Canopy Laboratory Hoods Manual
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco Canopy Hoods, models 4799000 (4-foot) and 7099000 (6-foot), as shipped with part numbers 47990 and 70990 respectively. Document revision 4809400 Revision C / ECO E427.
Section-by-Section Technical Breakdown
Lists the carton contents for the Canopy Hood. The shipment includes the hood assembly and the instruction manual. Users are instructed to check for transit damage before discarding packaging.
- Part 47990: 4 Foot Canopy Hood
- Part 70990: 6 Foot Canopy Hood
- Part 48094: Instruction Manual
- Do not discard carton or packing material until all components are checked and equipment is installed and tested
Describes the Canopy Hood as intended for special purpose functions generating heat, steam, or odors, accommodating large or bulky apparatus such as tissue processors, steam baths, or open Kjeldahls. The epoxy-coated superstructure includes an integral baffle to increase slot velocities and capture efficiency. All parts are epoxy-coated except stainless steel baffle support brackets.
- Intended for heat, steam, or odor-generating special purpose functions
- Suitable for large or bulky apparatus (tissue processor, steam baths, open Kjeldahls)
- Epoxy-coated superstructure with integral baffle
- Baffle support brackets are stainless steel
- Not intended for flammable or explosive atmospheres unless equipped with explosion-proof components and installed per electrical codes
Specifies maximum static pressure loss through the hood at given airflow rates for the 4-foot and 6-foot units. The baffle design enhances perimeter air velocities and capture efficiency.
- 4-foot unit: static pressure loss shall not exceed 0.64 inches at 907 CFM
- 4-foot unit: static pressure loss shall not exceed 1.25 inches at 1,360 CFM
- 6-foot unit: static pressure loss shall not exceed 0.90 inches at 1,289 CFM
- 6-foot unit: static pressure loss shall not exceed 1.65 inches at 1,934 CFM
- Baffle design enhances perimeter air velocities and capture efficiency
Recommends mounting the back against a wall with the front supported by 3/8 inch threaded rod from the ceiling; wire or cable may be substituted. Hood effectiveness depends on proximity of the work area to the hood face, with performance based on a theoretical face 1 foot below the hood.
- Back mounted against a wall; front supported by 3/8 inch threaded rod to ceiling
- Wire or cable may be substituted for 3/8 inch threaded rod
- Face velocity and static pressure based on theoretical hood face at 1 foot below the hood
- Efficiency drops if working apparatus is below the 1-foot face; increases as apparatus enters the 1-foot face
- Distance between work surface and hood is extremely important for maximum efficiency
Section heading for dimensional information; the document contains a dimensions section but no numeric values are present in the extracted text.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| 4-foot unit static pressure loss shall not exceed 0.64 inches at 907 CFM. p.5 ↗ | Use this limit when sizing exhaust blower and ductwork for the 4-foot hood. | 4799000 / 47990 (4 Foot Canopy Hood) |
| 4-foot unit static pressure loss shall not exceed 1.25 inches at 1,360 CFM. p.5 ↗ | Defines the upper airflow operating point and required fan static pressure for the 4-foot hood. | 4799000 / 47990 (4 Foot Canopy Hood) |
| 6-foot unit static pressure loss shall not exceed 0.90 inches at 1,289 CFM. p.5 ↗ | Use this limit when sizing exhaust blower and ductwork for the 6-foot hood. | 7099000 / 70990 (6 Foot Canopy Hood) |
| 6-foot unit static pressure loss shall not exceed 1.65 inches at 1,934 CFM. p.5 ↗ | Defines the upper airflow operating point and required fan static pressure for the 6-foot hood. | 7099000 / 70990 (6 Foot Canopy Hood) |
| Face velocity and static pressure performance is based on a theoretical hood face at 1 foot below the hood. p.6 ↗ | Mount the hood so the work area is at or within 1 foot of the hood face to achieve rated capture efficiency. | whole series |
| Efficiency drops off if the working apparatus height is below the theoretical 1-foot face; efficiency increases as apparatus enters inside the 1-foot face. p.6 ↗ | Keep apparatus as close to the hood face as practical; avoid low work surfaces. | whole series |
| Recommended installation: back mounted against a wall, front supported by 3/8 inch threaded rod mounted to the ceiling. p.6 ↗ | Plan ceiling anchoring and wall placement for the 3/8 inch rod support. | whole series |
| Wire or cable may be substituted for the 3/8 inch threaded rod. p.6 ↗ | Provides an alternative front suspension method where threaded rod is impractical. | whole series |
| All component parts are epoxy-coated except the baffle support brackets, which are stainless steel. p.5 ↗ | Account for mixed materials when assessing chemical compatibility and cleaning. | whole series |
| Canopy Hoods are not provided with explosion-proof electrical components; flammable or explosive atmospheres should not be used unless explosion-proof components are fitted and installed per all pertinent electrical codes. p.8 ↗ | Do not install in classified hazardous locations without a certified explosion-proof configuration. | whole series |