ASM 340 — instruction manual
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum ASM 340 helium/hydrogen leak detector, covering three variants: ASM 340 Wet (rotary vane 15 m³/h pump, part numbers JSVA02AxMx9x), ASM 340 Dry (diaphragm pump, KSBA02AxMM9A), and ASM 340 Integrable (without backing pump, MSXA02AxMM9A). Options include Smart/Standard sniffer probes, purge/vent lines, communication interfaces (RS-232, USB, Ethernet, PROFIBUS DP-V1, PROFINET, EtherNet/IP), RC 10 remote control, external oil mist separator.
Section-by-Section Technical Breakdown
Identifies the three variants (Wet 15 m³/h rotary vane, Dry diaphragm, Integrable without backing pump) and lists the delivery contents. Describes the connection interface (inlet, exhaust, sniffer/purge connectors, I/O, serial, network ports) and control panel (touchscreen, START/STOP, CAL, ZERO buttons).
- Delivery includes calibration certificate for internal calibrated leak, maintenance kit, lifting handles, power cable(s)
- Wet model delivery additionally includes funnel, 1-liter oil can, draining connector
- Storage box maximum permitted weight: 5 kg
- Communication options: RS-232 D-Sub 9, USB, Ethernet, PROFIBUS DP-V1, PROFINET, EtherNet/IP
Installation on a flat horizontal surface with ≥10 cm free space all around and unobstructed ventilation grids. Covers oil filling (Wet), purge/neutral gas connection, exhaust connection, electrical connection, test part connection, backing pump connection (Integrable) and external oil mist separator installation.
- Neutral gas purge: 200 hPa relative excess pressure, 50 sccm flow at 1 bar abs; usable range 0–0.3 bar relative (1–1.3 bar abs)
- Exhaust pressure must not exceed 200 hPa (relative); Dry model exhaust filter must never be removed
- Maximum permitted weight at detector inlet: 15 kg; maximum torque: 10 N·m
- Connected parts must withstand a negative pressure of 1·10³ hPa relative to atmospheric
- Never use rigid or flexible plastic pipework (compressed-air style tube) at the inlet
- Backing pump (Integrable): DN 25 ISO-KF, hose ≤2 m, ultimate vacuum <1 hPa (5·10⁻² hPa recommended), min flow 1 m³/h (3 m³/h with Smart sniffer probe), max 100 m³/h
- Only manufacturer-recommended oil may be used in the Wet model pump
Start-up sequence (cable, switch to I, language/unit/date/time on first start, wait for Stand-by), optional automatic test launch after start-up, and shutdown procedure. Detector auto-restarts after a mains power failure.
- Power-off: switch to O, unplug mains cable, wait 5 minutes before any work
- After mains failure the detector shuts down and switches on again automatically when power is restored
- 'Run Test after start?' checkbox appears at each power-up; last selection is not saved
Use conditions (hydrogen ≤5%, ventilated tracer-gas-free area, inlet filter for particle-laden applications), prerequisites for optimal use, fault signaling (warning / error / critical error E244), and test start/stop procedures for hard vacuum (spray and bombing methods) and sniffer tests.
- Critical faults display 'Critical error - E244'; contact service center
- In Stand-by mode the displayed leak rate corresponds to the detector's background
- Two test methods: Hard Vacuum and Sniffer; hard vacuum supports spray and bombing methods
- Remote test start possible via RC 10 remote control accessory
The Settings menu comprises six sub-menus covering tracer gas selection, set points, correction factor, calibrated leak settings, test method/mode, cycle end, purge valve, crossover pressures, units/language/password, maintenance history and counters, file manager, and I/O/service. Default settings are tabulated in the appendix (Tbl. 3–10).
- Measurement: tracer gas, set points, correction factor, calibrated leak settings, target value
- Test: method (hard vacuum/sniffer), mode, probe type, cycle end, inlet vent, memo, zero, regeneration, massive mode, calibration check/dynamic calibration, purge valve, start-up time delay, crossover pressures
- Maintenance: history, counters, turbo pump/cell maintenance, burn-in, internal Pirani calibration, parameter save/load
- Appendix provides menu tree (p.99) and default settings tables (pp.100–111)
A troubleshooting guide (chapter 9) and maintenance/replacement procedures (chapter 10) are provided for operational faults and consumable replacement.
- Troubleshooting guide: pages 87–93
- Maintenance/Replacement chapter: page 94; detailed procedures in Maintenance instructions doc. 128863M
Chapter 11 lists accessories and chapter 12 gives general data, technical characteristics, pressure unit conversion and gas throughput conversion tables, and dimensions. Appendix includes conformity declarations (UL/CSA, EC, partly completed machinery, UKCA).
