ATP 2300 M — instruction manual
Technical Reference Summary
Covered Models & Document Scope
Operating instructions (manual ref. 112419, ed. 06 June 2013) for the adixen ATP 2300 M magnetically levitated turbopump, in ISO-K DN 200 and DN 250 inlet flange variants (also documented: ISO-F and CFF flange mounting options). Powered by OBC or MAGPOWER controllers (covered in separate manuals). Includes accessory part numbers (inlet screens, purge valve, air inlet valve, water valve, bolt kits).
Section-by-Section Technical Breakdown
The ATP 2300 M is a multi-staged magnetically levitated turbomolecular pump for light gases, with 5 active axes of magnetic suspension and Unbalanced Force Rejection Control for low noise/vibration. It is maintenance-free and battery-free: on power failure the motor acts as a generator to keep the magnetic bearings energized until the rotor lands on dry back-up ceramic ball bearings (used only during air in-rushes, shocks, or power failure). Speed can be set between a STANDBY speed and the nominal (factory) speed to optimize pumping for the application.
- Rotor suspended by 5 active magnetic bearings: 3 controlled translations (X, Y, Z) and 2 controlled tilts (X, Y), monitored by position sensors
- Unbalanced force rejection control lets rotor rotate about its inertia axis and auto-compensates deposit build-up over pump life
- Dry-lubricated ceramic back-up ball bearings only engage on accidental events; replacement interval depends on number of landings (see D 150)
- N2 purge protects magnetic spindle from corrosion and assists rotor cooling
- Works with OBC or MAGPOWER controllers; this manual only covers pump-to-equipment connections
Full specification table for the ATP 2300 M in DN 200 and DN 250 ISO-K variants: pumping speeds per gas, compression ratios, throughputs, electrical data, cooling and environmental limits. Also gives flange dimension tables for ISO-F, ISO-K and CF-F inlet flanges and the port/cooling connection layout.
- Pumping speed N2: 1450 l/s (DN200) / 1950 l/s (DN250); He 1750/2050 l/s; H2 1600/1700 l/s; Ar 1400/1900 l/s (without inlet screen)
- Compression ratios: N2 >1e8, Ar >1e8, He >1e5, H2 >7e3; ultimate pressure <4e-9 hPa (ISO/F flange, better with CFF)
- Gas throughput at nominal speed, 25 °C water: N2 2200 sccm, Ar 900 sccm, H2 >3000 sccm, He >3000 sccm
- Fore vacuum max: N2 2.1 hPa, Ar 1.7 hPa, He 1.5 hPa, H2 1.9 hPa (with inlet/exhaust pressure ratio >100)
- Max inlet vacuum (throughput condition): Ar 0.011 hPa, N2 0.04 hPa, He 0.2 hPa
- Nominal speed 31,000 rpm (517 Hz); standby speed adjustable 15,000–31,000 rpm; run-up to 90% nominal <8 min (exhaust <0.1 hPa)
- Controller supply 200–240 V, 50/60 Hz; max leakage current <12 mA; max start-up <1000 W / <1250 VA; consumption at ultimate pressure <130 W
- Sound level <46 dB(A); vibration <0.01 µm; IP42; integral leak rate <5e-8 hPa·l/s; weight 52 kg (112 lb)
- Water cooling: 60 l/h, line pressure max 7 bar, water temperature 15–40 °C (derate above 25 °C); ambient 5–45 °C; max baking 120 °C
- Motor (max): 36 Vrms between phases, 1033 Hz, 9.7 A; mounting orientation: any; recommended backing pumps adixen ADP/ADS
- Ports: inlet ISO-K DN200/DN250, exhaust DN40 ISO-KF, N2 purge DN16 ISO-KF, water in/out 1/4 NPT female; 6 x M10-6H holes along 190 mm circle for handling bracket
Lists optional accessories with part numbers: inlet screens, purge flow reduction device, bolt kits, purge valve (50 sccm), air inlet valve, water valve, and their cables. Dimension drawings are provided for the purge valve, air inlet valve, and water valve.
