HEPA filtration, and optional accessory Odor Control carbon filters (organic vapor, formaldehyde, ammonia/amines, acid-sulfur) and Canopy Exhaust Connections. XPert Station model series 3940200, 3940300, 3940400, 3946500 and 3946600 are identified.
Section-by-Section Technical Breakdown
Defines airflow, exhaust volume, static pressure loss, noise and maximum external static pressure for each enclosure width and sash opening at 60, 80 and 100 fpm. XPert Systems use an integral motorized impeller and may recirculate or be exhausted outside via a Canopy Exhaust Connection and remote blower; XPert Stations have no integral impeller and must be exhausted by a remote blower.
- 2' 9.4" sash: 60/80/100 fpm correspond to 85/115/145 CFM, static loss .35-.41/.48-.57/.62-.72 in w.g., noise 48-53/50-56/53-58 dBA.
- 3' 9.4" sash: 60/80/100 fpm correspond to 130/175/220 CFM, static loss .37-.43/.53-.62/.68-.79 in w.g., noise 48-53/53-57/58-61 dBA.
- 4' 9.4" sash: 60/80/100 fpm correspond to 175/235/295 CFM, static loss .40-.46/.57-.66/.76-.87 in w.g., noise 49-54/58-61/62-66 dBA.
- 4' 12" sash (tall/deep): 60/80/100 fpm correspond to 230/305/380 CFM, static loss .56-.65/.77-.87/.92-1.04 in w.g., noise 58-61/62-66/67-70 dBA.
- 5' 12" sash (10" duct): 60/80/100 fpm correspond to 290/385/480 CFM, static loss .38-.46/.55-.67/.72-.85 in w.g., noise 52-56/58-63/63-67 dBA.
- 6' 12" sash (10" duct): 60/80/100 fpm correspond to 350/465/580 CFM, static loss .46-.56/.67-.81/.87-1.03 in w.g., noise 55-57/61-63/67-69 dBA.
- Balances typically work best between 60-90 fpm with the balance shield closed tightly.
- If connected directly to a house exhaust system, an adjustable damper (or valve) must be installed to control airflow properly.
Specifies power connection for 115V and 230V models and the remote-blower relay circuit for XPert Stations. XPert Systems have no provision for controlling a remote blower; XPert Stations provide a blower switch and relay with a maximum circuit load of 6 amps.
- 115V models: connect supplied power cord to IEC inlet on back of enclosure; at 50 Hz maximum airflow is reduced by 17%.
- 230V models: shipped without a plug; install plug per local codes.
- XPert Station remote blower is switched by an electrical relay via the blower switch; maximum circuit load for the relay is 6 amps.
- Only remote blower wiring is connected inside the electrical junction box; lights and airflow monitor are powered from a nearby standard duplex receptacle.
- Grounding connection shall not be made to the terminal box cover.
- Building supply for the remote blower should include overload protection (switch or circuit breaker) close to and within easy reach of the operator, marked as the disconnecting device.
Describes leveling and securing the work surface, positioning the enclosure, and fastening it with four mounting screws and RTV silicone sealant. The work surface must be smooth, rigid, non-porous and chemical-resistant and at least as wide as the enclosure.
- Front of enclosure should be aligned within 0.3" of the front of the work surface.
- Insert supplied mounting screws in four holes, allowing a minimum of 1/8" clearance under the screw head for positioning.
- For 5' and 6' stainless work surfaces use #10-24 x 3/8" hex washer head screw, Labconco part number 1885806.
- Apply RTV silicone sealant between work surface and enclosure for additional support.
- Seal enclosure to work surface with a bead of silicone sealant to prevent spilled materials collecting under walls and to add rigidity.
Covers duct connections for XPert Stations and optional Canopy Exhaust Connections for XPert Systems. 2', 3' and 4' Stations accept 6" duct; 5' and 6' Stations accept 10" duct. Ductwork weight must be supported independently of the enclosure.
- 2', 3' and 4' XPert Stations: 6.06" ID exhaust connection; 5' and 6': 10.09" ID.
