QMG 250 F1 — instruction manual qmg 250
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum QMG 250 PrismaPro compact quadrupole mass spectrometer operating instructions (90 pages). Covers the complete system: QMA 250 analyzers (PT M25 411–439) in F (Faraday) and M (Faraday/EM) versions, mass ranges 1–100 / 1–200 / 1–300 u (F1/F2/F3, M1/M2/M3), with open, crossbeam, grid or gas-tight ion sources, tungsten (W) or Ir-Y2O3 filaments; and QME 250 electronic units (PT M28 641–673) in standard/expanded (with IO 250) and straight/90° angled versions. Power supply SP 250 and PV MassSpec software included.
Section-by-Section Technical Breakdown
Describes the four main components (analyzer, electronic unit, SP 250 power pack, PV MassSpec software), analyzer structure and ion source types, the QME 250 front panel, and all interfaces with pinouts. Operating principle is electron-impact ionization with quadrupole mass filtering and Faraday/EM detection.
- Scope of delivery: QMG 250, SP 250 supply, copper gasket DN 40 CF, assembly set, Ethernet cable 3 m, D-Sub plugs, test report on device, PV MassSpec via PV Cloud
- QMA 250 F (Faraday only) supplied with open ion source only; QMA 250 M optionally open, crossbeam, grid or gas-tight ion source
- Faraday detector for RGA at total pressure > 10^-7 hPa; EM used at very low total pressure and high measuring speed
- Ion source types: open (max sensitivity), crossbeam (molecular beam), grid (UHV RGA < 10^-8 hPa), gas-tight (quantitative gas analysis)
- M analyzer EM variant requires DN 40 CF-F intermediate piece, 116 mm length
- Interfaces: 24 V=30W power socket, RJ-45 LAN, 15-pin AUX I/O, 62-pin HD EXT I/O (with IO 250), 6-pin Amphenol TP GAUGE, USB (service only)
- AUX I/O: relay (24 V DC / 0.5 A), analog input 0 to +10 V, digital inputs DI0/DI1 preset for emission ON/OFF control
- EXT I/O: 8× 16-bit analog outputs 0–10 V, 8× 16-bit differential analog inputs ±10 V, 10 open-collector digital outputs, 5 low-active digital inputs, +24 V max 1 A
- TP GAUGE socket accepts Pfeiffer ActiveLine/DigiLine gauges (AR option) for pressure-protection interlock; USB reserved for service
Covers mechanical installation of the PrismaPro and analyzer, assembly of the electronic unit, electrical and Ethernet connections, gauge connection, network configuration (IP addresses, sub-networks) and PV MassSpec installation. High-voltage safety warnings apply throughout.
- Analyzer fastening requires defined tightening torques (Tbl. 9) and proper grounding of the analyzer
- Ethernet connection via RJ-45: TD+/TD- (pins 1/2), RD+/RD- (pins 3/6); green LED = link, yellow = activity
- IP address configuration: DHCP or manual; procedure given for changing PrismaPro and host PC IP addresses; sub-network examples provided
- Gauges connectable via TP GAUGE socket (Pfeiffer ActiveLine/DigiLine)
- PV MassSpec requires Windows 7 or higher
Commissioning and switching on the device, then operation via PrismaPro Web UI or PV MassSpec software. Data storage, evaluation and comparison of measurements is only possible with PV MassSpec.
- Operation via Web UI or PV MassSpec; detailed Web UI instructions in document BG 6002
- PV MassSpec required for saving/evaluating measured data
Maintenance divided into three levels (1 customer, 2 technically educated customer/PV service, 3 PV-service-trained). Includes visual inspection of the QMA 250, bakeout, filament replacement for all four ion source types, and HF/mass-scale calibration with gas mixtures (FC-5311 up to 624 u) or process gases.
- Bakeout temperatures limited per ion source type (Tbl. 13); surfaces can exceed 100 °C — wear protective gloves
- Filament units replaceable for open, crossbeam, grid and gas-tight ion sources (dedicated procedures per source)
- Mass calibration (TUNE): stabilize operating temperature first, select calibration gas mixture, set high frequency for both polarities
- Test gas mixture FC-5311 allows mass scale calibration up to 624 u
- Helium leak detection calibration procedure included
- Test report accessible on device
Error tables for system/communication errors, hardware errors, and operating/measuring errors, plus PV MassSpec software error messages. Includes contact diagrams for QMA 250/QME 250 and measurement-based analyzer checking procedures.