- Technical data: pages 96–98; dimensions: page 97–98
- Included declarations: UL/CSA, EC Declaration of Conformity, EC Declaration of incorporation of partly completed machinery, UKCA
- Related accessory documents: sniffer probes (121780, BG5268B), spray gun (121781), RC 10 (124628), HLT I/O module (122864), communication interface (130417)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Maximum permitted hydrogen concentration in the tracer gas is 5%; recommended mix is 95% N2 / 5% H2 (hydrogenated nitrogen) p.12 ↗ | Forming gas must be specified for hydrogen-mode leak testing to stay outside explosive limits | All ASM 340 models |
| Backing pump for Integrable model: DN 25 ISO-KF connection, hose max 2 m, ultimate vacuum <1 hPa (5·10⁻² hPa recommended), min flow 1 m³/h (3 m³/h with Smart sniffer probe), max flow 100 m³/h p.27 ↗ | Pump sizing outside this envelope degrades performance; flows >100 m³/h bring no benefit | ASM 340 Integrable model |
| Neutral gas purge supply: 200 hPa relative excess pressure, 50 sccm flow (at 1 bar abs inlet); usable pressure 0–0.3 bar relative (1–1.3 bar absolute); manual lockout valve required within 3 m p.25 ↗ | Over-pressure can keep the inlet vent valve closed; plumbing must include an accessible shutoff | Dry and Integrable models (Wet purge line is permanently closed) |
| Detector exhaust pressure must not exceed 200 hPa (relative); the line must always be at slightly negative pressure p.26 ↗ | Long or restricted exhaust ducting can damage the detector; keep exhaust unobstructed | All models (Dry model exhaust filter must never be removed) |
| Maximum permitted weight at the detector inlet is 15 kg and maximum tightening torque is 10 N·m p.27 ↗ | Heavy or over-torqued test connections risk damaging the inlet flange/turbo assembly | All models |
| Parts connected to the inlet must withstand a negative pressure of 1·10³ hPa relative to atmospheric; inlet pressure must not exceed atmospheric p.22 ↗ | Test objects must be mechanically rated for full vacuum before connection | All models |
| Leave ≥10 cm free space all around the detector, keep ventilation grids clear, store nothing under the detector p.22 ↗ | Prevents overheating of internal components in enclosed installations | All models |
| Wet model pumps are shipped without oil; oil must be filled to the maximum level mark before first start-up using only Pfeiffer Vacuum recommended oil p.23 ↗ | First commissioning of a Wet unit requires an oil-filling step; other oils degrade pump performance and void support | ASM 340 Wet model |
| Wait 5 minutes after power-off before working on the product, removing covers or moving it p.12 ↗ | Internal components remain live/hot after switch-off; service scheduling must include this delay | All models |
| Transport components weighing more than 20 kg only with lifting gear; unpacking uses a 3-strand strap with >500 mm strands and >100 kg load per strand p.12 ↗ | Lifting equipment and slings must be specified for receipt/installation planning | All models |
| Wet model uses a 15 m³/h rotary vane backing pump; Dry model uses a diaphragm pump for oil-free applications p.19 ↗ | Variant selection depends on contamination tolerance of the test application | Wet vs Dry variants |
| Storage box on the detector has a maximum permitted load of 5 kg p.20 ↗ | Accessories carried in the box must not exceed 5 kg | All models |
| Critical faults are signaled as 'Critical error - E244' requiring service-center intervention; warnings and errors are displayed on the control panel via [i Next] p.33 ↗ | Field troubleshooting limited to warnings/errors; E244 escalates to service | All models |
| Extended storage procedure keeps the detector under vacuum: blank-off inlet, hard vacuum method, most sensitive mode, inlet vent 'Operator', run test then power off p.18 ↗ | Reduces degassing time at next start-up; recommended for shutdowns >3 months | All models |
| After a mains power failure the detector switches on again automatically when power is restored p.32 ↗ | Automated installations must account for unattended restart behavior | All models |
| Use flexible metal pipework at the inlet; rigid or flexible plastic tubing (compressed-air style) is prohibited; pipes should be short and matched to inlet diameter p.27 ↗ | Plastic tubing outgasses/permeates and kills detection sensitivity; specify metal connections | All models |
| External backing pump contamination (particles, oil backstreaming) damage is excluded from warranty; user must protect the detector p.28 ↗ | Integrable installations should include filtration/traps on the external pump line | ASM 340 Integrable model |
| Only plastic DN 25 ISO-KF blanking plate (supplied) may be used when dispatching the Integrable detector — never a metal blanking plate with clamp p.28 ↗ | Shipping configuration requirement to avoid damage | ASM 340 Integrable model |
| Detector is Class I equipment and must be earthed; only original manufacturer mains cables may be used p.26 ↗ | Installation must include a properly grounded mains supply; replacement cables must be ordered from Pfeiffer | All models |
| Communication interface options include RS-232 (D-Sub 9), USB, Ethernet, PROFIBUS DP-V1, PROFINET and EtherNet/IP, ordered per configuration p.20 ↗ | PLC/integration connectivity must be selected at ordering time | All models, order-dependent option |