- Canopy Exhaust Connection part numbers: 3924400 (2'), 3924401 (3'), 3924402 (4'), 39294403 (5'), 3924404 (6').
- Canopy connection for 2', 3' and 4' is 6.09" ID / 6.19" OD for 6" nominal duct; 5' and 6' are 10.09" ID / 10.19" OD for 10" nominal duct.
- Exhaust duct material must match the enclosure, procedures and chemical applications.
- Exhaust system should be installed by a qualified HVAC contractor.
HEPA filters ship installed with gasket on the downstream side and are leak-checked at Labconco; a second leak check is recommended before use and at least annually. Carbon filters are gravity-installed with gasket side down on the exhaust side of the HEPA filter and provide odor control only, not particulate protection.
- HEPA filters are 99.99% efficient for 0.3-micron particles.
- HEPA part numbers: 2' 3707900, 3' 3707901, 4' 3707902, 5' 3707903 (2 ea), 6' 3707901 (2 ea).
- Organic vapor carbon: 2' 3937200, 3' 3937300, 4' 3937400, 5' 3981400 (2 ea), 6' 3937300 (2 ea).
- Formaldehyde carbon: 2' 3937201, 3' 3937301, 4' 3937401, 5' 3981401 (2 ea), 6' 3937301 (2 ea).
- Ammonia/amines carbon: 2' 3937202, 3' 3937302, 4' 3937402, 5' 3981402 (2 ea), 6' 3937302 (2 ea).
- Acid-sulfur carbon: 2' 3937203, 3' 3937303, 4' 3937403, 5' 3981403 (2 ea), 6' 3937303 (2 ea).
- Carbon filter use requires work-area concentration not exceed IDLH and exhaust not exceed TWA.
Explains adjusting the speed control on XPert Systems to achieve 60-100 fpm for weighing, and describes airflow validation by anemometer readings across the sash opening. Labconco reports ASHRAE 110-1995 containment ratings and NSF/ANSI 49 biological containment test results.
- Speed control is behind the front panel; turn counterclockwise to increase face velocity and clockwise to decrease.
- Face velocity should be 60-100 fpm for weighing operations; lowest appropriate velocity gives quietest operation.
- Average face velocity readings: two rows, 6" from ends and every 12", first row 3" down from upper sash foil, second row 3" up from work surface.
- As-manufactured ASHRAE 110-1995 rating: less than 0.05 ppm at 4 lpm.
- Biological containment test at 75 fpm inflow typically 1 or fewer spores escaped per test.
- Balance stability tests were excellent at 78 fpm and 100 fpm with a five-decimal-place balance.
Covers operation and programming of the Guardian Digital 1000 airflow monitor, including alarm set point calibration and alarm silencing. The monitor alarms when airflow deviates from the calibrated set point.
- Alarm can be temporarily silenced using the SILENCE ALARM or enter buttons.
- Audible disable requires a continuously present alarm condition; fluctuating flow near the set point resets the monitor.
- Calibration steps are given in Chapter 6 and in the supplied Digital 1000 Monitor Manual.
Provides the routine maintenance schedule, criteria for HEPA replacement, bag-in/bag-out HEPA replacement procedure, HEPA leak test, carbon filter life calculation and replacement, and replacement of LED lights, motorized impeller and speed control.
- HEPA filter leak test recommended before use and at least annually thereafter.
- Bag-in/bag-out HEPA filter allows safe disposal of hazardous filters.
- Carbon filter life must be calculated based on chemical loading; replace when loaded.
- Motorized impeller and speed control are replaceable items described in Chapter 6.
A problem/cause/corrective-action table covering containment, blower, lighting and airflow monitor malfunctions. It lists symptoms such as contaminants outside the enclosure, blower not operating, low face velocity, lights not working, and airflow monitor alarms.
- Contaminants outside enclosure may be caused by improper technique, blocked baffle slots or exhaust outlet, external air disturbances, or improper face velocity.
- Blower won't operate: check power, circuit breaker/GFI, blower wiring, blower switch, motorized impeller.
- Low face velocity or poor containment: close sash, replace clogged HEPA filter or increase speed, replace loaded carbon filter.