- Inlet screens: DN200 stainless steel convex filter + clip, mesh 4.5 mm, P/N 108872; DN250 mesh 5 mm, P/N 108762
- Purge flow reduction device DN16, 50 sccm, P/N 066752
- Air inlet valve, 24 V DC, P/N 114280 (NC valve on exhaust, opens below 10,000 rpm to reduce braking time); purge valve 50 sccm 24 V DC P/N 111921S; water valve P/N 108668
- Cables: 1 m A462403-010, 3.5 m A462403-035, 5 m A462403-050, 10 m A462403-100, 20 m A462403-200
- Bolt kits for ATP 2300 M: DN200 ISO-F P/N 114510 (12 studs M10x35); DN250 ISO-F P/N 110034 (12 CHC M10x30); DN250 ISO-F without centering ring P/N 110676S (12 CHC M10x35 with centering ring); DN250 CFF P/N 118690 (32 studs M8x35)
- Valve cables require SELV-compatible power line
Safety requirements for handling and installing the pump: heavy-lift precautions, hot surfaces (>70 °C), electrical hazards with capacitor/motor-generator residual charge, pressurized N2 purge and water circuit, LO/TO per OSHA 29 CFR 1910.147, and hazardous process gas handling including LEL monitoring and purge-loss interlocks. Also covers magnetic field and radiation limits.
- Handling: use lifting rings provided, slings longer than 500 mm; pump weighs 52 kg
- After power-off, wait at least 5 minutes after complete rotor stop before working on connector pins (capacitors >60 VDC; motor may act as generator)
- Hazardous energies: release N2 purge line and water circuit pressure before servicing (place ≥1 l vessel under water outlet)
- No LO/TO device on pump: lock out/tag out the entire tool per OSHA 29 CFR 1910.147; if IEC connector access is restricted, add an accessible isolation device; customer main isolator must interrupt ≥10 kA
- Flammable gases: hardware LEL detection at 25% LEL in exhaust extraction required to stop gas supply; pyrophoric above LEL requires N2 dilution below LEL with purge-loss interlock
- Magnetic field: guaranteed operation up to 0.5 mT; 0.5–5 mT depends on cooling/gas load; >5 mT requires shielding (eddy-current rotor heating); standalone radiation resistance up to 1e3 Gy
- All connected parts must withstand 100 kPa negative pressure below atmosphere; always connect inlet before starting (rotor cut hazard)
- EMC: use shielded cables in noisy environments; electrical installation per local/national rules with fuses, breaker, proper earth
The pump operates in any orientation and its enclosure withstands 1.1 MPa static overpressure for 1 minute. The frame must absorb rotor crash kinetic energy; worst-case loads were measured with the rotor splitting in two at nominal speed. Strict inlet flange bolting specifications (bolt type, number, torque) apply, and ISO-K flanges are discouraged.
- Worst-case crash at 31,000 rpm, rotor energy 116 kN·m; transient loads: torque max 50 kN·m (3.0 ms, 2.5 ms delay), bending moment max 35 kN·m (1.5 ms), axial force up to 650 kN (1.0 ms) only if inlet obstructed with high-stiffness parts (no load on low-stiffness pendulum valve)
- DN250 ISO-F / DN200 ISO-K with centering ring: 12 x M10 bolts, grade 12-9, length 35 mm, torque 30 N·m, total clamping force 161,500 N
- DN250 ISO-F without centering ring: 12 x M10, 30 mm, 30 N·m, only special washer bolts kit P/N 110676S; DN250 CFF: 32 x M8, length 40 mm, torque 20 N·m, clamping force 355,000 N
- Use a torque wrench: lower torque risks loosening, higher torque damages bolts
- ISO-K flanges not recommended: no visual reminder of clamp count, harder to fasten clamps, no anti-rotation in a crash (could damage foreline/purge line)
- Do not use bellows on the pump inlet flange; if frame cannot bear crash loads, add a bracket to the 6 x M10 holes at pump bottom (contact manufacturer for calculation)
- Materials in gas path: stainless steel/aluminium alloy; O-rings fluorinated elastomer (FPM)
Inlet must always have the inlet screen fitted (delivered with pump); an isolation valve is recommended at the inlet, and a venting/roughing valve (V1) close to the exhaust, opened via the controller START contact. Missing vent valve increases slow-down time and reduces back-up bearing service life.
