ASI 35 — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ASI 35 integrable industrial helium/hydrogen leak detector (part number pattern Sxxx0x0xMM9A, all models). Modules: vacuum module + ½-rack 19" 3U electronic module; optional ½-rack industrial control panel. Options/accessories: 'Internal calibration' kit, sniffer kit (standard and Smart sniffer probes), spray gun, ASI 20 MD module, communication interface. Interconnecting cables: vacuum-module/electronic-module 15-pin and 25-pin D-Sub in lengths 1.5 m / 3.5 m / 5 m / 10 m; control-panel cable 1.8 m / 5 m / 10 m. Related doc refs: ASI 20 MD/ASI 30/ASI 35 specific features (annex), ASI 35 Maintenance instructions 127801M.

Section-by-Section Technical Breakdown

pp. 17-21
Product Description, Connection Interface and Test Methods

Describes the three-module architecture (vacuum module with SplitFlow 50 turbo pump and analyzer cell; ½-rack electronic module; optional ½-rack industrial control panel), the connector/label layout, and the two test methods (hard vacuum test: spray or bombing; sniffer test). Lists full scope of delivery including mains cables (EU + US), ½-rack plate, 4 fixing brackets, storage-valve lever and sniffer adapter.

  • Inlet ports: Gross Leak 2 x DN 16 ISO-KF; Normal mode 1 x DN 16 ISO-KF; High Sensitivity 1 x DN 16 ISO-KF + 1 x DN 25 ISO-KF.
  • Detection module does not include backing pumping (primary pump ordered separately).
  • Electronic/vacuum module cables: 15-pin and 25-pin D-Sub, length per option; module-to-panel cable lengths 1.8 / 5 / 10 m.
  • Response time in spray method: T = V/S (s, liters, l/s) — time to reach 63% of final leak-rate value.
  • Sniffer measurement is not an exact leak value; result depends on probe distance and leak direction.
  • Control panel: START/STAND-BY key, function keys, SD card slot, remote-control and buzzer connectors.
pp. 22-27
Installation — Securing the Modules

Mounting instructions for the vacuum module (any position, 4-bracket M8 attachment), electronic module (any position, brackets or rubber feet, ½-rack plate) and the optional industrial control panel. Requires integration into a customer system with protection index ≥ IP4x; the detector must not be used standalone on a workbench. Ventilation clearances and rack-format precautions are specified.

  • Vacuum module attachment must absorb a braking torque of 620 N·m; mount with M8 socket-head screws and Ø8 mm washers.
  • Leave 10 cm free space around the vacuum module; 2 cm minimum under the electronic module fan.
  • Electronic module and industrial control panel must be attached separately to the customer's system — never fixed together via a plate in 1-rack format.
  • System incorporating the leak detector must have protection index at least IP4x.
  • Mains switch acts as emergency stop; keep access free.
  • Effective integration volume of vacuum module: 198 × 264 × 279 mm.
pp. 27-35
Installation — Connecting the Installation to be Tested (Vacuum Circuit)

Detailed connection schematics for Gross Leak, Normal and High Sensitivity mode inlet ports plus sniffer mode and primary-pump connection. Specifies pipe sizing, primary pump characteristics, and safety rules (pressure limits, hydrogen mix, single-port connection, inlet filtering). Calibration with the 'Internal calibration' option requires the corresponding test valve closed during calibration.

  • Primary pump: wet or dry, DN 16 ISO-KF connection, hose max 2 m, ultimate vacuum <1 hPa (5·10⁻² hPa recommended), min flow 1 m³/h (3 m³/h with Smart probe in sniffer), max 100 m³/h.
  • Connected parts must withstand a negative pressure of 1·10³ hPa relative to atmospheric pressure.
  • Backing pump must always be connected to one of the two Gross Leak mode inlet ports.
  • Never connect the test part to several inlet ports simultaneously; never exceed 10 kg load on an inlet port.
  • Pressure inside the product must never exceed 100 kPa absolute; pressure at detector exhaust must never exceed 18 hPa.
  • Use flexible (not rigid/plastic compressed-air style) pipework with diameter equal to the detector inlet, as short as possible.
  • Install an inlet filter for applications generating particles.
pp. 38-54
Commissioning and Operation (Start-up, Test Start/Stop, Calibration, Touchscreen)

Covers detector start-up and power-off, control panel familiarization, test start/stop, operation monitoring, and the full range of calibration procedures (internal/external calibrated leak, machine calibration, parallel pumping system, sniffer concentration and dynamic calibration, calibration check). Also describes the touchscreen interface: navigation, main screen, graph screen (parameters, scales, recording/save/view), settings, synoptic and measurement windows.