- Structured troubleshooting tables: system/communication (Tbl. 15), hardware (Tbl. 16), operating/measuring (Tbl. 17), software errors (Tbl. 18)
- Analyzer check measurements (Tbl. 19) with contact locations (Fig. 32) for diagnosis
Safe shipping of possibly contaminated products, disposal guidance for the compact mass spectrometer, and Pfeiffer Vacuum service solutions. Decontamination required for parts exposed to toxic process media.
- Decontaminate contaminated parts before maintenance/shipping/disposal
- Pfeiffer Vacuum service and training offerings (maintenance levels 2 and 3)
Ordering data for QME 250 electronic units, QMA 250 analyzers, filament units, spare parts and small parts sets with part numbers. Maintenance level 1 parts are customer-replaceable.
- Small parts sets: PT 163 544 (open IS), PT 163 543 (crossbeam IS), PT 163 534 (grid IS), PT 163 550 (gas-tight IS)
- QME 250 ordering matrix: mass range M1/M2/M3 × standard/expanded (IO 250) × straight/90° angled
Technical data tables for QMG 250 F and QMG 250 M, pressure/gas-throughput unit conversion tables, and dimension drawings for all combinations of ion source type and electronic unit orientation (straight/90°). Includes NRTL certification and EC/UK declarations of conformity.
- Separate data tables for QMG 250 F (Tbl. 30) and QMG 250 M (Tbl. 31)
- Dimension drawings for open, crossbeam, grid and gas-tight ion sources, each in straight and 90° QME 250 versions
- Dimensions in mm unless stated otherwise
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| QMG 250 PrismaPro analyzers are available in mass ranges 1–100 u (F1/M1), 1–200 u (F2/M2) and 1–300 u (F3/M3). p.8 ↗ | Select the mass-range variant according to the highest gas species mass that must be resolved; 300 u covers FC-5311 calibration up to 624 u only partially. | QMA 250 F/M series |
| QMA 250 F analyzers are supplied only with open ion source; QMA 250 M offers open, crossbeam, grid or gas-tight ion sources. p.21 ↗ | Faraday-only systems can only be ordered in the open-source configuration; specialized sources require the EM (M) analyzer. | QMA 250 F vs QMA 250 M |
| Grid ion source is specified for residual gas analysis in UHV (< 10^-8 hPa); Faraday detector for RGA at total pressure > 10^-7 hPa. p.21 ↗ | Choose grid source for UHV applications; below ~10^-7 hPa use EM-equipped analyzer for adequate sensitivity. | QMA 250 M |
| Filaments are tungsten (W) or Ir-Y2O3, with nickel wiring for open/gas-tight/grid sources and silver-plated copper wiring for crossbeam sources. p.8 ↗ | Ir-Y2O3 tolerates higher oxygen exposure/rarer burnout; wiring material must match the crossbeam variant when replacing parts. | all QMA 250 analyzers |
| Power input socket is 24 V DC, 30 W (4-pin lockable mains part socket, internally isolated from system earth). p.23 ↗ | Ensure the SP 250 power supply pack delivers 24 V/30 W; do not substitute supplies. | QME 250 |
| AUX I/O relay output contact load: 24 V DC at 0.5 A; +24 V output on pin 8 limited to max 1 A. p.24 ↗ | Size external loads/relays accordingly; do not exceed 1 A through the 24 V auxiliary output. | QME 250 (AUX I/O) |
| Emission can be hard-controlled by digital inputs: emission ON = pin 14 + pin 15 connected; emission OFF = pin 13 + pin 15 connected; this bypasses all software and hardware locks. p.25 ↗ | An external interlock (e.g., pressure switch) is mandatory to prevent filament activation at excessive pressure. | QME 250 (AUX I/O) |
| EXT I/O (with IO 250): 8× 16-bit analog outputs (0–10 V), 8× 16-bit differential analog inputs (±10 V), 10 open-collector digital outputs (DO 2–11), 5 low-active digital inputs (DI 13–17), shared +24 V limited to 1 A total. p.25 ↗ | Process control integration (PLC/analog logging) via the 62-pin HD D-Sub; observe 16-bit resolution and voltage ranges in system design. | QME 250 with IO 250 (PT M28 651–653, 671–673) |