- Install screen: place screen in inlet housing groove with bent side opposite rotor, press retaining ring in manually all around
- Isolation valve between chamber and pump recommended for venting chamber without stopping pump; at exhaust, isolation valve (closed at power-off) highly recommended between turbo and roughing circuit
- Leak-check entire connection after installation; use dry clean pipelines and gloves (pollution risk)
- Turbo cannot evacuate at atmospheric pressure — a roughing pump is mandatory; transient start at atmospheric pressure is allowed
A filtered dry nitrogen supply of specified purity is connected continuously to the DN16 purge port; a flow reduction device guarantees 50 sccm at 110 kPa (or use a mass flow meter). Recommended purge flow is 50 sccm; purge pressure must never exceed 200 kPa.
- N2 supply requirements: H2O <1 ppm, O2 <1 ppm, dew point < -22 °C (documented as '< 22 °C'), dust <1 µm, oil vapor <0.1 ppm, absolute pressure 100–120 kPa
- Flow reduction device P/N 066752 gives 50 sccm at 110 kPa; max purge pressure 200 kPa; purge must connect directly to the pump purge port
- Adjust flow per supplied pressure/flow scale (30–70 sccm over ~1.0–1.5 kPa inlet absolute pressure)
- Cooling water: 60 l/h, max line pressure 7 bar (7e3 hPa), water 15–40 °C, derate above 25 °C; air inlet valve mounted on exhaust reduces braking time
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pumping speed for N2 is 1,450 l/s (DN200 inlet) and 1,950 l/s (DN250 inlet), measured without inlet screen p.2300 ↗ | Choose DN250 variant when higher conductance/speed is needed; account for inlet screen losses in sizing | ATP 2300 M, both inlet variants |
| Nominal rotation speed is 31,000 rpm (517 Hz); standby speed is settable from 15,000 rpm (250 Hz) to nominal p.12 ↗ | Speed can be reduced via controller for high-pressure or application-optimized pumping | whole ATP 2300 M series |
| Run-up time to 90% of nominal speed is <8 min at exhaust pressure <0.1 hPa p.13 ↗ | Tool cycle-time budgets and interlock timing must allow for this start-up period | whole series (controller-dependent) |
| Ultimate pressure <4e-9 hPa (ISO/F flange; better with CFF flange); compression ratios: N2 and Ar >1e8, He >1e5, H2 >7e3 p.12 ↗ | For H2-rich processes the low H2 compression ratio mandates an adequate backing pump | whole series |
| Controller power supply 200–240 V, 50/60 Hz; max start-up <1,000 W / <1,250 VA; consumption at ultimate pressure <130 W; max leakage current <12 mA p.12 ↗ | Size supply circuit and leakage-current protection (RCD) accordingly | whole series (values depend on controller type) |
| Cooling water flow 60 l/h at max line pressure 7 bar (7e3 hPa), water temperature 15–40 °C with derating above 25 °C; max baking temperature 120 °C p.13 ↗ | Chilled-water loop and flow monitoring must be designed to these limits; bakeout above 120 °C not permitted | water-cooled ATP 2300 M |
| Recommended N2 purge flow 50 sccm; purge supply 100–120 kPa absolute with H2O and O2 <1 ppm, dust <1 µm, oil vapor <0.1 ppm; max purge pressure 200 kPa p.13 ↗ | Install the flow reduction device (P/N 066752) or an MFC; connect purge directly to the pump purge port for corrosion protection | whole series |
| Worst-case rotor crash (rotor split in two at 31,000 rpm, energy 116 kN·m) produces torque up to 50 kN·m (3 ms), bending moment up to 35 kN·m (1.5 ms), and axial force up to 650 kN only if the inlet is obstructed with stiff parts p.30 ↗ | Equipment frame, attachment devices and flange bolting must be designed for these transient loads; avoid stiff obstacles close above the inlet | ATP 2300 M installation |