  • Calibration variants: internal calibrated leak, external calibrated leak, machine calibration, with parallel pumping system (hard vacuum); external leak and concentration (sniffer); dynamic calibration.
  • Zero function available for background suppression; Memo function for result storage.
  • Graph screen supports recording, saving and reviewing curves with adjustable scales.
pp. 55-85
Settings Menus (Set Points, Test, Spectro, Maintenance, Configuration, Advanced)

Complete menu structure of the detector: audio alarm/reject points (hard vacuum, sniffer, probe clogged, background max, pollution), test parameters (method, correction factor, test mode, probe type, automatic cycle end), spectro settings (tracer gas 4He/3He/H2, filament, calibrated leak), maintenance timers and histories (events, calibration, pump info), configuration (time/date/language, passwords, function keys), and advanced settings (I/O, serial links, secondary pump speed, SD card, service).

  • Tracer gas selectable: He (4He/3He) or H2 — modes labelled He_PU/He_MU, H2_PU/H2_MU on the electronic module label.
  • Set points include: hard vacuum reject point, sniffer reject point, probe clogged threshold, background maximum, pollution function.
  • Advanced menu: Start Up timer, detector/external pressure gauge, analyzer cell, signal processing, serial links 1 & 2, I/O connector, secondary pump speed.
  • Password protection and function-key assignment configurable.
  • Default settings tables provided in the annex (Tbl. 3–9, pages 102–112).
pp. 86-92
Troubleshooting Guide

A dedicated chapter guiding the user through diagnosis of detector faults; complements the event history stored in the Maintenance menu.

  • Event history and calibration history retrievable via Maintenance menu for diagnostics.
pp. 93-96
Maintenance, Service and Accessories

Refers to the separate ASI 35 Maintenance instructions (part no. 127801M) for maintenance/replacement work; lists Pfeiffer Vacuum service solutions and the accessories catalogue with ordering references (sniffer kits, probes, spray gun, internal calibration kit, inlet filter, communication interface, gauge).

  • Maintenance instructions document: ASI 35 Maintenance instructions, part number 127801M.
  • Accessories include inlet filter, sniffer kit (standard probe 121780 / Smart probe BG5268), spray gun 121781, internal calibration kit 123588.
pp. 97-99
Technical Data, Units, Gas Throughputs and Dimensions

General and technical characteristics of the ASI 35, pressure unit and gas throughput conversion tables, and dimensional drawings of the three modules. Rating plate data: supply 90/240 V, 50–60 Hz, max 300 W (per module labels).

  • Operating voltage 90–240 V, 50–60 Hz, maximum power consumption 300 W (module label).
  • Pressure unit and gas throughput conversion tables included (Tbl. 1–2).
  • Dimensional drawings for vacuum module, electronic module and control panel; module drawings also on the documentation USB.
pp. 100-114
Annex: ASI 20 MD / ASI 30 / ASI 35 Specific Features and Default Settings Tables

Describes model-specific differences among ASI 20 MD, ASI 30 and ASI 35, provides the full settings menu tree, and default settings tables for all menus plus initial graph parameters. Includes EC and UK Declarations of Conformity.

