DuoVane 36 — instruction manual 70
Technical Reference Summary
Covered Models & Document Scope
Instruction manual for Pfeiffer Vacuum + Fab Solutions DuoVane 36 and DuoVane 70 two-stage oil-sealed rotary vane vacuum pumps. Covers four variants: Standard (DuoVane 36/70), C version (operating fluid F5, changed vane material, hose connection and dosage spindle on gas ballast valve, oiler for shaft feedthrough, leak rate ≤1×10⁻⁷ Pa m³/s), M version (magnetic coupling, leak rate ≤1×10⁻⁷ Pa m³/s), and MC version (combines C-version features with magnetic coupling). Pumping speeds: DuoVane 36 = 32 m³/h (50 Hz) / 36 m³/h (60 Hz); DuoVane 70 = 62 m³/h (50 Hz) / 70 m³/h (60 Hz).
Section-by-Section Technical Breakdown
Specifies installation and operating limits (protection classes, temperatures, pressures, inclination), intended use for vacuum generation only, and prohibited uses. Defines personnel qualification requirements and safety precautions.
- Installation location: weatherproof (internal space); protection class I; pollution degree 2; overvoltage category II (IEC 61010)
- Protection class: IP40 (single-phase motor), IP43 (three-phase motor); housing Type 1 (UL 50E)
- Permissible inclination: ±1° longitudinal, ±2° transverse
- Ambient temperature +12 °C to +40 °C; relative humidity max. 85%; intake temperature max. +40 °C
- Exhaust pressure: ≥ atmospheric, ≤ 1500 hPa absolute; constant intake pressure max. 50 hPa absolute
- Prohibited: explosive, corrosive (except C version), radioactive media, fluids, dust, sublimating media, pressure generation, Ex areas, strong magnetic/electromagnetic fields
- For oxygen concentration ≥ 21%, only perfluorinated synthetic oils (F4, F5, A113) as operating fluid; never mineral oil
Describes the two-stage, oil-sealed rotary displacement pump design with asynchronous motor and claw coupling, gas ballast, vacuum safety valve and operating fluid functions. Identifies pump features and delivery scope.
- Two-stage, oil-sealed rotary displacement pumps for low and medium vacuum
- DuoVane 36: 32 m³/h at 50 Hz, 36 m³/h at 60 Hz; DuoVane 70: 62 m³/h at 50 Hz, 70 m³/h at 60 Hz
- Integrated gas ballast supplies ambient air or inert gas to reduce condensate
- Vacuum safety valve separates pump from chamber at standstill, vents system so oil does not migrate; valve opens with delay after switch-on
- Motor options: three-phase with built-in thermal winding protection, or single-phase with reversible voltage range, thermal protection switch, mains switch and rubber connector
- Scope of delivery: pump with motor, operating fluid (other than F4/F5/A113), locking caps for both flanges, centering ring, instruction manual
Covers setup with inclination limits, repositioning vacuum/exhaust flanges, vacuum and exhaust-side connections, electrical safety measures and mains connection with 6-pin or 9-pin terminal boards, motor protection setting and rotation check. Includes requirements for frequency converter and soft-start operation.
- Surface can exceed 70 °C — provide touch protection if unrestrictedly accessible
- Flange repositioning: WAF 5 hex key, calibrated torque wrench (tightening factor ≤ 2.5), tightening torque 5 Nm
- Vacuum/exhaust line cross section ≥ nominal diameter of connection flange; support piping so no forces act on pump; route exhaust piping downwards
- Exhaust line without shut-off units; exhaust pressure ≥ atmospheric, ≤ 1500 hPa abs; negative exhaust pressure not permissible; condensate separator recommended at lowest point
- Pump has no mains switch: install mains disconnection device, MCB with interruption rating ≥ 10 kA, and RCCB
- Three-phase motors: direct on-line starting only — never star-delta start-up circuit
- IE2 motor: soft-start relay or frequency converter recommended; IE3 motor: soft start or frequency converter mandatory (magnetic coupling demagnetization risk)
- Recommended mains fuse: type 'K' circuit breaker with slow tripping characteristics
- Rotation direction (6-pin board): U1–L2, V1–L1, W1–L3 give clockwise shaft rotation viewed at motor fan (9-pin: T1–L2, T2–L1, T3–L3)
- Motor automatically restarts after power failure or thermal standstill — secure against restart during work
Covers commissioning, switching on/off, operating with gas ballast (standard valve and inert-gas connection version) and refilling operating fluid. Gas ballast operation reduces condensate in the pumping system.
- Gas ballast valve standard version and variant with inert gas connection described
- Gas ballast supports reduction of condensate accumulating in the pumping system
- Refill operating fluid during operation as required (sight glass level)
Maintenance instructions covering magnetic coupling care, inspection/maintenance checklist intervals, operating fluid change and aging determination, gas ballast valve removal/cleaning, silencer nozzle cleaning and vacuum safety valve diaphragm replacement.
- Dedicated maintenance instructions for the magnetic coupling (M/MC versions)
- Checklist for inspection and maintenance intervals provided
- Operating fluid change procedure including determination of aging degree of P3 operating fluid
- Oiler oil level check for C version only
- Rinsing/cleaning of rotary vane pump, gas ballast valve disassembly/cleaning, silencer nozzle cleaning
- Diaphragm replacement on vacuum safety valve described
Shutdown procedures for longer periods, recommissioning steps, disposal guidance, and a malfunction/troubleshooting chapter with a service solutions section.