| Inlet flange bolting: 12 x M10 grade 12-9 bolts torqued to 30 N·m for DN250 ISO-F/DN200 ISO-K; 32 x M8 bolts at 20 N·m for DN250 CFF; total clamping force 161,500 N (ISO) or 355,000 N (CFF) p.31 ↗ | Use a torque wrench at exactly these values — under-torque risks loosening, over-torque damages bolts | ATP 2300 M inlet flange mounting |
| ISO-K inlet flanges are not recommended for securing the pump; ISO-F or CF-F preferred p.32 ↗ | ISO-K lacks anti-rotation in a crash and offers no visual clamp-count reference — choose ISO-F/CF-F flanges in the chamber design | whole series |
| Pump enclosure withstands 1.1 MPa static internal overpressure for 1 minute; all connected parts must withstand 100 kPa below atmospheric pressure p.28 ↗ | Connected chambers/lines must be vacuum-impervious; enclosure rupture is not the failure mode in overpressure incidents | whole series |
| Fore vacuum maximum: 2.1 hPa (N2), 1.7 hPa (Ar), 1.5 hPa (He), 1.9 hPa (H2), with inlet/exhaust pressure ratio >100 p.12 ↗ | Backing pump and roughing-line sizing must keep foreline pressure below these values under full gas load | whole series |
| Environmental limits: ambient 5–45 °C, indoor use only, max altitude 2,000 m, max relative humidity 80% up to 31 °C decreasing linearly to 50% at 40 °C, pollution degree 2, overvoltage category II p.13 ↗ | Installations above 2,000 m or in outdoor/humid environments require special consideration | whole series |
| Pump weight 52 kg (112 lb); handling requires lifting rings and slings longer than 500 mm; center of gravity given in A400 dimensions p.14 ↗ | Lifting gear and frame design must be rated for 52 kg with tilt risk considered | whole series |
| After power-off, capacitors may hold >60 VDC and the motor can act as a generator; wait at least 5 minutes after the rotor has fully stopped before touching connector pins p.23 ↗ | Service procedures and lockout instructions must include this 5-minute wait | whole series |
| Operation guaranteed in ambient magnetic fields up to 0.5 mT; 0.5–5 mT depends on cooling and gas load; above 5 mT shielding is required; standalone radiation resistance up to 1e3 Gy p.24 ↗ | For accelerators/physics installations, map the field at the pump location and add magnetic shielding if needed | whole series |
| For flammable gases, hardware-based LEL detection at 25% LEL in the exhaust extraction system is required to stop chemical supply; purge-loss interlock required if purge loss is a significant risk p.22 ↗ | Tool integrator must add exhaust monitoring and interlocks; these are not built into the pump | hazardous-gas installations |
| Air inlet valve (24 V DC, P/N 114280) is a NC valve on the exhaust that activates below 10,000 rpm to vent the pump and reduce braking time; without it, slow-down time increases and back-up bearing life shortens p.16 ↗ | Specify the vent valve on all systems to preserve back-up bearings; it must be SELV-powered | whole series |
| Integral leak rate <5e-8 hPa·l/s; vibration <0.01 µm; sound pressure <46 dB(A); protection index IP42 p.12 ↗ | Suitable for quiet, vibration-sensitive tools (electron microscopy, surface analysis); not suited for wet/dusty environments beyond IP42 | whole series |
| New pumps store up to 2 years at -5 to +50 °C; pumps run on clean process store max 6 months (after water-line blow-out and N2 purge); pumps from corrosive processes must never be stored p.26 ↗ | Return corrosive-process pumps to service center (E100 shipping procedure) instead of warehousing them | whole series |