  • Settings menu tree diagram (pages 101–112) with default values for Set Points, Test, Spectro, Maintenance, Configuration and Advanced menus.
  • EC and UK Declarations of Conformity included (pages 113–114).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Tracer gas with hydrogen content must never exceed 5%; recommended gas is 95% N2 / 5% H2 (hydrogenated nitrogen). p.13 ↗ Specify a 5% H2/N2 mix supply for hydrogen leak testing to stay within safe/ex合规 limits; pure H2 is prohibited. All ASI 35 models
Primary pump requirements: wet or dry, DN 16 ISO-KF, hose ≤2 m, ultimate vacuum <1 hPa (5·10⁻² hPa recommended), flow 1–100 m³/h; ≥3 m³/h required for sniffer with Smart probe. p.29 ↗ Size the backing pump within this window — flows above 100 m³/h give no performance benefit; undersized flow degrades sniffer operation. ASI 35 all models (Smart probe sniffer option specifically)
Pressure at the detector exhaust must never exceed 18 hPa; pressure inside the product must never exceed 100 kPa absolute. p.28 ↗ Exhaust line design must prevent back-pressure buildup (long/damped exhaust lines) to avoid damage or loss of performance. ASI 35 all models
Parts/installations connected to the detector inlet must withstand a negative pressure of 1·10³ hPa relative to atmospheric pressure. p.27 ↗ Verify mechanical strength of the test part/vacuum chamber before hard vacuum testing. All test configurations
Vacuum module attachment must absorb a braking torque of 620 N·m; fixing with M8 socket-head screws and Ø8 mm washers at 4 bracket points. p.22 ↗ Design the integration frame/brackets to withstand 620 N·m reaction torque from the SplitFlow 50 turbo pump. Vacuum module, all models
Vacuum module can be mounted in any position; 10 cm free space required around it; 2 cm clearance under the electronic module fan. p.22 ↗ Flexible machine integration possible, but ventilation clearance must be respected to prevent overheating. All modules
Maximum load on an inlet port is 10 kg; never connect the test part to multiple inlet ports; backing pump must always connect to one of the two Gross Leak ports. p.28 ↗ Pipework to the detector inlet must be independently supported; reserve one Gross Leak port for the forepump. ASI 35 all models
Inlet ports: Gross Leak 2 x DN 16 ISO-KF; Normal 1 x DN 16 ISO-KF; High Sensitivity 1 x DN 16 ISO-KF + 1 x DN 25 ISO-KF. p.18 ↗ Match connection flange sizing per desired mode: DN 25 on High Sensitivity gives highest tracer gas pumping speed/sensitivity; Gross Leak gives best analyzer cell protection. ASI 35 vacuum module
Response time in spray method follows T = V/S (T in s, V in liters, S in l/s) — time to reach 63% of final value. p.20 ↗ Estimate test cycle time from test volume and effective pumping speed; larger volumes require proportionally longer signal settling. Hard vacuum test, all models
Electronic module/vacuum module cables and control panel cable are selectable: module cables 1.5 / 3.5 / 5 / 10 m; panel cable 1.8 / 5 / 10 m. p.17 ↗ Choose cable lengths at ordering time to allow remote panel mounting; maximum separation up to 10 m. Ordering options, all models
Power supply: 90–240 V, 50–60 Hz, maximum power consumption 300 W. p.8 ↗ Single-phase mains supply sizing; must be earthed — detector must not be operated with ungrounded power cable. Electronic module, all models
Wait 5 minutes after switching mains to O before working on or opening the product (live components remain charged). p.12 ↗ Include a 5-min lockout delay in maintenance procedures/lockout-tagout planning. All models
Pacemaker safety distance: at least 130 mm from the product due to magnetic fields. p.12 ↗ Personnel with implanted pacemakers must keep >130 mm distance or use field shielding. Vacuum module (magnetic field source), all models
System incorporating the leak detector must achieve protection index at least IP4x; the detector is a fuse-type detector intended for integration, not standalone bench use. p.22 ↗ The integrator is responsible for enclosing/protecting the detector to IP4x within the machine. ASI 35 all models
Purge/neutral gas circuit requires a manual lockout valve installed at max 3 m from the product. p.13 ↗ Add a manual shut-off valve in the purge gas line during machine integration for safe servicing. All models using purge gas
Blanking plate with manual storage valve (1/8" female connection) must be removed before use; lever must be removed for transport; module should be stored under vacuum via this valve. p.15 ↗ Retain the storage valve assembly; evacuate the vacuum module before long-term storage to protect performance and shorten degassing on restart. Vacuum module, all models
Calibration with the 'Internal calibration' option requires the corresponding test valve (1/2/3 for Gross Leak/Normal/High Sensitivity) closed during calibration. p.30 ↗ Automated test sequences must interlock test valves during calibration phase to get valid calibration. ASI 35 with Internal calibration option
Inlet filter recommended when the application generates particles; testing of soiled/wet parts (water, vapor, paint, detergent) is forbidden misuse. p.14 ↗ Add an inlet filter to the accessories order for dusty applications; implement part-cleanliness checks upstream. All models
Tracer gases selectable in Spectro menu: helium (4He/3He) and hydrogen (H2); test modes include hard vacuum and sniffer with standard or Smart probe. p.60 ↗ One instrument covers He and H2 testing; select tracer gas and probe type via [Test]/[Spectro] menus per application. ASI 35 all models (sniffer option for probe use)
Maintenance work requires the separate ASI 35 Maintenance instructions (part no. 127801M); opening the warranty-sealed pump body voids the warranty. p.7 ↗ Schedule maintenance per the dedicated manual and contact service for intervals shorter than warranty period — do not open the pump in-house during warranty. ASI 35 vacuum module