| TP GAUGE socket is a 6-pin Amphenol C 091 A accepting Pfeiffer ActiveLine/DigiLine gauges with Analog-Relay (AR) option; measurement signal 0 to +10 V DC, supply +24 V DC. p.27 ↗ | Use an AR gauge as a hardware protection interlock against unintentional pressure rises before operating the MS. | QME 250 with IO 250 |
| Ethernet is a standard RJ-45 (pins 1/2 TD, 3/6 RD); delivered cable is a 3 m red UTP crossed patch cable; green LED = link, yellow LED = activity. p.24 ↗ | Crossed cable supplied suggests direct PC connection; for switched networks use standard patch cabling; LED states aid link diagnostics. | QME 250 |
| QMA 250 M analyzers with EM require a DN 40 CF-F intermediate piece of 116 mm length between analyzer and electronic unit. p.21 ↗ | Reserve this extra axial length in vacuum-system layout and ordering (part included in delivery of EM variants). | QMA 250 M |
| Analyzer mounting flange connection is DN 40 CF with copper gasket (supplied in scope of delivery); defined tightening torques apply (Tbl. 9). p.19 ↗ | Use only the supplied copper CF gasket and observe torque table to ensure vacuum tightness and analyzer grounding. | whole series |
| After switching off the device, wait 60 seconds before disconnecting all cables (power cable last); the QME must never be operated when opened, and the 24 V supply must never be connected unless the QMA/QME connection is fully screwed tight. p.13 ↗ | Mandatory lockout procedure for any installation or maintenance work; sequencing prevents lethal capacitive/analyst voltage hazards. | whole series |
| Mains cable must be 3-pin with protective earth, double insulation per IEC 61010 / IEC 60950; mains cable connected first (before all other cables). p.14 ↗ | Ensures continuous protective earthing; non-compliant cabling risks fatal electric shock via housing leakage. | SP 250 / whole series |
| During/after bakeout, metal surfaces can exceed 100 °C; maximum bakeout temperatures per ion source type are given in Tbl. 13. p.15 ↗ | Wear protective gloves; do not exceed the ion-source-specific bakeout limit to avoid damaging filaments/insulators. | QMA 250 all ion source types |
| Mass scale calibration uses FC-5311 (perfluoro(tetradecahydrophenantrene) isomer mixture) covering up to 624 u; HF must be set for both polarities before mass calibration (TUNE). p.57 ↗ | For full-range quantitative work, calibrate after thermal stabilization using FC-5311 or process gases of known composition. | QMG 250 F and M |
| The quadrupole MS is a relative measuring instrument requiring calibration with reference gases for accurate quantitative analysis. p.28 ↗ | Do not treat raw ion currents as absolute partial pressures; plan calibration gases for the target species. | whole series |
| Improper use excludes corrosive or explosive media, outdoor use, safety-critical measurements and use outside technical data limits. p.16 ↗ | For reactive/corrosive gas analysis only the gas-tight ion source is intended for quantitative work, and any hazardous-media use voids warranty if outside limits. | whole series |
| USB mini port is reserved exclusively for Pfeiffer Vacuum service personnel. p.27 ↗ | Do not use the USB port for user integration; all PC communication is via Ethernet/PV MassSpec. | QME 250 |
| Maintenance levels: Level 1 = trained customer (spare parts in section 14.2), Level 2 = technically educated customer/PV service, Level 3 = PV-service-trained customer/PV service; PV MassSpec requires Windows 7 or higher. p.18 ↗ | Filament replacement and visual inspection are customer-level tasks; verify PC OS compatibility before software rollout. | whole series |
| Small parts sets for filament maintenance: PT 163 544 (open IS), PT 163 543 (crossbeam IS), PT 163 534 (grid IS), PT 163 550 (gas-tight IS). p.75 ↗ | Stock the set matching the installed ion source to avoid downtime during filament replacement. | QMA 250 per ion source type |