- Lights do not work: check lamp installation and wiring, replace lamp, light switch or ballast.
- Airflow monitor malfunction: check power, cables, fuses, alarm set point and sensor airway.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| HEPA filters are rated at least 99.99% efficient for 0.3-micron particles. p.5 ↗ | Suitable for hazardous particulate containment; carbon filters add odor control only and provide no particulate protection. | All XPert and RXPert models |
| Face velocity for weighing operations should be 60-100 fpm; balances typically work best between 60-90 fpm with the balance shield closed tightly. p.10 ↗ | Set speed control within this range and verify with anemometer; lower end gives quieter operation. | XPert Filtered Balance System |
| XPert Station 2', 3' and 4' models require a 6" exhaust connection (6.06" ID); 5' and 6' models require 10" (10.09" ID). p.16 ↗ | Duct sizing and remote blower selection must match enclosure width. | XPert Filtered Balance Station |
| Maximum circuit load for the XPert Station remote-blower relay is 6 amps. p.11 ↗ | Remote blower electrical load must not exceed 6 A on this circuit. | XPert Filtered Balance Station |
| At 50 Hz operation, blower maximum airflow is reduced by 17%. p.16 ↗ | Account for reduced airflow when sizing and setting face velocity on 50 Hz supplies. | 115V models |
| Enclosure front should be aligned within 0.3" of the front of the work surface. p.13 ↗ | Critical for proper airflow entry and sealing. | All models |
| Mounting screws must allow a minimum of 1/8" clearance under the head for positioning the enclosure. p.14 ↗ | Ensures flanges can slide under screw heads during installation. | All models |
| For 5' and 6' stainless work surfaces, use #10-24 x 3/8" hex washer head screw, Labconco part number 1885806. p.14 ↗ | Correct fastener for large stainless work surfaces. | 5' and 6' models with stainless work surface |
| Maintain humidity between 45 and 60% to control electrostatic effects. p.9 ↗ | Reduces weighing errors and powder attraction in the enclosure. | All models |
| Temperature drift of 1-2°C is generally tolerable for balances. p.8 ↗ | Avoid installing near heating sources or direct sunlight. | All models |
| HEPA filter leak test is recommended before use and at least annually thereafter. p.19 ↗ | Scheduled leak testing is required to maintain containment assurance. | All models |
| Carbon filters are gravity-installed with gasket side down on the exhaust side of the HEPA filter. p.19 ↗ | Correct orientation is essential for odor control performance. | Models with accessory carbon filter |
| Carbon filter use requires work-area concentration not exceed IDLH and exhaust not exceed TWA. p.19 ↗ | Chemical application must be reviewed by a Labconco product specialist for suitability. | Models with accessory carbon filter |
| Canopy Exhaust Connection part numbers: 3924400 (2'), 3924401 (3'), 3924402 (4'), 39294403 (5'), 3924404 (6'). p.18 ↗ | Order correct canopy for outside exhaust of XPert Systems. | XPert Filtered Balance System |
| As-manufactured ASHRAE 110-1995 containment rating is less than 0.05 ppm at 4 lpm. p.18 ↗ | Baseline containment performance for field validation comparison. | All models |
| Biological containment testing at 75 fpm inflow typically showed 1 or fewer spores escaping per test. p.19 ↗ | Supports use for biohazardous procedures at appropriate face velocity. | All models |
| Balance stability tests were excellent at 78 fpm and 100 fpm with a five-decimal-place balance. p.19 ↗ | Confirms suitability for high-precision weighing within the recommended velocity range. | XPert Filtered Balance System |
| Each enclosure model weighs between 125 and 350 lbs (55 to 152 kg). p.12 ↗ | Work surface and supporting structure must safely support this weight plus equipment. | All models |
| An adjustable damper (or valve) must be installed when connected directly to a house exhaust system. p.11 ↗ | Required to control airflow properly, especially when multiple enclosures share a house exhaust. | Models ducted to house exhaust |
