ASM 310 — instruction manual
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum ASM 310 helium/hydrogen leak detector, part number BSAA0200MM9A (single model; manual covers all models/variants, some features optional). Accessories referenced but sold separately: standard sniffer probe (121780), spray gun (121781), RC 10 remote control (124628), maintenance instructions (128864M), communication interface instructions (130417), transport case and cart.
Section-by-Section Technical Breakdown
Identifies the product (rating plate, labels) and lists scope of delivery. Describes all connectors: inlet, USB, purge/vent, exhaust, sniffer, RC 10, RS-232, 15-pin I/O, power and main switch. Control panel has touchscreen, START/STOP, CAL and ZERO buttons.
- Scope of delivery: detector, documentation (USB stick), mains cable (EU and/or US), internal calibrated leak certificate, QC certificate, maintenance kit, 15-pin D-Sub connector and cover.
- Connectors: SNIFFER probe (pneumatic), VENT/N2 PURGE, EXHAUST (with filter, never remove), RS-232 9-pin D-Sub, 15-pin D-Sub I/O, RC 10 remote control, USB stick port.
- CAL button launches internal calibration, external calibration or calibration check depending on assigned [CAL. TYPE] function key.
- Product labels include grounding point, firmware (name/version/checksum/date), rating plate (weight, voltage, frequency, max power, P/N, S/N).
Installation requirements: flat horizontal surface, ≥10 cm free space all around, ventilated area free of tracer gas, inlet pressure not above atmospheric. Covers purge gas connection, exhaust line and electrical connection (Class I, must be earthed). Defines mechanical limits for parts connected to the inlet.
- Purge gas: dry, filtered neutral gas (e.g. N2) at 200 hPa relative excess pressure, flow 5 sccm; use pressure 0–0.3 bar relative (1–1.3 bar absolute). Excess pressure may keep the inlet vent valve closed.
- Exhaust pressure must not exceed 200 hPa relative; exhaust line must be under slightly negative pressure; exhaust filter must never be removed or obstructed.
- Maximum weight at detector inlet: 5 kg; maximum tightening torque: 3 N·m.
- Connected parts must withstand a negative pressure of 1·10³ hPa relative to atmospheric.
- Never use rigid or flexible plastic pipework (compressed-air style tube); use shortest possible sealed pipes with inlet-port diameter.
- Use shielded cables for interfaces in interference-prone environments (EMC).
Start-up sequence: connect mains cable, set main switch to I, set language/unit/date/time on first start, wait for Stand-by mode. Power-off requires switch to O, cable disconnection, and 5 minutes wait. After mains failure the detector restarts automatically when power is restored.
- First start-up requires setting language, unit, date and time.
- Detector automatically switches on again after a mains power failure.
- Wait 5 minutes after power-off before maintenance, cover removal or moving the detector.
Two test methods: Hard Vacuum (spray or bombing method) and Sniffer (with probe accessory). Describes test start/stop procedure, set point configuration and fault notification types on the control panel (warning, error, critical error E244).
- In Stand-by mode the displayed leak rate corresponds to the detector's background.
- Critical error 'E244' requires contacting the service center.
- Hard vacuum test: vent via [VENT] key, connect part, start with START/STOP, spray tracer gas or use bombing (pressurization chamber) method.
- Sniffer test: connect sniffer probe accessory, run probe slowly over suspect areas; display shows quantitative leak rate.
- Before commissioning: check connections, verify tracer-gas-free environment, perform calibration, test only clean dry parts.
Calibration ensures correct tracer-gas response and leak-rate display. Internal calibrated leak is 4He, range ≈10⁻⁷ mbar·l/s, with temperature sensor and integrated diaphragm; other tracer gases (4He Mass 3, H2) require external calibration. Recommended at least daily or each 8-hour shift; detector prompts calibration 20 minutes after switch-on.
- Internal calibrated leak: 4He only, range ≈10⁻⁷ mbar·l/s (≈10⁻⁸ Pa·m³/s), temperature-compensated, supplied with certificate.
- External calibrated leak must contain the selected tracer gas (4He, Mass 3 or H2); manufacturer does not supply Mass 3 or H2 leaks; choose leak in same range as the leak to be measured.