- Shutting down for longer periods and recommissioning procedures given
- General disposal information and disposal procedure per chapter 9
- Malfunction troubleshooting table in chapter 10; Pfeiffer service solutions in chapter 11
Ordering information for spare part packages with their contents, and accessory ordering information with compatible accessories from the Pfeiffer range.
- Spare part packages with defined contents (chapter 12)
- Accessory information and ordering data (chapter 13)
General technical data, substances in contact with media, dimension drawings for all six pump versions (36/36MC/36M and 70/70MC/70M) and EU/UK declarations of conformity.
- Dimension drawings for DuoVane 36, 36MC, 36M, 70, 70MC, 70M
- Substances in contact with media table provided
- EU and UK Declarations of Conformity included as chapters 15/16
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| DuoVane 36 pumping speed is 32 m³/h at 50 Hz and 36 m³/h at 60 Hz; DuoVane 70 is 62 m³/h (50 Hz) / 70 m³/h (60 Hz) p.15 ↗ | Size the pump for the actual mains frequency; 60 Hz operation gives ~10-13% higher throughput | whole series |
| Constant intake pressure max. 50 hPa absolute; exhaust pressure ≤ 1500 hPa absolute and ≥ atmospheric; max. intake medium temperature +40 °C p.9 ↗ | Applications requiring continuous operation below 50 hPa or throttled exhaust above 1500 hPa abs are outside the allowed envelope | whole series |
| Ambient temperature range +12 °C to +40 °C, relative humidity max. 85%, permissible inclination ±1° longitudinal / ±2° transverse p.9 ↗ | Installation site must be climate-controlled, weatherproof and level within these tilt limits to safeguard operating fluid supply | whole series |
| Protection class IP40 with single-phase motor and IP43 with three-phase motor; housing Type 1 (UL 50E); overvoltage category II, pollution degree 2 p.9 ↗ | Pump is not suited for dusty or wet environments without additional enclosure; verify electrical installation category accordingly | whole series |
| M and MC versions have leakage rate ≤ 1×10⁻⁷ Pa m³/s; C and MC use operating fluid F5 with changed vane material p.5 ↗ | Choose M/MC for tight applications (e.g. helium leak-test background) and C/MC where corrosive media demand special fluid/vanes | DuoVane 36/70 M, C, MC versions |
| For media with oxygen concentration ≥ 21%, only perfluorinated synthetic oils (F4, F5, A113) may be used; mineral oil is forbidden at > 21% O₂ p.10 ↗ | Prevents oil oxidation/ignition risk; specify F-series oil at ordering for oxygen service | whole series |
| Pump has no built-in mains switch; a mains disconnection device with MCB (interruption rating ≥ 10 kA) and RCCB must be installed p.22 ↗ | Panel design must include labelled disconnecting means within reach of the pump | whole series |
| Three-phase motors must always be started direct on-line; star-delta start-up is prohibited and causes engine damage p.24 ↗ | Do not fit star-delta starters in the supply concept; use DOL contactors | three-phase motor versions |
| With IE2 motors a soft-start relay or frequency converter is recommended; with IE3 motors it is mandatory due to magnetic coupling demagnetization risk p.23 ↗ | For M/MC (magnetic coupling) versions with high-efficiency motors, budget a soft-start relay or VFD in the drive spec | M and MC versions (magnetic coupling), esp. IE3 motors |
| Recommended mains fuses are type 'K' circuit breakers with slow tripping characteristics p.23 ↗ | Avoid fast-curve breakers which nuisance-trip on pump start current | whole series |
| Rotation direction wiring: 6-pin board U1–L2, V1–L1, W1–L3; 9-pin board T1–L2, T2–L1, T3–L3 give clockwise shaft rotation viewed at motor fan p.24 ↗ | Use these assignments to get correct rotation at commissioning before checking direction of rotation | three-phase motor versions |
| The motor restarts automatically after power failure or thermal standstill p.23 ↗ | Implement lock-out/tag-out and secure against restart before any work near the vacuum flange | whole series |
| Interchangeable flange screws are tightened to 5 Nm using WAF 5 hex key and a calibrated torque wrench (tightening factor ≤ 2.5) p.19 ↗ | Enables safe repositioning of vacuum/exhaust flanges between vertical and horizontal positions without seal damage | whole series |
| Vacuum and exhaust line cross sections must be ≥ nominal flange diameter; piping must be supported so no forces act on the pump; exhaust routed downwards with no shut-off units p.20 ↗ | Prevents piping stress on the housing and condensate return; negative exhaust pressure is not permissible | whole series |
| Surface temperature can exceed 70 °C; touch protection required for unrestricted access; keep cables away from surfaces > 70 °C p.9 ↗ | Plan guarding/clearances and cable routing for burn protection in the layout | whole series |
| The vacuum safety valve separates the pump from the chamber at standstill and vents it so oil does not migrate into the chamber; valve opens with delay after switch-on p.14 ↗ | No external anti-suck-back valve is required for basic isolation, but account for the opening delay in process sequencing | whole series |
| Operating fluid must be changed after storage periods longer than 2 years; pump is stored filled to upper edge of sight glass with gas ballast valve closed p.17 ↗ | Include oil-change and level checks in storage/recommissioning SOPs | whole series |
| Magnetic coupling does not provide motor overload protection; motor protection switch setting procedure is required p.25 ↗ | Motor protection must be set per nameplate current even for M/MC versions | whole series, esp. M/MC versions |