| Grounding connection shall not be made to the terminal box cover. p.17 ↗ | Electrical installation must follow this grounding restriction. | XPert Filtered Balance Station |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Contaminants outside of enclosure. p.62 ↗ | Improper user techniques for the enclosure. | See “Certifying the Enclosure” Chapter 3 and “Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix D). | Ref. Appendix D |
| Contaminants outside of enclosure. p.62 ↗ | Restriction of the baffle air slots or blockage of the exhaust outlet. | Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. | |
| Contaminants outside of enclosure. p.62 ↗ | External factors are disrupting the enclosure airflow patterns or acting as a source of contamination. | See “Location Requirements” Chapter 2, “Certifying the Enclosure” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix D). | Ref. Appendix D |
| Contaminants outside of enclosure. p.62 ↗ | Enclosure has improper face velocity. | Have enclosure certified and check exhaust system. Check HEPA filters for loading. Adjust enclosure speed control. Enclosure should have an average face velocity of 60-100 fpm for weighing procedures and 75-105 fpm for biohazardous procedures. | |
| Blower won’t operate. p.62 ↗ | Unit not plugged into outlet. | Plug the enclosure into appropriate electrical service. | |
| Blower won’t operate. p.62 ↗ | Circuit breaker(s) or Ground Fault Interrupter. | Reset circuit breaker. | |
| Blower won’t operate. p.62 ↗ | Blower wiring is disconnected. | Inspect blower wiring. | |
| Blower won’t operate. p.62 ↗ | Blower switch is defective. | Replace switch. | |
| Blower won’t operate. p.62 ↗ | Motorized impeller or blower is defective. | Replace motorized impeller or blower. See Chapter 6. | Chapter 6 |
| Low face velocity or poor containment of contaminants. p.62 ↗ | Enclosure sash not closed. | Close sash to the lowest position. | |
| Low face velocity or poor containment of contaminants. p.62 ↗ | HEPA filter clogged. | Replace HEPA filter or increase speed. | |
| Low face velocity or poor containment of contaminants. p.62 ↗ | Carbon filter loaded with chemicals. | Replace carbon filter. | |
| Blower and lights won’t operate. p.62 ↗ | Unit not plugged into outlet. | Plug enclosure into appropriate electrical service. | |
| Blower and lights won’t operate. p.62 ↗ | Circuit breaker(s) tripped. | Reset or replace circuit breaker. | |
| Lights do not work. p.63 ↗ | Lamp not installed properly. | Inspect lamp installation. | |
| Lights do not work. p.63 ↗ | Lamp wiring disconnected. | Inspect lamp wiring. | |
| Lights do not work. p.63 ↗ | Defective lamp. | Replace lamp. | |
| Lights do not work. p.63 ↗ | Light switch is defective. | Replace light switch. | |
| Lights do not work. p.63 ↗ | Defective electronic ballast. | Replace ballast. | |
| Airflow monitor malfunction. p.63 ↗ | No power. No lights. No display. | Power cable to circuit board is disconnected and needs to be connected. Sensor cable needs to be connected. Power supply is not plugged into proper voltage; plug in power supply. Verify that all airflow monitor interface cables are connected. Check fuses on enclosure. | |
| No audible alarm. p.63 ↗ | Alarm has been temporarily silenced using “SILENCE ALARM” or “enter” buttons. | Wrong alarm set point. Airflow monitor was not properly adjusted. Repeat calibration steps outlined in this manual in Chapter 6 or in the supplied Digital 1000 Monitor Manual. | Chapter 6 |
| Constant audible alarm. p.63 ↗ | Check airflow and calibration of airflow monitor. | See Chapter 6. | Chapter 6 |
| Continuous alarm. p.63 ↗ | Check the face velocity of the enclosure as the airflow of the system may have changed. | If incorrect, adjust the speed control to increase face velocity. The HEPA filter may have become loaded. If face velocity is correct, calibrate the airflow monitor outlined in this manual in Chapter 6. | Chapter 6 |
| Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. p.63 ↗ | Airway to airflow monitor sensor is blocked by insects, dust or debris. | Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. | |
| Audible disable will not stay operational. p.63 ↗ | An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. | Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower. |