- To abort a calibration: press the CAL button 3 times within 5 seconds.
- Calibration possible only in Stand-by mode; calibration types: Internal, External, Check, and Concentration (sniffer only).
- Sniffer external-leak calibration uses DN 16 ISO-KF or DN 25 ISO-KF adaptor (Standard sniffer probe only); wait at least 10 s before reading the leak rate.
- Concentration calibration uses a known tracer-gas volume mixture at atmospheric pressure, sniffer mode only.
Full settings structure: Measurement (tracer gas, set points, correction factor, calibrated leak, target value), Test (method, mode, probe type, cycle end, inlet vent, memo, zero, regeneration, massive mode, calibration check/function, start-up delay), Configuration (units/date/time/language, sound, function keys, screen, password), Maintenance (history, counters, turbo pump/cell, burn-in, Pirani calibration, parameter save/load), File Manager and Advanced (I/O, service). Default settings tables are in the appendix.
- Tracer gas selection, reject/warning set points, correction factor and target value configurable in [Measurement] menu.
- Test menu includes 'Massive Mode', 'Regeneration', 'Pollution' function (recommended with purge) and inlet vent valve (Operator setting).
- Maintenance menu: history, last operations, counters before next maintenance, turbo pump and cell maintenance, burn-in, internal Pirani gauge calibration, save/load of detector parameters.
- Access control via password in Configuration menu; function keys user-assignable (e.g. [CAL. TYPE], [SWITCH SETPOINT]).
- Appendix contains default-settings tables for all menus (Tbl. 3–8) and function key/graph initial settings (Tbl. 9–10).
Dedicated troubleshooting chapter for fault diagnosis during operation. Control panel notifies user of warnings, errors and critical errors (E244 → contact service center).
- Warnings: press [i Next] to display fault.
- Errors: message displayed, press [i Next] for detail.
- Critical error E244: contact Pfeiffer Vacuum service center.
Maintenance operations referenced to a separate maintenance instruction manual (part number 128864M). Accessories chapter lists original-manufacturer accessories (sniffer probe, spray gun, RC 10 remote, transport case/cart, inlet filter, ISO-KF calibration adaptors, spare internal calibrated leak).
- Maintenance instructions for ASM 310: part number 128864M.
- Only original manufacturer accessories/spare parts permitted; use of others is foreseeable misuse.
- Recommended: inlet filter when the application generates solid particles; spare internal calibrated leak for intensive use.
Chapter 12 contains general data, technical characteristics, pressure unit conversion table, gas throughput conversions and dimensional drawings. Specific numeric values are in the chapter tables (not fully quoted in the extract).
- Includes pressure-unit conversion table (Tbl. 1) and gas-throughput conversion table (Tbl. 2).
- Dimensional data for installation planning (see 12.5).
- Environmental conditions and power supply data referenced from 'Technical characteristics' (12.2).
Appendix describes the 15-pin D-Sub Input/Output interface (cable characteristics, interface functions, save, setting, formulas) and the RS-232 serial link via 9-pin D-Sub (cable characteristics, interface, setting). A separate manual (130417) covers the communication interface in depth.
- 15-pin D-Sub I/O interface: cable characteristics, save, setting and formula functions documented (13.2).
- RS-232 serial link via 9-pin D-Sub connector (13.3).
- Communication interface operating instructions: part number 130417.
EC Declaration of Conformity included with the operating instructions; UKCA Declaration of Conformity also included.
