Introduces Bond Elut Lipid Extraction product line
Procedurep. 2
Recommended protocol steps for lipid extraction
p. 2table, 3 rows
Bond Elut Lipid Extraction ordering information
Description
Quantity
Part Number
Agilent Bond Elut Lipid Extraction, 1 mL cartridge
100/pk
5610-2041
Agilent Bond Elut Lipid Extraction, 96-well plate
1 plate
5610-2042
Agilent Bond Elut Lipid Extraction, 96-well plate
5 plates
5610-2043
Engineering Facts
12 facts · every claim linked to a page
Selection
The 1 mL cartridge format is intended for small batch sample preparation, so it should be chosen only when sample throughput is low.
Determines whether this cartridge format matches the expected batch size and workflow scale.
Source
“Bond Elut Lipid Extraction 1 mL cartridges are suitable for small batch sample preparation.”
Operation
The crashing solvent must be ACN with 1 to 5% MeOH to produce finer protein precipitates and enable pipette transfer.
Defines the required precipitation solvent composition before loading the cartridge.
Source
“Crashing solvent: ACN with 1 to 5% MeOH. A small percentage of MeOH helps to generate finer protein precipitates, which allows for easy pipette transfer of the homogenate.”
Operation
The wash solvent must be ACN with 10 to 20% water.
Specifies the wash composition needed to remove matrix interferences without eluting retained lipids.
Source
“Washing solvent: a mixture of ACN with 10 to 20% of water.”
Operation
The elution solvent must contain at least 50% MeOH to release lipids, with DCM/MeOH 1:2 or chloroform/MeOH 1:1 recommended.
Sets the minimum organic modifier content and preferred solvent ratios for complete lipid elution.
Source
“A minimum of 50% MeOH is important for the release of lipids. Agilent recommends DCM/MeOH (v/v 1:2) or chloroform/MeOH (v/v 1:1).”
Operation
The sample mixture must contain at least 10% aqueous for proper loading onto the EMR-Lipid sorbent.
Prevents over-drying or incorrect solvent composition that would impair sorbent retention during loading.
Source
“The EMR—Lipid sorbent requires a minimum of 10% aqueous in the sample mixture for loading.”
Operation
The entire homogenate must be transferred without centrifugation because lipids can be trapped in the protein precipitate.
Avoids loss of lipid analytes that would otherwise remain in the discarded precipitate.
Source
“It is important to transfer the entire homogenate, as lipids can be trapped in the protein precipitate. Do not centrifuge before transferring.”
Operation
A slow flow rate of approximately 3 to 5 seconds per drop should be maintained during cartridge operation.
Controls residence time for adequate lipid retention and reproducible extraction.
Source
“Maintain a slow flow rate of approximately 3 to 5 seconds per drop.”
Operation
After washing, higher vacuum or pressure must be applied to dry the cartridge before elution.
Ensures residual wash solvent is removed so it does not dilute or interfere with the elution step.
Source
“After the wash has eluted, apply higher vacuum or pressure to dry the cartridge.”
Operation
The eluate must be dried under nitrogen at 30 °C.
Specifies the evaporation temperature to avoid thermal degradation of lipids during concentration.
Source
“Dry the eluent with N2 at 30 °C.”
Operation
Wide bore pipette tips are recommended for transferring the homogenate.
Reduces clogging and shear of the protein-rich homogenate during transfer to the cartridge.
Source
“Wide bore pipette tips are recommended for transfer.”
Operation
The reconstitution solvent is n-BuOH/MeOH (v/v, 1:1) at 100 µL, followed by vortexing and sonication.
Defines the final sample solvent and volume required before MS analysis or storage.
Source
“Reconstitute into 100 μL of n-BuOH/MeOH (v/v, 1:1), vortex for two minutes, and sonicate for 10 minutes.”
Operation
The final extract can be stored in a –20 °C freezer if not analyzed immediately.
Provides the storage condition to preserve extract stability before analysis.
Source
“Transfer to a 2 mL vial with an insert, and cap for MS analysis or store in a –20 °C freezer.”