- EC and UKCA Declarations of Conformity ship with the product.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Only tracer gas with hydrogen concentration ≤5% permitted; recommended mixture is 95% N2 / 5% H2 (hydrogenated nitrogen). p.12 ↗ | Standard forming-gas (5% H2) can be used safely; any richer H2 mix is misuse and voids warranty/safety protection. | whole series |
| Internal calibrated leak is 4He, in the range ≈10⁻⁷ mbar·l/s (≈10⁻⁸ Pa·m³/s), temperature-compensated with integrated diaphragm. p.27 ↗ | Factory calibration is 4He; applications with reject points far from 10⁻⁷ mbar·l/s require an external calibrated leak in the same leak-rate range. | whole series |
| Calibration recommended at least once a day (or at start of each 8-hour shift for intensive use); detector prompts calibration 20 min after switch-on. p.27 ↗ | Plan calibration into shift procedures to maintain measurement accuracy. | whole series |
| Purge neutral gas supply: 200 hPa relative excess pressure, 5 sccm flow (at 1 bar abs inlet), use range 0–0.3 bar relative (1–1.3 bar absolute). p.21 ↗ | Regulate N2 supply to this window; excessive pressure keeps the inlet vent valve closed and purge fails to reduce background. | whole series |
| Manual lockout valve required on the purge gas circuit within 3 m of the product. p.11 ↗ | Specify a manual shutoff valve during installation design for pressurized-gas safety compliance. | whole series |
| Exhaust pressure must not exceed 200 hPa relative; exhaust line must be under slightly negative pressure and the exhaust filter never removed or obstructed. p.22 ↗ | Any ducted exhaust into a customer installation needs sizing to avoid backpressure exceeding 200 hPa. | whole series |
| Maximum permitted weight at the detector inlet: 5 kg; maximum torque: 3 N·m; connected parts must withstand −1·10³ hPa relative to atmospheric. p.23 ↗ | Support heavy test parts externally and hand-tighten/limit clamping torque on the inlet flange to avoid damage. | whole series |
| Inlet pressure must never exceed atmospheric pressure. p.20 ↗ | Never connect pressurized parts directly to the inlet; bombing-method parts must be vented to the test chamber, not the detector. | whole series |
| Maintain ≥10 cm free space around the detector; do not obstruct/clean regularly the ventilation grids; store nothing under the detector. p.20 ↗ | Reserve installation footprint with clearance to prevent internal overheating. | whole series |
| Wait 5 minutes after power-off before any maintenance, cover removal or moving the detector (live parts between mains input and circuit breaker). p.11 ↗ | Add a 5-min lockout step to service procedures; do not use ungrounded power cables. | whole series |
| Unpacking/hoisting requires 3-strand strap, each strand >500 mm long, rated >100 kg per strand. p.13 ↗ | Specify lifting gear ratings before receipt/installation; product handled by 2 people via handle or lifting device. | whole series |
| Maximum incline for moving the product: 3° (6%); transport cart load limit 26 kg. p.16 ↗ | Ramps and floor slopes above 3° require alternative handling to prevent tipping/dragging. | whole series |
| Other tracer gases (Mass 3 helium isotope, H2) require external calibration with an external leak of that gas; manufacturer does not supply Mass 3 or H2 calibrated leaks. p.27 ↗ | For H2 or Mass-3 testing, source a suitable external calibrated leak before commissioning. | whole series |
| To abort an active calibration press the CAL button 3 times within 5 seconds. p.29 ↗ | Operator escape procedure if a calibration is started in error. | whole series |
| Sniffer-mode external calibration requires DN 16 ISO-KF or DN 25 ISO-KF adaptor fitted to the calibrated leak (Standard sniffer probe only); wait ≥10 s after removing probe before reading. p.30 ↗ | Include the ISO-KF adaptor and clamp/centering ring in the calibration kit for sniffer applications. | whole series with Standard sniffer probe |
| Critical error code E244 requires contacting Pfeiffer Vacuum service; warnings and errors are viewable via [i Next]. p.25 ↗ | Field troubleshooting limited to warnings/errors; E244 is service-level only. | whole series |
| Never use rigid or flexible plastic (compressed-air style) pipework to connect test parts. p.23 ↗ | Elstp_lastic tubing outgasses/permeates and corrupts leak readings; use metal vacuum pipework sized to the inlet diameter. | whole series |
| Product is Class I equipment and must be earthed; only original manufacturer mains cables may be used. p.22 ↗ | Verify grounding at installation; order replacement cables only per maintenance manual (128864M). | whole series |
| The detector auto-restarts after a mains power failure once power is restored. p.24 ↗ | Design test-cell safety interlocks assuming the detector can come back online unattended after an outage. | whole series |
| Communication options: RS-232 via 9-pin D-Sub and a 15-pin D-Sub I/O interface; dedicated manual part number 130417. p.22 ↗ | PLC/remote integration uses these interfaces; use shielded cables in EMC-critical environments. | whole series |